We have reviewed 70 Harris-Galante uncemented acetabular components implanted as hybrid hip replacements with cemented stems between 1991 and 1995 in 53 patients whose mean age was 40 years (19 to 49). The mean follow-up was for 13.6 years (12 to 16) with no loss to follow-up. We assessed the patients both clinically and radiologically. The mean Oxford hip score was 20 (12 to 46) and the mean Harris hip score 81 (37 to 100) at the final review. Radiologically, 27 hips (39%) had femoral osteolysis, 13 (19%) acetabular osteolysis, and 31 (44%) radiolucent lines around the acetabular component. Kaplan-Meier survival curves were constructed for the outcomes of revision of the acetabular component, revision of the component and polyethylene liner, and impending revision for progressive osteolysis. The cumulative survival for revision of the acetabular component was 94% (95% confidence interval 88.4 to 99.7), for the component and liner 84% (95% confidence interval 74.5 to 93.5) and for impending revision 55.3% (95% confidence interval 40.6 to 70) at 16 years. Uncemented acetabular components with polyethylene liners undergo silent lysis and merit regular long-term radiological review.
منابع مشابه
Cemented polyethylene and cementless porous-coated acetabular components have similar outcomes at a mean of seven years after total hip replacement: a prospective randomised study.
The aim of this prospective randomised study was to compare the clinical and radiological results of a cemented all-polyethylene Ultima acetabular component with those of a cementless porous-coated acetabular component (PFC) following total hip replacement (THR). A total of 287 patients received either a polyethylene acetabular component (group A) or a cobalt-chromium porous-coated co...
متن کاملOutcome of a hemispherical porous-coated acetabular component with a proximally hydroxyapatite-coated anatomical femoral component: a 12- to 15-year follow-up study.
We reviewed 111 hemispherical Duraloc series-500 acetabular components with a minimum follow-up of 12 years. The mean clinical and radiological follow-up was 13.4 years (12 to 15). A Profile hydroxyapatite-coated anatomical femoral component was used in each case. Six patients had a late dislocation, for whom the polyethylene liner was exchanged. Each acetabular component was well fixed and all...
متن کاملFixation, polyethylene wear, and pelvic osteolysis in primary total hip replacement.
A multicenter retrospective review was performed analyzing 1081 primary total hip replacements in 944 patients using the Harris Galante-I cementless acetabular component with screw fixation. All patients were followed up for a minimum of 5 years with a mean followup of 81 months. Linear polyethylene wear averaged 0.11 mm/year (range, 0-0.86 mm/year). Pelvic osteolysis was seen in 25 patients (2...
متن کاملRevision of the acetabulum with a contemporary cementless component.
We evaluated the performance of a contemporary cementless acetabular component at a minimum of 5 years postoperatively. One hundred eighty-seven consecutive acetabular component revisions were performed using a hemispherical porous-coated component. Patients were followed prospectively with radiographs and Harris hip scores. Twenty patients died, leaving 158 patients (166 hips) available for fo...
متن کاملHarris Galante cementless acetabular replacement in avascular necrosis.
We describe the results of 76 total arthroplasties of the hip for stage-III or stage-IV avascular necrosis of the femoral head. Harris Galante Porous cups were used in 63 patients between 1986 and 1994 and followed prospectively. We reviewed 70 hips with a follow-up of more than five years (mean 7.6). At the latest review the mean Harris Hip Score had improved from a preoperative value of 29 +/...