Analysis of infusion pump error logs and their significance for health care.
نویسندگان
چکیده
Infusion therapy is one of the largest practised therapies in any healthcare organisation, and infusion pumps are used to deliver millions of infusions every year in the NHS. The aircraft industry downloads information from 'black boxes' to help design better systems and reduce risk; however, the same cannot be said about error logs and data logs from infusion pumps. This study downloaded and analysed approximately 360 000 hours of infusion pump error logs from 131 infusion pumps used for up to 2 years in one large acute hospital. Staff had to manage 260 129 alarms; this accounted for approximately 5% of total infusion time, costing about £1000 per pump per year. This paper describes many such insights, including numerous technical errors, propensity for certain alarms in clinical conditions, logistical issues and how infrastructure problems can lead to an increase in alarm conditions. Routine use of error log analysis, combined with appropriate management of pumps to help identify improved device design, use and application is recommended.
منابع مشابه
An Observational Study of Errors Related to the Preparation and Administration of Medications Given by Infusion Devices in a Teaching Hospital
Since there is no detailed hospital based incident reporting system, this study was designed to evaluate the medication errors associated with infusion pumps in intensive care unit (ICU). The investigation was conducted in a Teaching hospital in the form of a prospective, observational study. A sample size of 43 doses administered to ICU patients was chosen to enable reliable estimate of error ...
متن کاملAn Observational Study of Errors Related to the Preparation and Administration of Medications Given by Infusion Devices in a Teaching Hospital
Since there is no detailed hospital based incident reporting system, this study was designed to evaluate the medication errors associated with infusion pumps in intensive care unit (ICU). The investigation was conducted in a Teaching hospital in the form of a prospective, observational study. A sample size of 43 doses administered to ICU patients was chosen to enable reliable estimate of error ...
متن کاملDifficulty Using Smart Pump Logs to Recreate a Patient Safety Event: Case Study and Considerations for Pump Enhancements
The authors present a case in which a physical anomaly with an infusion pump resulted in an unforeseen fault that the nurse's attempts to resolve unknowingly exacerbated. This case study presents the first report in the literature to detail the difficulty in recreating a patient safety event using smart pump logs, support server continuous quality improvement (CQI) data, and the drug order entr...
متن کاملCollaboration Fuels Success of Infusion Management Interoperability Initiative.
Great River Medical Center (GRMC) is a 378-bed regional facility offering acute and intermediate care to residents of southeast Iowa, west-central Illinois, and northeast Missouri. The hospital, which admits more than 6,500 patients and logs more than 190,000 outpatient visits annually, offers comprehensive medical and extensive surgical services. In 2013, GRMC began a patient safety initiative...
متن کاملTime-dependent analysis of dosage delivery information for patient-controlled analgesia services
Pain relief always plays the essential part of perioperative care and an important role of medical quality improvement. Patient-controlled analgesia (PCA) is a method that allows a patient to self-administer small boluses of analgesic to relieve the subjective pain. PCA logs from the infusion pump consisted of a lot of text messages which record all events during the therapies. The dosage infor...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- British journal of nursing