Intra-cavitary dosimetry for IMRT head and neck treatment using TLDs in a naso-oesophageal tube

نویسندگان

  • F M Gagliardi
  • K J Roxby
  • P E Engström
  • J C Crosbie
چکیده

Complex intensity-modulated radiation therapy (IMRT) treatment plans require rigorous quality assurance tests. The aim of this study was to independently verify the delivered dose inside the patient in the region of the treatment site. A flexible naso-gastric tube containing thermoluminescent dosimeters (TLDs) was inserted into the oesophagus via the sinus cavity before the patient’s first treatment. Lead markers were also inserted into the tube in order that the TLD positions could be accurately determined from the lateral and anterior– posterior electronic portal images taken prior to treatment. The measured dose was corrected for both daily linac output variations and the estimated dose received from the portal images. The predicted dose for each TLD was determined from the treatment planning system and compared to the measured TLD doses. The results comprise 431 TLD measurements on 43 patients. The mean measured-to-predicted dose ratio was 0.988 ± 0.011 (95% confidence interval) for measured doses above 0.2 Gy. There was a variation in this ratio when the measurements were separated into low dose (0.2–1.0 Gy), medium dose (1.0–1.8 Gy) and high dose (>1.8 Gy) measurements. The TLDloaded, naso-oesophageal tube for in vivo dose verification is straightforward to implement, and well tolerated by patients. It provides independent reassurance of the delivered dose for head and neck IMRT. (Some figures in this article are in colour only in the electronic version)

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

In vivo dose verification of IMRT treated head and neck cancer patients.

An independent in vivo dose verification procedure for IMRT treatments of head and neck cancers was developed. Results of 177 intracavitary TLD measurements from 10 patients are presented. The study includes data from 10 patients with cancer of the rhinopharynx or the thyroid treated with dynamic IMRT. Dose verification was performed by insertion of a flexible naso-oesophageal tube containing T...

متن کامل

Dose Distribution Analysis of Rapid Arc and Intensity Modulated Radiotherapy Plan in Head and Neck Cancer

Introduction: Globally, intensity-modulated radiation therapy (IMRT) is considered as highly precise and accurate method of radiotherapy planning. This technique amplifies spatial dose distribution conformity by modulating the intensity of radiation beams in each sub-volume. Additionally, it can reduces the dose to surrounding critical organs and deliver the planned dose to targets with the nom...

متن کامل

Assessment of a 2D EPID-based Dosimetry Algorithm for Pre-treatment and In-vivo Midplane Dose Verification

Introduction: The use of electronic portal imaging devices (EPIDs) is a method for the dosimetric verification of radiotherapy plans both pretreatment and in-vivo. The aim of this study was to test a 2D EPID-based dosimetry algorithm for dose verification of some plans inside a homogenous and anthropomorphic phantom and in-vivo, as well. Materials and Methods: </strong...

متن کامل

Design and Implementation of Head and Neck Perspex Phantom for Intensity Modulated Radiation Therapy

The aims of this study are to design, construct and evaluate the applicability of a head and neck Perspex phantom for dosimetric verification of intensity modulated radiation therapy (IMRT). Thermoluminescent dosimeters (TLDs) was used in this study. Consistency of the fabricated phantom readings was checked by multiple IMRT irradiations. The inter-fraction standard deviation among all TLDs’ me...

متن کامل

The effect of high energy photon beam in dose- volume parameters and integral dose of normal structures in Head and neck IMRT

Introduction: The aim of this study is to investigate the effect of partially used high photon energies on dosimetric parameters and integral doses of nasopharynx IMRT plans.   Materials and Methods: Two 7 coplanar IMRT plan were created for eleven patients (6 MV only IMRT plan in both phase1 and phase2 and 6 MV phase1-mixed energies (6 MV and 15 MV) phase2...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2009