Characterizations of a CERN NbTi Reference Wire at LBNL∗
نویسندگان
چکیده
We report on the measurement of two NbTi reference samples, obtained from CERN through BNL, wire number 02R00056A01UX.265. The purpose of these characterizations is two-fold: 1) To establish an inter-laboratory comparison of short sample tests – and analysis – between LBNL, BNL and Fermilab within the LARP collaboration. Due to differences in reaction and mounting procedures and their resulting strain differences, Nb3Sn is not suited for this purpose. 2) To determine the accuracy of LBNL’s system with regard to magnetic field, temperature and current after a major system overhaul. The results show that at μ0H = 8 T, the critical current result falls within the standard deviation of the CERN database, whereas at μ0H = 5 T, it falls outside the CERN standard deviation, but within the range of critical current values that are reported by CERN for this wire. It is concluded that the LBNL results fall within the characterization variance as reported in the CERN database. 1 System changes LBNL’s short sample test facility has been substantially modified. Improvements involved the commissioning of a new system control and data acquisition system, a new magnet power supply, a new nano-volt meter and improved wiring. The data acquisition software was developed by Dr. B. ten Haken at the University of Twente, the Netherlands, and has gone through more than 15 years of development. Efforts were made to improve system performance with respect to accuracy, noise, ease of operation, consistent data storage and measurement. Voltage current characterizations and data reduction are now almost fully automated, yielding consistency, ease of operation and high sample throughput. 2 Voltage-current results Measurements of electric field as a function of current are best represented on a double logarithmic scale since physics, through the Anderson-Kim flux creep model [1, 2], dictates linearity on a double logarithmic scale [3]: E = Ec ( J Jc (H,T ) )n(H,T ) . (1) Measurement errors are effectively amplified on a double logarithmic plot through non-linearities that can arise from offset drift, sample heating, and current re-distribution and entrance effects (i.e. current sharing with the matrix in general). One can argue whether some non-linearities should be seen as a property of a wire, as opposed to probing solely the properties of the superconducting volume inside the wire. Usually, however, non-linearities in a double logarithmic plot can be traced back to a faulty measurement [3]. Two NbTi samples were characterized. Sample a was used to analyze and test the system, and optimize performance with respect to noise. Measurements are performed in DC mode, to prevent inductive voltage pickup and AC artifacts, creating a better defined characterization. The measured electric field as a function of sample current for both samples is depicted on a double logarithmic scale in Figure 1. Two voltage tap pairs were used over a 5 turn and a 7 turn section (i.e. about 0.5 m and 0.7 m sample length respectively). The voltages from both pairs are plotted in ∗Supported by the Director, Office of Science, High Energy Physics, U.S. Department of Energy under contract No. DE-AC02-05CH11231 †Electronic address: [email protected] 2 Godeke et al., report number LBNL-62392 Figure 1: Electric field as function of current for sample a (left) and sample b (right). Table 1: Critical current results CERN Sample a: 5 turn / 7 turn Sample b: 5 turn / 7 turn T = 4.222 K T = not recorded T = 4.233 K 4.222 K μ0H Ic n Ic n Ic Ic n Ic Ic 10−14 Ωm 10−5 V/m 10−14 Ωm 10−5 V/m 10−14 Ωm 10−14 Ωm [T] [A] [A] [A] [A] [A] [A] 5 538.8 52 535/537 47/47 535/537 532/533 49/48 532/533 534/535 6 426.2 48 425/427 44/44 423/424 423/423 46/45 421/421 423/423 7 313.6 43 315/315 41/41 311/311 312/313 8 201.7 35 210/211 30/30 203/204 208/208 33/33 202/202 203/204 Figure 1. LBNL’s data acquisition software uses a selectable electric field criterion since this is less ambiguous than a resistive criterion since the latter involves a choice with respect to the area to be used (e.g. total or non-Cu). A usual criterion for measurements on a barrel is 10−5 V/m. Data are collected over one electric field decade above and below this criterion, thereby preventing the collection of irrelevant data far below the transition and thus saving time. This range was extended down for the transition at μ0H = 6 T for sample a, demonstrating a noise level of E ∼= 10−7 V/m, or about 50 nV. 3 Critical current results Critical current is determined by fitting the measured electric field as function of current data to relation (1) (replacing J by I) over a selectable electric field range and calculating the intersect with the electric field criterion. Here, E(I) data that fall within a window Ec/3 < Ec < 3Ec were fitted to (1) to calculate Ic at Ec = 10−5 V/m. The resulting critical currents are summarized in Table 1, together with the reference results from CERN [4]. The E(I) data are linear on a double logarithmic plot (Figure 1) and thus follow (1). The critical current results, derived using an electric field criterion, can thus be translated to a resistive criterion through a combination of (1) with Ohm’s law, yielding: Iρ = ( ρIn c EcA ) 1 n−1 , (2) Godeke et al., report number LBNL-62392 3 4 5 6 7 8 9 0 100 200 300 400 500 600 CERN reference data LBNL results T = 4.222 K Cr itic al cu rre nt [A ] Applied magnetic field [T] Figure 2: Critical current as a function of magnetic field at T = 4.222 K for the CERN reference data and LBNL results for the 5 turn section of sample b. where Iρ is the critical current at a resistive criterion ρ . These calculated critical currents at a criterion ρ = 10−14 Ωm, using a wire diameter of 0.825 mm, are also given in Table 1. 4 Temperature corrections The Helium bath temperature during measurement of sample a was not recorded. LBNL’s system is equipped with vapor cooled current leads. The Helium gas flow through these leads is typically adjusted to run at an overpressure of 0.15 to 0.2 psi. For now, the bath temperature is influenced by variations in atmospheric pressure. Improvements are planned to register the Helium bath temperature during all characterization runs. For sample b, the bath temperature was recorded using two calibrated Carbon Glass Resistor (CGR) thermometers positioned on the top (Ttop) and bottom (Tbot ) flange of the magnet. A readout at the start of the measuring session yielded Ttop = 4.231 K and Tbot = 4.236 K. The sample temperature is taken as the average between these two measurements. The CERN reference data are given at a temperature T = 4.222 K and the data for sample b are corrected for this small discrepancy in temperature. Ideally, the samples should be parameterized using well established scaling relations [5] but the required scaling parameters cannot be accurately determined due to the limited data range. Also, since the temperature error is small, a limited range – linear – correction will be sufficient [6]: Ic (4.222 K) = Ic (Tbath) ( Tc (H)−4.222 Tc (H)−Tbath ) , (3) where: Tc (H) = 9.2 ( 1− μ0H 14.5 )0.59 . (4) Note that relation 4 implicitly assumes Tc(0) = 9.2 K, μ0Hc2(0) = 14.5 T and Hc2(T )/Hc2(0) = 1− (T/Tc (0)). These re-normalized data are also given in Table 1 and the resulting Ic(H) data for the 5 turn section are, together with the CERN reference data, plotted in Figure 2. Corrections for self-field were considered, but these will always be subjective and only become relevant below μ0H . 3 T. No self-field corrections were therefore applied on the results.
منابع مشابه
OPTICS IMPLICATIONS OF IMPLEMENTING Nb3Sn MAGNETS IN THE LHC PHASE I UPGRADE*
CERN has encouraged the US-LARP collaboration to participate in Phase I of the LHC luminosity upgrade by analyzing the benefits gained by using Nb3Sn technology to replace the functionality of select NbTi magnets that CERN is committed to construct. Early studies have shown that the much higher gradients (shorter magnetic lengths) and temperature margins (quench stability) of Nb3Sn magnets comp...
متن کاملRadiation and Thermal Analysis of Superconducting Quadrupoles in the Interaction Region of Linear Collider*
CERN has encouraged the US-LARP collaboration to participate in Phase I of the LHC luminosity upgrade by analyzing the benefits gained by using Nb3Sn technology to replace the functionality of select NbTi magnets that CERN is committed to construct. Early studies have shown that the much higher gradients (shorter magnetic lengths) and temperature margins (quench stability) of Nb3Sn magnets comp...
متن کاملSuperconducting transmission lines for use in a He dilution refrigerator
We compare two superconducting transmission lines for use in a He dilution refrigerator using a NbTi wire inner conductor with NbTi foil and CuNi tubing outer conductors. The heat transfer is calculated using the Wiedermann-Franz Law, and the impedance and power loss are measured using a network analyzer at room temperature and at 4K. The transmission line with CuNi tubing is easy to construct,...
متن کاملNBTI-Induced Delay Degradation Analysis of FPGA Routing Structures
Reliability issues, such as soft errors, process variations and Negative Bias Temperature Instability (NBTI), become dominant on Field Programmable Gate Arrays (FPGAs) fabricated in a nanometer process. We focus on aging degradation by NBTI, which causes threshold voltage shifts on PMOS transistors. We characterize delay degradation in the routing structures on FPGAs. The rising and falling del...
متن کاملEUROPEAN ORGANISATION FOR NUCLEAR RESEARCH CERN – PS DIVISION CERN NUFACT Note 084 CERN/PS 2001-050 (RF) DESIGN OF THE SUPERCONDUCTING SECTION OF THE SPL LINAC AT CERN
In order to set up a powerful proton source for a future Neutrino Factory, increasing at the same time the flux of protons available for new and existing facilities, CERN is studying a 2.2 GeV superconducting Hlinac for 4 MW beam power, called SPL. The superconducting part of this linac covers the energy range from 120 MeV to 2.2 GeV. Three sections made of 352 MHz cavities with nominal ß of 0....
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره شماره
صفحات -
تاریخ انتشار 2007