Analytic performance of the iQ200 automated urine microscopy analyzer and comparison with manual counts using Fuchs-Rosenthal cell chambers.
نویسندگان
چکیده
Automated instruments that can examine urine for cells and particles have reduced the need for labor-intensive manual microscopy. We evaluated the performance of the Iris iQ200 Automated Urine Microscopy Analyzer (Iris Diagnostics, Chatsworth, CA) and compared results with manual cell and particle counts using Fuchs-Rosenthal counting chambers. Within-run imprecision (coefficient of variation) of the iQ200 for urine samples ranged from 3.0% to 45% for RBC counts between 1,029 and 3 x 10(6) cells/L, 3.4% to 40% for WBC counts between 1,006 and 4 x 10(6) cells/L, and 8.9% to 35% for epithelial cell counts between 93 and 4 x 10(6) cells/L. Between-run imprecision was 3.3% at 1,017 x 10(6) cells/L and 19.2% at 28 x 10(6) cells/L. There was good agreement between the iQ200 and manual cell counts (r > or = 0.94); however; the iQ200 produced lower results based on slopes of 0.92 (RBC count), 0.81 (WBC count), and 0.94 (epithelial cell counts). The iQ200 has satisfactory performance and correlates well with manual cell counts. Most urine samples containing RBCs, WBCs, and epithelial cells can be reported without review of captured images.
منابع مشابه
A multicenter evaluation of the Iris iQ200 automated urine microscopy analyzer body fluids module and comparison with hemacytometer cell counts.
Manual hemacytometer cell counting of body fluids is labor-intensive and requires skilled testing personnel. We performed a multicenter evaluation of the Iris iQ200 automated microscopy analyzer Body Fluids Module (Iris Diagnostics, Chatsworth, CA) and compared 350 iQ200 body fluid cell counts with manual hemacytometer cell counts. Within-run imprecision, expressed as coefficient of variation (...
متن کاملComparison of Cobas 6500 and Iris IQ200 fully-automated urine analyzers to manual urine microscopy
INTRODUCTION Urine screening is achieved by either automated or manual microscopic analysis. The aim of the study was to compare Cobas 6500 and Iris IQ200 urine analyzers, and manual urine microscopic analysis. MATERIALS AND METHODS A total of 540 urine samples sent to the laboratory for chemical and sediment analysis were analyzed on Cobas 6500 and Iris IQ200 within 1 hour from sampling. One...
متن کاملAutomated urinalysis: first experiences and a comparison between the Iris iQ200 urine microscopy system, the Sysmex UF-100 flow cytometer and manual microscopic particle counting.
BACKGROUND Automated analysis of insoluble urine components can reduce the workload of conventional microscopic examination of urine sediment and is possibly helpful for standardization. We compared the diagnostic performance of two automated urine sediment analyzers and combined dipstick/automated urine analysis with that of the traditional dipstick/microscopy algorithm. METHODS A total of 3...
متن کاملThe clinical reliability of automated cerebrospinal fluid cell counts on the Beckman-Coulter LH750 and Iris iQ200.
Although automation has improved the accuracy and precision of blood cell counts and is more rapid and less labor-intensive, cerebrospinal fluid (CSF) samples are still counted manually. We compared the IRIS iQ200 Body Fluids Module (Iris Diagnostics, Chatsworth, CA) and the Beckman-Coulter LH750 (Beckman-Coulter, Brea, CA) with manual counts and evaluated the impact of automation on the labora...
متن کاملLeukocyte counts in cerebrospinal fluid with the automated hematology analyzer CellDyn 3500 and the urine flow cytometer UF-100.
BACKGROUND The counting of leukocytes and erythrocytes in cerebrospinal fluid (CSF) is still performed microscopically, e.g., using a chamber in most laboratories. This requires sufficient practical experience, is time-consuming, and may constitute a problem in emergency diagnostics. Specific automated systems for CSF cell counting are not available at present. METHODS We tested the hematolog...