Modular titanium alloy neck adapter failures in hip replacement - failure mode analysis and influence of implant material
نویسندگان
چکیده
BACKGROUND Modular neck adapters for hip arthroplasty stems allow the surgeon to modify CCD angle, offset and femoral anteversion intraoperatively. Fretting or crevice corrosion may lead to failure of such a modular device due to high loads or surface contamination inside the modular coupling. Unfortunately we have experienced such a failure of implants and now report our clinical experience with the failures in order to advance orthopaedic material research and joint replacement surgery.The failed neck adapters were implanted between August 2004 and November 2006 a total of about 5000 devices. After this period, the titanium neck adapters were replaced by adapters out of cobalt-chromium. Until the end of 2008 in total 1.4% (n = 68) of the implanted titanium alloy neck adapters failed with an average time of 2.0 years (0.7 to 4.0 years) postoperatively. All, but one, patients were male, their average age being 57.4 years (36 to 75 years) and the average weight 102.3 kg (75 to 130 kg). The failures of neck adapters were divided into 66% with small CCD of 130 degrees and 60% with head lengths of L or larger. Assuming an average time to failure of 2.8 years, the cumulative failure rate was calculated with 2.4%. METHODS A series of adapter failures of titanium alloy modular neck adapters in combination with a titanium alloy modular short hip stem was investigated. For patients having received this particular implant combination risk factors were identified which were associated with the occurRence of implant failure. A Kaplan-Meier survival-failure-analysis was conducted. The retrieved implants were analysed using microscopic and chemical methods. Modes of failure were simulated in biomechanical tests. Comparative tests included modular neck adapters made of titanium alloy and cobalt chrome alloy material. RESULTS Retrieval examinations and biomechanical simulation revealed that primary micromotions initiated fretting within the modular tapered neck connection. A continuous abrasion and repassivation process with a subsequent cold welding at the titanium alloy modular interface. Surface layers of 10 - 30 microm titanium oxide were observed. Surface cracks caused by fretting or fretting corrosion finally lead to fatigue fracture of the titanium alloy modular neck adapters. Neck adapters made of cobalt chrome alloy show significantly reduced micromotions especially in case of contaminated cone connection. With a cobalt-chromium neck the micromotions can be reduced by a factor of 3 compared to the titanium neck. The incidence of fretting corrosion was also substantially lower with the cobalt-chromium neck configuration. CONCLUSIONS Failure of modular titanium alloy neck adapters can be initiated by surface micromotions due to surface contamination or highly loaded implant components. In the present study, the patients at risk were men with an average weight over 100 kg. Modular cobalt chrome neck adapters provide higher safety compared to titanium alloy material.
منابع مشابه
Comment on: Modular titanium alloy neck failure in total hip replacement
To the Editor of the SICOT-J Orthopaedic Journal, I am writing to you as I would like to kindly express some considerations on the case-report of Dr. Marco Ceretti on ‘‘Modular titanium alloy neck failure in total hip replacement’’ published in your on-line journal last April (M. Ceretti and F. Falez SICOT J 2016, 2, 20). The Authors report on ‘‘a case of a repetitive modular femoral neck prost...
متن کاملModular titanium alloy neck failure in total hip replacement: analysis of a relapse case.
Modular neck hip prosthesis born in the 1990 with the aim of allowing the surgeon to modify CCD angle, offset and femoral anteversion intra-operatively restoring patient's original biomechanics. In order to achieve the best biomechanics of the reconstructed hip, preoperative planning is essential. In the last few years modularity has been questioned and an argument made for the return to mono b...
متن کاملPrimary hip replacement stem taper fracture due to corrosion in 3 patients
© 2016 The Author(s). Published by Taylor & Francis on behalf of the Nordic Orthopedic Federation. This is an Open Access article distributed under the terms of the Creative Commons Attribution-Non-Commercial License (https://creativecommons.org/licenses/by-nc/3.0) DOI 10.3109/17453674.2015.1128780 (McGrory et al. 2015, Bishop et al. 2013). No mechanical implant failures (i.e. fractures) have b...
متن کاملCorrosion and adverse local tissue reaction after total hip arthroplasty with a modular titanium alloy femoral neck
This report describes a case of mechanically assisted crevice corrosion and secondary adverse local tissue reaction in a patient following a total hip arthroplasty, utilizing a modular neck (bi-modular) femoral component. Radiographic evaluation demonstrated a well-positioned, stable, cementless arthroplasty. Upon further evaluation, the patient had elevated serum cobalt and chromium levels, an...
متن کاملTrunnion Failure of the Recalled Low Friction Ion Treatment Cobalt Chromium Alloy Femoral Head.
BACKGROUND Gross trunnion failure (GTF) is a rare complication in total hip arthroplasty (THA) reported across a range of manufacturers. Specific lots of the Stryker low friction ion treatment (LFIT) anatomic cobalt chromium alloy (CoCr) V40 femoral head were recalled in August 2016. In part, the recall was based out of concerns for disassociation of the femoral head from the stem and GTF. ME...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 11 شماره
صفحات -
تاریخ انتشار 2010