Evaluation of n Order Polynomial Phase Correction in Reprojected Line Scan Phase Contrast MRA
نویسندگان
چکیده
E. J. Offerman, I. Koktzoglou, C. Glielmi, and R. R. Edelman Radiology, Northshore University HealthSystem, Evanston, IL, United States, Radiology, The University of Chicago, Chicago, IL, United States, Siemens Healthcare, Chicago, IL, United States, Radiology, Northwestern University, Chicago, IL, United States Introduction: Reprojected Line Scan Phase Contrast (rLSPC) MRA is a rapid line scan phase contrast method that produces a coronal cine of blood flow and can provide information on the hemodynamic significance of a stenosis [1]. It is acquired in line by line projections in real-time over the cardiac cycle. Although the method can rapidly demonstrate flow velocity over large vascular territories, it suffers from spatially varying phase shifts due to eddy currents and pre-steady-state instabilities. Phase correction (PC) is needed to improve image quality and consistency of the velocity quantification. Presented is a PC method that employs a polynomial curve fitting technique. The merit of the method is assessed and orders of fit are compared. Methods: rLSPC was performed with a 1.5T MR scanner (MAGNETOM Avanto, Siemens Healthcare, Erlangen, Germany) in five healthy subjects and one patient with partial occlusion in the right iliac artery. The sequence consisted of a spoiled gradient echo readout with the following parameters: 9 stations; 60 3-mm-thick slices per station; flip angle 90°; TR=20ms; TE=4.3ms; bandwidth=625 Hz/pixel; VENC=20-150(cm/s). The polynomial fitting technique is a derivation of the 2D polynomial PC method employed by Lankhaar et al. in [2]. An nth order 1D polynomial was fit to pixels of stationary background phase and then subtracted from the same line in the final cine. This was performed for each line in each frame of the cine. PC of different polynomial orders (n = 1 to 8) were compared to themselves and to uncorrected rLSPC using two criteria: (a) flow velocity measurements at four locations within the vessel lumen; and (b) amplitude of background phase fluctuations which was quantitated as the time-series standard deviation of the background phase. Results: Flow velocity in the aorta and anterior tibial artery was not substantially altered by any fitting order (Fig. 1), however measurements at proximal and distal femoral locations showed convergence to a common velocity, suggesting that correction of intra-station phase gradients (see Fig. 3) may improve velocity quantification. Higher fitting orders did not significantly alter proximal and distal femoral velocities. The standard deviation of background phase was reduced 40% with 1st order PC (Fig. 2), which manifested as improved image quality with the reduced background phase variation over time. High order PC (> 3rd order) demonstrated little additional improvement. Figure 3 displays the obvious improvement in image quality of uncorrected rLSPC (top) with 3rd order PC (bottom). Inter-station and intra-station phase shifts were nearly completely eliminated with polynomial PC. Discussion and Conclusion: One-dimensional polynomial PC of rLSPC provides vast improvement over uncorrected rLSPC by eliminating background phase shifts in stationary tissue which can alter the velocity values and reduce image quality. However, this method has several drawbacks. First, in order to achieve optimal fit, the intra-vessel velocities must be segmented from background velocities, which is done using an empirically determined velocity threshold (values > 40 cm/s are ignored in the fit). Velocities in smaller vessels may not be removed, which can reduce the accuracy of the fit. Secondly, the PC process can be computationally intensive and time consuming when high order fitting is used. For instance, 8th order PC took 83% longer than 2rd order PC to perform. Considering this and the marginal improvement seen at high fitting orders, it may be prudent to use low fitting orders (< 3rd order). References: [1] Edelman et al. MRA Club 2010. [2] Lankhaar et al. JMRI 2005;22:73-79. Figure 3. Montage of the first nine frames of the cine without PC (top) and with 3rd order polynomial PC (bottom) in a healthy subject. Temporal resolution is 20ms. With polynomial PC, the substantial phase fluctuations in frames 3-6 of uncorrected rLSPC are reduced and intrastation phase gradients are eliminated. Figure 2. Standard deviation of background (non-vessel) phase vs. fitting order of PC. Each point is the average of five healthy subjects and one patient. Figure 1. Peak flow velocity vs. fitting order of PC. Each point is the average of five healthy subjects and one patient.
منابع مشابه
Flow-encoded raster line scanning (FERAL) of the peripheral arteries
Methods Imaging was performed on a 1.5T Siemens Avanto system. FERAL MRA was acquired as a series of real-time, time-resolved (20 ms/frame) transversal 3mm-thick, one-dimensional phase contrast (paired rephased/ dephased) acquisitions, with each cine series spanning a single R-R interval. Overlapping line scan data are reprojected into the coronal plane for cine viewing and quantitative evaluat...
متن کاملReducing the respiratory motion artifacts in PET cardiology: A simulation study
Introduction: There are several technical features that make PET an ideal device for the noninvasive evaluation of cardiac physiology. Organ motion due to respiration is a major challenge in diagnostic imaging, especially in cardiac PET imaging. These motions reduce image quality by spreading the radiotracer activity over an increased volume, distorting apparent les...
متن کاملBackground phase correction using k-space filters in phase contrast velocity encoded MRI
Figure 5: In-vivo background phase correction. (a) Magnitude image, (b) the original phase image, with the dashed blue line defining the area and (c) the red contour of stationary tissue for polynomial surface subtraction, (d) is ground truth, (e) 5th order polynomial subtraction, and (f) k-space filtering with a 64×64 pixel Gaussian kernel. Figure 1: phantom with steady-state flow. Red lines s...
متن کاملContrast-enhanced whole-heart coronary magnetic resonance angiography at 3 T using interleaved echo planar imaging.
OBJECTIVES The goal of this work was to reduce the scan time of contrast-enhanced whole-heart coronary magnetic resonance angiography (MRA) by using a gradient echo interleaved echo planar imaging (GRE-EPI) sequence at 3 T field strength. MATERIALS AND METHODS A GRE-EPI sequence was optimized to acquire contrast-enhanced whole-heart coronary MRA at 3 T. First-order phase correction was used f...
متن کاملNon-contrast enhanced MR angiography: physical principles.
Noncontrast-enhanced magnetic resonance angiography (NCE-MRA) methods have been demonstrated in anatomies throughout the body. Previously established NCE-MRA techniques suffered from long scan times or low sensitivity. Advances in hardware and software have made NCE-MRA scan times clinically feasible. Recent concerns over the safety of gadolinium-based contrast material combined with the expens...