Effect of a Novel Intracycle Motion Correction Algorithm on Dual-Energy Spectral Coronary CT Angiography: A Study with Pulsating Coronary Artery Phantom at High Heart Rates
نویسندگان
چکیده
OBJECTIVE Using a pulsating coronary artery phantom at high heart rate settings, we investigated the efficacy of a motion correction algorithm (MCA) to improve the image quality in dual-energy spectral coronary CT angiography (CCTA). MATERIALS AND METHODS Coronary flow phantoms were scanned at heart rates of 60-100 beats/min at 10-beats/min increments, using dual-energy spectral CT mode. Virtual monochromatic images were reconstructed from 50 to 90 keV at 10-keV increments. Two blinded observers assessed image quality using a 4-point Likert Scale (1 = non-diagnostic, 4 = excellent) and the fraction of interpretable segments using MCA versus conventional algorithm (CA). Comparison of variables was performed with the Wilcoxon rank sum test and McNemar test. RESULTS At heart rates of 70, 80, 90, and 100 beats/min, images with MCA were rated as higher image scores compared to those with CA on monochromatic levels of 50, 60, and 70 keV (each p < 0.05). Meanwhile, at a heart rate of 90 beats/min, image interpretability was improved by MCA at a monochromatic level of 60 keV (p < 0.05) and 70 keV (p < 0.05). At a heart rate of 100 beats/min, image interpretability was improved by MCA at monochromatic levels of 50 keV (from 69.4% to 86.1%, p < 0.05), 60 keV (from 55.6% to 83.3%, p < 0.05) and 70 keV (from 33.3% to 69.3%, p < 0.05). CONCLUSION Low-keV monochromatic images combined with MCA improves image quality and image interpretability in CCTAs at high heart rates.
منابع مشابه
Improvement of Image Quality and Diagnostic Performance by an Innovative Motion-Correction Algorithm for Prospectively ECG Triggered Coronary CT Angiography
OBJECTIVE To investigate the effect of a novel motion-correction algorithm (Snap-short Freeze, SSF) on image quality and diagnostic accuracy in patients undergoing prospectively ECG-triggered CCTA without administering rate-lowering medications. MATERIALS AND METHODS Forty-six consecutive patients suspected of CAD prospectively underwent CCTA using prospective ECG-triggering without rate cont...
متن کاملA novel dual energy CT-based attenuation correction method in PET/CT systems: A phantom study
In present PET/CT scanners, PET attenuation correction is performed by relying on the information given by CT scan. In the CT-based attenuation correction methods, dual-energy technique (DECT) is the most accurate approach, which has been limited due to the increasing patient dose. In this feasibility study, we have introduced a new method that can implement dual-en...
متن کاملProspective-triggered high-pitch spiral versus sequential dual-source CT coronary angiography: comparison of image quality and radiation dose
Background: Prospectively electrocardiography (ECG)-triggered high-pitch spiral coronary computed tomography angiography (CCTA) is a unique scan mode for dual-source CT (DSCT). Our reports aim to compare image quality and radiation dose of CCTA using high-pitch spiral or sequential acquisition mode in patients with low and stable heart rates. Materials and Methods: Patients with low and stable ...
متن کاملEffect of Heart Rate on CT Angiography Using the Enhanced Cardiac Model of the 4D NCAT
We investigate the effect of heart rate on the quality and artifact generation in coronary artery images obtained using multi-slice computed tomography (MSCT) with the purpose of finding the optimal time resolution for data acquisition. To perform the study, we used the 4D NCAT phantom, a computer model of the normal human anatomy and cardiac and respiratory motions developed in our laboratory....
متن کاملLow-dose CT coronary angiography with a novel IntraCycle motion-correction algorithm in patients with high heart rate or heart rate variability.
AIMS Motion artefacts due to high or irregular heart rate (HR) are common limitations of coronary computed tomography (CT) angiography (CCTA). The aim of the study was to evaluate the impact of a new motion-correction (MC) algorithm used in conjunction with low-dose prospective ECG-triggering CCTA on motion artefacts, image quality, and coronary assessability. METHODS AND RESULTS Among 380 pa...