The Accuracy of New and Aged Mechanical Torque Devices Employed in Five Dental Implant Systems

نویسندگان

  • Mehmet Ali Erdem
  • Burcin Karatasli
  • Onur Dinçer Kose
  • Taha Emre Kose
  • Erhan Çene
  • Serhan Aydın Aya
  • Abdulkadir Burak Cankaya
چکیده

Purpose Friction-style and spring-style torque wrenches are used to tighten implant abutments and prosthetic screws. The mechanical stability of these torque wrenches is crucial for the implant-abutment connection. The purposes of this study were to assess the performance of five brands (Straumann, Zimmer, Implant KA, Bredent, and Biohorizons) of wrench and to evaluate possible changes in applied torque values of aged wrenches. Materials and Methods Five new and aged wrenches that had been used approximately 250 times in a 1-year period were tested. The torque applied by friction- and spring-style wrenches was measured with a specially designed strain gauge indicator. Descriptive statistics, the one-sample t-test, and the independent-samples t-test were used to analyze values obtained from all torque wrenches. Results The accuracy of new and aged torque devices of all brands except Bredent differed significantly from the target values, but the mean values for aged and new wrenches did not differ significantly from each other (p > 0.05). Values for the spring- and friction-type torque wrenches deviated from the target values by 11.6% and 10.2%, respectively. Conclusion The accuracy of aged torque wrenches is adequate for prosthetic screw tightening, but that of new torque wrenches is unsatisfactory and must be examined carefully before delivery.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The accuracy of various torque wrenches used in dental implant systems

Introduction: The purpose of this study was to compare the accuracy of the torque wrenches used in different dental implant systems. Methods: We evaluated 42 torque wrenches used in different dental clinics in Mashhad, Iran, using a digital torque meter (Mark 10). High (25, 30 and 35 N·cm) and low (15 N·cm) levels of torque were examined. Ten tests were performed on each wrench, and the mean va...

متن کامل

The effect of steam sterilization on the accuracy of spring-style mechanical torque devices for dental implants

BACKGROUND Mechanical torque devices (MTDs) are one of the most commonly recommended devices used to deliver optimal torque to the screw of dental implants. Recently, high variability has been reported about the accuracy of spring-style mechanical torque devices (S-S MTDs). Joint stability and survival rate of fixed implant supported prosthesis depends on the accuracy of these devices. Currentl...

متن کامل

The Effect of Aging on the Accuracy of New Friction-Style Mechanical Torque Limiting Devices for Dental Implants

OBJECTIVE High variability in delivering the target torque is reported for friction-style mechanical torque limiting devices (F-S MTLDs). The effect of aging (number of use) on the accuracy of these devices is not clear. The purpose of this study was to assess the effect of aging on the accuracy (±10% of the target torque) of F-S MTLDs. MATERIALS AND METHODS Fifteen new F-S MTLDs and their ap...

متن کامل

Evaluation of the Accuracy and Related Factors of the Mechanical Torque-Limiting Device for Dental Implants

OBJECTIVE Accurate delivery of torque to implant screws is critical to generate ideal preload in the screw joint and to offer protection against screw loosening. Mechanical torque-limiting devices (MTLDs) are available for this reason. In this study, the accuracy of one type of friction-style and two types of spring-style MTLDs at baseline, following fatigue conditions and sterilization process...

متن کامل

بررسی تأثیر دفعات کارکرد بر دقت محدودکننده‌های مکانیکی تورک از نوع فنری در ایمپلنت‌های دندانی□

Objective: Recently, a wide variation has been reported in delivery of the target torque by spring-style mechanical torque limiting devices (MTLDs) but the effect of aging on their accuracy has not been independently evaluated. The purpose of this study was to determine the effect of aging on the accuracy of spring-style MTLDs. Methods: In this in-vitro experimental study, 15 new spring-style M...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 2017  شماره 

صفحات  -

تاریخ انتشار 2017