ECG/Navigator-Free 4D Whole-Heart Coronary MRA with Simultaneous Visualization of Cardiac Function and Anatomy
نویسندگان
چکیده
Purpose: Cardiac and respiratory motion artifacts are major challenges in cardiac MRI, and are conventionally suppressed by prospective ECG and diaphragm navigator gating, respectively. However, these approaches require time-consuming setup and prolonged acquisition times, and are susceptible to variations of heart rate or respiratory pattern. Further, they can become unreliable at higher field strengths [1]. In this work, we propose to address these limitations by using retrospective self-gating (SG) with continuous 3D radial acquisition. The proposed method offers cardiac phase resolved whole-heart visualization in high isotropic spatial resolution and flexible temporal resolution, providing functional and anatomical information simultaneously. Here we demonstrate and discuss our initial healthy volunteer results in cine and coronary applications. Methods: We used a spoiled gradient-echo sequence with Gd-BOPTA contrast-enhancement on a Siemens 3T scanner: TR/TE = 5.5/3.0ms, waterselective hard pulse, FOV=(400mm), matrix=384, flip angle=15 ̊, and a 3D radial trajectory with 2D golden means ordering [2]. Imaging was performed continuously without any prospective gating or prep-pulses. Superior-inferior SG projections were inserted every 15 imaging lines. A total of 100,000 projections were collected with a fixed free-breathing scan time of ~10 minutes. In the offline reconstruction, we firstly extract the cardiac and respiratory motion signals from the SG profiles by principal component analysis. We then find the cardiac triggers by valley detection, reject the arrhythmic heartbeats, and bin the data into 9 cardiac and 6 respiratory phases. We subsequently perform respiratory motion correction separately for each cardiac phase to combine all respiratory phases [3], and reconstruct the 4D image series using a self-calibrating CG-SENSE method [4]. Results and Discussions: Shown in Fig. 1, the SG profiles displayed intensity variations from both cardiac and respiratory motion. The extracted motion signal showed excellent correspondence with the ECG and respiratory bellow waveforms. Fig. 2 shows the two reconstruction modes. In the performancedriven mode, the data was re-binned into 16 cardiac phases without view sharing. Each phase contained ~6,000 projections with a temporal footprint of ~50ms. The isotropic spatial resolution enabled arbitrary 3D reformatting. In the quality-driven mode, the quiescent window, which may span more than one cardiac phase (2 in this particular case), was retrospectively determined from the 4D image series. The image quality can therefore be improved by including more data into the reconstruction. The location and size of this acquisition window was automatically adjusted on a beat-by-beat basis to account for moderate heart rate variations. In this subject, the retrospective triggering showed improved RCA visualization than that from the simulated prospective triggering with a fixed trigger delay. Conclusion: We have demonstrated the feasibility of a fully selfgated 4D whole-heart imaging technique with high isotropic spatial resolution and near 100% imaging efficiency through respiratory motion correction and retrospective cardiac self-gating. No ECG or diaphragm navigator is needed. The 4D visualization allows flexible reconstructions that prioritize either temporal resolution or image quality. Future efforts will be focused on optimizing the pulse sequence and reconstruction parameters, as well as comparing the dual-mode reconstruction with existing cardiac cine and coronary MRA protocols.
منابع مشابه
ECG and navigator-free 4D whole-heart coronary MRA
Background Cardiac and respiratory motion artifacts are major challenges to whole-heart coronary MRA. The conventional motion suppression strategies often involve prospective gating based on motion surrogates, e.g. ECG and navigator, which complicates scan setup and prolongs scan time significantly. To address these limitations, an ECG and navigator-free 4D whole-heart coronary MRA technique wa...
متن کاملDynamic self-navigated 3D whole-heart radial coronary MRA with retrospective acquisition window selection
Background Navigator gated coronary MRA is highly time inefficient as the data collection duty cycle (DC) is only 2% [1]. Here, we report a new, dynamic self-navigated coronary MRA technique with isotropic spatial resolution that a) further improves DC by continuously acquiring image data without ECG triggering and that b) enables retrospective, flexible, and individual selection of the acquisi...
متن کامل3T Coronary MRA using 3D Multi-interleaved Multi-echo Acquisition and VARPRO Fat-water Separation
Introduction: Coronary MR Angiography is a valuable tool for non-invasive assessment of coronary arteries. Presently, contrastenhanced, fat-saturated, ECG-triggered and navigator-gated 3D spoiled gradient-echo sequence is employed for whole-heart Coronary MRA at 3T. However, large static field variations at 3T frequently result in non-uniform fat-suppression over the field-of-view (FOV), obscur...
متن کامل3.0T whole-heart coronary magnetic resonance angiography performed with 32-channel cardiac coils: a single-center experience.
BACKGROUND Whole-heart coronary magnetic resonance angiography (MRA) is a promising method for noninvasive, radiation-free detection and exclusion of obstructive coronary artery disease; however, the required imaging time and robustness of the technique are not yet satisfactory. We evaluated the value of whole-heart coronary MRA at 3.0T using a 32-channel cardiac coil, which reduces image-acqui...
متن کاملAssessment of coronary artery disease using 3.0T magnetic resonance coronary angiography: a national multicenter trial
Background 3.0T contrast enhanced whole-heart coronary magnetic resonance angiography (MRA) is a promising method for noninvasive, radiation-free detection and exclusion of obstructive coronary artery disease (CAD); however, the accuracy of this approach has not been determined in a multicenter trial. Methods An ECG-triggered, navigator-gated, inversion-recovery prepared, segmented gradient-ech...