Evaluation of six oxygen concentrators.

نویسندگان

  • D P Johns
  • P D Rochford
  • J A Streeton
چکیده

Examples of six oxygen concentrators (DeVO2, Dom 10, Econo 2, Hudson, Permox, and Roomate) were evaluated over a 9-28 day period to determine (1) the oxygen yield (% O2) over the flow range 1-4 l min-1; (2) 90% oxygen rise time (90% RT) from a cold start when they were operated at 2 l min-1; (3) accuracy and readability of the flow device; (4) static outlet pressure; (5) major components comprising the product gas (Hudson only); and (6) general characteristics. At an outlet flow of 2 l min-1 the mean % O2 generated by all models, except the Permox (which was lower, mean (SD) 90.5% (3.1%), were between 94% and 95% with a range of less than +/- 0.5%. The Dom 10, Econo 2, and Hudson consistently generated higher oxygen concentrations than the other models at flow rates greater than 2 l min-1. The 90% RT was less than 10.5 minutes for all models. Deviations of up to 22% were observed between predicted and measured flow rates in all models except the DeVO2, Hudson, and Permox. It was possible to set the orifice type flow devices fitted to the Permox and Roomate between indicated flow settings, resulting in cessation of flow. Spectral analysis of the output of one device showed that argon and oxygen were concentrated to similar extents, indicating that the maximal attainable oxygen yield for a molecular sieve concentrator is about 96%.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

An evaluation of the use of concentrators for domiciliary oxygen supply for less than 8 h day-1.

Since their introduction in 1985, oxygen concentrators have only been recommended when domiciliary oxygen is used for over 8 h day-1. Subsequent changes in the prices of oxygen merit a reappraisal of the prescribing of concentrators and cylinders when oxygen is used for less than 8 h day-1. Twenty-six patients in two health districts who used oxygen for less than 8 h day-1 completed a crossover...

متن کامل

Meeting oxygen needs in Africa: an options analysis from the Gambia.

OBJECTIVE To compare oxygen supply options for health facilities in the Gambia and develop a decision-making algorithm for choosing oxygen delivery systems in Africa and the rest of the developing world. METHODS Oxygen cylinders and concentrators were compared in terms of functionality and cost. Interviews with key informants using locally developed and adapted WHO instruments, operational as...

متن کامل

Insufficient oxygen concentration obtained at domiciliary controls of eighteen concentrators.

The oxygen concentration (O2%) produced by 12 type A concentrators, with a working time of 28-18,099 h, and 6 type B concentrators, with a working time of 0-3,033 h, was measured over a 12 month period in the user's home, at a flow rate of 2 l.min-1. One hundred and two measurements of O2% (mean 82.9), made at least once monthly by a visiting nurse, showed that type A concentrators were usually...

متن کامل

Oxygen concentrators performance with nitrous oxide at 50:50 volume.

BACKGROUND AND OBJECTIVES Few investigations have addressed the safety of oxygen from concentrators for use in anesthesia in association with nitrous oxide. This study evaluated the percent of oxygen from a concentrator in association with nitrous oxide in a semi-closed rebreathing circuit. METHODS Adult patients undergoing low risk surgery were randomly allocated into two groups, receiving a...

متن کامل

Technical and clinical assessment of oxygen concentrators.

One membrane oxygen enricher (Oxygen Enrichment Company OE-4E) and four molecular sieve (MS) concentrators (Mountain Medical Econo2, De Vilbiss MINI DeVO2, Cryogenic Roomate III, and Mountain Medical Mini O2) have been studied to assess technical and clinical performance. During four weeks of continuous operation at a flow rate of 2 l min-1 (6 l min-1 for OE-4E) consistent oxygen levels were de...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Thorax

دوره 40 11  شماره 

صفحات  -