Turn geometry for minimizing band broadening in microfabricated capillary electrophoresis channels.
نویسندگان
چکیده
Turns in microfabricated capillary electrophoresis channels generally result in degraded separation quality. To circumvent this limitation, channels were constructed with different types of turns to determine the design that minimizes turn-induced band broadening. In particular, tapered turns were created by narrowing the separation channel width before the start of a turn and widening the channel after the turn is complete. The radius of curvature of the turn, the length over which the channel is tapered, and the degree of tapering were explored. The column efficiencies were determined by examining the resolution of the 271/281 base pair doublet in the separation of a phiX174 HaeIII DNA sizing ladder. Tapered turns with the smallest radius of curvature (250 microm), the shortest tapering length between the separation and turn widths (55 microm), and the largest tapering ratio (4:1 separation channel width to turn channel width) produced the highest resolution separations. These results are discussed by comparison to theoretical predictions of the effect of tapers and turns on analyte band dispersion in capillary electrophoresis.
منابع مشابه
System-oriented dispersion models of general-shaped electrophoresis microchannels.
This paper presents a system-oriented model for analyzing the dispersion of electrophoretic transport of charged analyte molecules in a general-shaped microchannel, which is represented as a system of serially connected elemental channels of simple geometry. Parameterized analytical models that hold for analyte bands of virtually arbitrary initial shape are derived to describe analyte dispersio...
متن کاملFast separation of oligonucleotide and triplet repeat DNA on a microfabricated capillary electrophoresis device and capillary electrophoresis.
A laser-induced fluorescence detection system coupled with a highly sensitive silicon-intensified target (SIT) camera is successfully applied to the imaging of a band for DNA fragment labeling by fluorescence dye in a microchannel, and to the visualizing of the separation process on a microfabricated chip. We demonstrated that an only 6 mm separation channel is sufficient for the separation of ...
متن کاملMulti-layer Microfluidic Devices for Amino Acid Analysis: the Mars Organic Analyzer
Sensitive amino acid composition and chirality analysis is demonstrated using the Mars Organic Analyzer (MOA), a portable microfabricated capillary electrophoresis (CE) instrument. The MOA integrates all high voltage power supplie s, pneumatic controls, and fluorescence detection optics. The microfabricated device is a novel multi-layer structure combining glass separation channels [1] and micr...
متن کاملFully integrated PCR-capillary electrophoresis microsystem for DNA analysis.
A fully integrated genomic analysis microsystem including microfabricated heaters, temperature sensors, and PCR chambers directly connected to capillary electrophoretic separation channels has been constructed. Valves and hydrophobic vents provide controlled and sensorless sample positioning and immobilization into 200 nL PCR chambers. The use of microfabricated heating and temperature sensing ...
متن کاملMicrofabricated bioprocessor for integrated nanoliter-scale Sanger DNA sequencing.
An efficient, nanoliter-scale microfabricated bioprocessor integrating all three Sanger sequencing steps, thermal cycling, sample purification, and capillary electrophoresis, has been developed and evaluated. Hybrid glass-polydimethylsiloxane (PDMS) wafer-scale construction is used to combine 250-nl reactors, affinity-capture purification chambers, high-performance capillary electrophoresis cha...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Analytical chemistry
دوره 72 14 شماره
صفحات -