Modified Jendrassik--Grof method for bilirubins adapted to the Abbott Bichromatic Analyzer.
نویسنده
چکیده
I describe a semi-micro method for use with a discrete analyzer, the Abbott ABA-100, requiring 75 microliter of sample for determination of both the total and direct bilirubins. This method utilizes a serum blank. It is simple and reasonably rapid. The azobilirubin formed by both the total and the direct bilirubins produces a reddish-violet color, which is measured at 550 nm. Values obtained are comparable to those by the automated Jendrassik and Grof procedure adapted by Gambino and Schreiber (total r = 0.999, direct r = 0.990). The method is linear to 300 mg/liter. Day-to-day precision (CV) for the total bilirubin was 8.8% for a 8 mg/liter sample (n = 45) and 4.7% for a 57 mg/liter sample (n = 46). The absorbance of hemoglobin and its derivatives (540 nm) caused a false decrease in the measured bilirubin content, the direct being more affected than the total. This problem was alleviated by the use of caffeine and acid blanks.
منابع مشابه
Automated Jendrassik-Grof method for measurement of bilirubin in serum with the Greiner Selective Analyzer (GSA II D), and comparison with the method involving diazotized 2,4-dichloroaniline.
We automated the Jendrassik-Grof method for measurement of bilirubin with the Greiner Selective Analyzer GSA II D. Comparison with the manual Jendrassik-Grof method showed a highly significant correlation coefficient of 0.994 for total and 0.980 for "direct" bilirubin measurement. The automated method showed good precision and was linear to 667 mu mol/L, the highest concentration tested. Additi...
متن کاملIndican interference with six commercial procedures for measuring total bilirubin.
We have studied the effect of indican on six commercial procedures for the measurement of total bilirubin in serum. Total bilirubin measured by the Bilirubin A-Gent (Abbott) 2,4-dichlorophenyl diazonium procedure increased by 50 mg/L for each 1 mmol/L of added indican. Similarly, total bilirubin measured by the Bilirubin C-System (Boehringer Mannheim) 2,5-dichlorophenyl diazonium procedure incr...
متن کاملCentrifugal analyzer method for total bilirubin in serum by use of diazotized 2-chloroaniline-5-sulfonic acid.
We describe two centrifugal analyzer methods for measuring total bilirubin in serum. Diazotized 2-chloroaniline-5-sulfonic acid is used in both. In the first procedure, ethylene glycol and methanol are used as an accelerator solvent; results correlate well with those by a Jendrassik and Grof method adapted to the centrifugal analyzer, but there is considerable hemolysis interference. In the sec...
متن کاملAdaptation of blank-corrected methods for measurement of total and conjugated bilirubin and uric acid to the centrifugal analyzer.
The blank-corrected determination in serum of total and conjugated bilirubin and of uric acid has been adapted to the centrifugal analyzer. Bilirubin is measured by a modified Jendrassik and Grof procedure; uric acid is determined kinetically by using a reaction based on reduction of ferric ion, including a uricase-treated blank. Sensitivity and precision of both methods are excellent, and resu...
متن کاملMicro-scale enzymic determination of ethanol in plasma with a discrete analyzer, the ABA-100.
An enzymic method for estimating ethanol has been adapted to the Abbott Bichromatic Analyzer-100. Fully automated, the method can accommodate up to 192 assays per hour. It requires only 2.5 muL of plasma and therefore is applicable to analysis for ethanol in blood sampled by finger-prick.