Middle-term results of threaded acetabular cups. High failure rates five years after surgery.
نویسندگان
چکیده
We report our results using three different threaded acetabular components (Mecring A, Mecring B and Weill) in 715 hips with a follow-up of between one and ten years (median: 99.1, 56.5, 38.3 months, respectively). All cups were implanted with one type of cementless stem. The clinical results were good or acceptable in about 70% of the hips, but signs of loosening with radiolucency and/or migration were found in 10.1%. Radiological evidence of loosening did not correlate significantly with the clinical outcome. Pain was not a reliable indicator of loosening and its absence sometimes allowed severe osteolysis to develop. Twenty-five hips were revised (3.5%) for aseptic loosening of the acetabular component. Kaplan-Meier estimates of the cumulative rate of failure showed a rapid increase five years after the initial operation, but no significant correlation with gender, age or weight. The high rate of failure indicates that further use of these acetabular components cannot be recommended. Annual radiographs are required to assess osteolysis even if the patients are free from pain.
منابع مشابه
The third generation of cementless cups consist of a metallic acetabular shell with a rough surface to improve osseous integration and therefore long- term survival
Prospectively the outcome after total hip replacement with a new threaded acetabular cup design was compared to an established press-fit cup. After 1, 2 and 5 years, the 36-item Short Form Health Survey, Western Ontario and McMaster University Osteoarthritis Index and Harris Hip Score revealed no significant differences between the two groups (each group : n = 42 patients), except for a higher ...
متن کاملCementless acetabular components.
Four hundred and fifteen patients with cementless acetabular components of either a smooth threaded (130) or porous surfaced (285) variety were compared for clinical symptoms and radiographic signs of component loosening. At a mean 4.8 year follow-up none of the patients with porous acetabular components had signs of component instability. At a mean 3.9 year follow-up 27 (21%) of the patients w...
متن کاملPinnacle revision cup in acetabular surgery results at 2 to 5 years.
Uncemented revision cups are widely used in revision hip arthroplasty; they have shown good results. We report the the short term outcome with the cementless Pinnacle revision cup. All acetabular revisions using a Pinnacle revision cup between January 2007 and March 2010 were included. In March 2012, clinical scores were determined and the latest radiographs were assessed. Revision and radiogra...
متن کاملPolyethylene wear in uncemented acetabular components.
We measured polyethylene wear in 231 porous-coated uncemented acetabular cups. We divided the hips into two groups according to the fixation of the femoral component, by cementing (n = 97) or press-fit (n = 134). Follow-up was from three to five years. The patients in two sub-groups were matched for weight, diagnosis, sex, age and length of follow-up. The linear wear rate of cups articulated wi...
متن کامل[Mid-term results of cementless ultima cups in primary total hip replacement].
PURPOSE OF THE STUDY To present the results of primary total hip arthroplasty (THA) with use of the hemispheric threaded Ultima cup at 8.5-year follow-up. MATERIAL Between 1996 and 1999, a total of 40 Ultima acetabular components were used in 33 patients (15 men and 18 women). The average age at the time of surgery was 61.2 years (range, 45 to 71). By the end of 2006, 36 cups were assessed. I...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- The Journal of bone and joint surgery. British volume
دوره 79 3 شماره
صفحات -
تاریخ انتشار 1997