A Strategy of Underexpansion and Ad Hoc Post-Dilation of Balloon-Expandable Transcatheter Aortic Valves in Patients at Risk of Annular Injury: Favorable Mid-Term Outcomes.

نویسندگان

  • John S Tan
  • Jonathon Leipsic
  • Gidon Perlman
  • Dion Stub
  • Danny Dvir
  • Nicolaj C Hansson
  • Bjarne L Norgaard
  • Philipp Blanke
  • Anson Cheung
  • Jian Ye
  • Christopher R Thompson
  • Robert R Moss
  • Sandra Lauck
  • David Wood
  • John Webb
چکیده

OBJECTIVES The aim of this study was to evaluate a strategy of intentional underexpansion of excessively oversized balloon-expandable transcatheter heart valves (THVs) in terms of clinical outcomes, valve function, and frame durability at 1 year. BACKGROUND Transcatheter aortic valve replacement requires the selection of an optimally sized THV to ensure paravalvular sealing and fixation without risking annular injury. However, some patients have "borderline" annular dimensions that require choosing between a THV that may be too small or another that may be too large. METHODS We evaluated 47 patients at risk of annular injury who underwent transcatheter aortic valve replacement (TAVR) with an oversized, but deliberately underexpanded, THV followed by post-dilation if required. Clinical evaluation, echocardiography, and cardiac computed tomography were performed pre-TAVR, post-TAVR, and at 1 year. RESULTS Deployment of oversized THVs with modest underfilling of the deployment balloon (<10% by volume) was not associated with significant annular injury. Paravalvular regurgitation was mild or less in 95.7% of patients, with post-dilation required in 10.7%. THV hemodynamic function was excellent and remained stable at 1 year. Computed tomography documented stent frame circularity in 87.5%. Underexpansion was greatest within the intra-annular THV inflow (stent frame area 85.8% of nominal). There was no evidence of stent frame recoil, deformation, or fracture at 1 year. CONCLUSIONS In carefully selected patients with borderline annulus dimensions and in whom excessive oversizing of a balloon-expandable SAPIEN XT valve (Edwards Lifesciences, Inc., Irvine, California) is a concern, a strategy of deliberate underexpansion, with ad hoc post-dilation, if necessary, may reduce the risk of annular injury without compromising valve performance.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

1-Year Outcomes After Transcatheter Aortic Valve Replacement With Balloon-Expandable Versus Self-Expandable Valves: Results From the CHOICE Randomized Clinical Trial.

BACKGROUND The use of a balloon-expandable transcatheter heart valve previously resulted in a greater rate of device success compared with a self-expandable transcatheter heart valve. OBJECTIVES The aim of this study was to evaluate clinical and echocardiographic outcome data at longer term follow-up. METHODS The investigator-initiated trial randomized 241 high-risk patients with symptomati...

متن کامل

Structural integrity of balloon-expandable stents after transcatheter aortic valve replacement: assessment by multidetector computed tomography.

OBJECTIVES This study sought to evaluate the structural integrity of balloon-expandable stents used in transcatheter aortic valve replacement. BACKGROUND Underexpansion, deformation, or fracture of stent frames may affect transcatheter heart valve (THV) function and durability. METHODS Patients >1 year after transcatheter aortic valve replacement underwent multidetector computed tomography....

متن کامل

Evaluation and management of paravalvular aortic regurgitation after transcatheter aortic valve replacement.

Paravalvular aortic regurgitation (PAR) negatively affects the prognosis after transcatheter aortic valve replacement (TAVR) with dramatically increased morbidity and mortality in patients with more than mild PAR. Because transcatheter heart valves are implanted in a sutureless fashion using oversizing to anchor the prosthesis stent frame at the level of the virtual aortic annulus, stent frame ...

متن کامل

Impact of balloon post-dilation on clinical outcomes after transcatheter aortic valve replacement with the self-expanding CoreValve prosthesis.

OBJECTIVES The aim of this study was to assess the incidence and clinical impact of balloon post-dilation (BPD) after transcatheter aortic valve replacement (TAVR) with the CoreValve prosthesis (Medtronic Inc., Minneapolis, Minnesota). BACKGROUND BPD is a widely adopted strategy to reduce the degree of paraprosthetic regurgitation in case of transcatheter heart valve underexpansion. However, ...

متن کامل

Balloon expandable transcatheter aortic valve implantation with or without pre-dilation of the aortic valve – rationale and design of a multicenter registry (EASE-IT)

BACKGROUND In patients with severe calcific aortic stenosis, balloon aortic valvuloplasty (BAV) is routinely performed in order to pre-dilate the stenosed aortic valve prior to transcatheter aortic valve implantation (TAVI). Although pre-dilation is considered to be essential for the preparation of the valve landing zone, there is no clear evidence to support its clinical value. In contrast, BA...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • JACC. Cardiovascular interventions

دوره 8 13  شماره 

صفحات  -

تاریخ انتشار 2015