Low dose of continuous – wave microwave irradiation did not cause temperature increase in muscles tissue adjacent to titanium alloy implants – an animal study

نویسندگان

  • Dongmei Ye
  • Yiming Xu
  • Tengfei Fu
  • Han Zhang
  • Xianxuan Feng
  • Gang Wang
  • Lan Jiang
  • Yuehong Bai
چکیده

BACKGROUND Research studies on the influence of radiofrequency electromagnetic radiation on implants in vitro have failed to investigate temperature changes in the tissues adjacent to the implants under microwave therapy. We therefore, used a rabbit model in an effort to determine the impact of microwave therapy on temperature changes in tissues adjacent to the titanium alloy implants and the safety profile thereof. METHODS Titanium alloy internal fixation plates were implanted in New Zealand rabbits in the middle of femur. Microwave therapy was performed by a 2450 MHz microwave generator 3 days after the surgery. Temperature changes of muscles adjacent to the implants were recorded under exposure to dose-gradient microwave radiation from 20w to 60w. RESULTS Significant difference between control and microwave treatment group at peak temperatures (T(peak)) and temperature gap (T(gap) = T(peak)-T(vally)) were observed in deep muscles (T(peak), 41.63 ± 0.21°C vs. 44.40 ± 0.17°C, P < 0.01; T(gap), 5.33 ± 0.21°C vs. 8.10 ± 0.36°C, P < 0.01) and superficial muscles (T(peak), 41.53 ± 0.15°C vs. 42.03 ± 0.23°C, P = 0.04; T(gap), 5.23 ± 0.21°C vs. 5.80 ± 0.17°C, P = 0.013) under 60 w, and deep muscles (T(peak), 40.93 ± 0.25°C vs. 41.87 ± 0.23°C, P = 0.01; T(gap), 4.73 ± 0.20°C vs. 5.63 ± 0.35°C, P = 0.037) under 50w, but not under 20, 30 and 40w. CONCLUSION Our results suggest that low-dose (20w-40w) continuous-wave microwave irradiation delivered by a 2450 MHz microwave generator might be a promising treatment for patients with titanium alloy internal fixation, as it did not raise temperature in muscle tissues adjacent to the titanium alloy implant.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Effects of Low-Dose Microwave on Healing of Fractures with Titanium Alloy Internal Fixation: An Experimental Study in a Rabbit Model

BACKGROUND Microwave is a method for improving fracture repair. However, one of the contraindications for microwave treatment listed in the literature is surgically implanted metal plates in the treatment field. The reason is that the reflection of electromagnetic waves and the eddy current stimulated by microwave would increase the temperature of magnetic implants and cause heat damage in tiss...

متن کامل

Enhancement of Apoptosis by Titanium Alloy Internal Fixations during Microwave Treatments for Fractures: An Animal Study

OBJECTIVE Microwaves are used in one method of physical therapy and can increase muscle tissue temperature which is useful for improving muscle, tendon and bone injuries. In the study, we sought to determine whether titanium alloy internal fixations influence apoptosis in tissues subjected to microwave treatments at 2,450 MHz and 40 W during the healing of fractures because this issue is not ye...

متن کامل

Investigating the Impact of Knee Prosthesis in Patients’ Body on Radiation Dose Distribution: A Monte Carlo Approach

Introduction: Metal prostheses in patients affect the radiotherapy dose distribution. Metal prostheses with high density and atomic number cause major changes in scattering and attenuation of radiation. The present study aims to assess the impact of metal knee prosthesis with various dimensions and materials on radiotherapy dose distribution. Material and Methods: In this research, the Varian L...

متن کامل

An investigation on heat transfer to the implant-bone interface due to abutment preparation with high-speed cutting instruments.

Excessive heat generation at the implant-bone interface may cause irreversible bone damage and loss of osseointegration. The effect of heat generation in vitro at the implant surface caused by abutment reduction with medium- and extra-fine-grain diamond and tungsten burs in a high-speed dental turbine was examined. Titanium-alloy abutments connected to a titanium-alloy cylindrical implant embed...

متن کامل

Machinability evaluation of Titanium alloy in Laser Assisted Turning

The use of titanium and its alloys has increased in various industries recently, because of their superior properties of these alloys. Titanium alloys are generally classified as difficult to machine materials because of their thermo-mechanical properties such as high strength-to-weight ratio and low thermal conductivity. Laser Assisted Machining (LAM) improves the machinability of high strengt...