Failure of an Implantable Defibrillator to detect transition from sinus tachycardia to slow ventricular tachycardia with 'Onset' discrimination algorithm activated.
نویسندگان
چکیده
A 74-year-old male with severe, inotrope-dependent congestive heart failure (CHF) and previous episodes of slow ventricular tachycardia (VT) was evaluated for ongoing VT which persisted despite several attempts at ablation. The patient had a dual-chamber, biventricular Medtronic defibrillator programmed with a VT-1 zone of 500 ms and with ‘Onset’ as the only active discriminator. He had a history of complete heart block, and thus, the ‘PR Logic’ algorithm was not programmed. On telemetry monitoring, he was noted to have a sustained, wide complex tachycardia (Panel A). Although the tachycardia cycle length (CL) of 480 ms was faster than the VT-1 cutoff of 500 ms, no therapy was given. Stored electrograms from the event showed a sudden onset tachycardia with a CL of 490 ms with AV dissociation (Panel B). Thus, this rhythm was indeed VT. The patient’s sinus CL was 580 ms, related to the use of dobutamine for inotropic support. The change in CL from sinus tachycardia compared with slow VT was 84.5% ([490/580] × 100). As this value exceeded the programmed cutoff of 81%, the rhythm was not classified as VT, and no therapy was delivered. Following inactivation of the onset discriminator, additional episodes of slow VT were appropriately treated. The patient ultimately underwent left ventricular-assist device implantation. This case illustrates an important limitation of the ‘onset’ discriminator criterion. This algorithm requires a programmable decrease in CL over four consecutive beats to detect VT. Supraventricular tachycardia algorithms, such as ‘PR Logic’, are activated only after a rhythm is determined to not be sinus tachycardia by an activated onset discriminator. When programmed on, the default value is 81% and programmable range is from 72 to 97% in increments of 3%. The algorithm has been demonstrated to be most useful in distinguishing between sinus tachycardia and VT. In patients with high resting heart rates, as would commonly be encountered in the advanced heart failure population with low cardiac output or ongoing inotropic support, the difference between a slow VT rate and the sinus rate may be too small to fall below the nominal value of 81%. The use of antiarrhythmic drugs may also contribute to slower VTs. Clinicians should be aware of the limitations of the onset algorithm when used in the advanced heart failure population, particularly those with fast resting heart rates and/or clinical need for lower VT rate cut-offs.
منابع مشابه
First postpacing interval variability during right ventricular stimulation: a single algorithm for the differential diagnosis of regular tachycardias.
BACKGROUND Failure to differentiate supraventricular from ventricular arrhythmias is the most frequent cause of inappropriate implantable cardioverter-defibrillator therapies. Although a sudden-onset criterion is available to differentiate sustained monomorphic ventricular tachycardias (SMVTs) and sinus tachycardias (STs), SMVTs arising during ST and SMVTs gradually accelerating above the cutof...
متن کاملReduced risk for inappropriate implantable cardioverter-defibrillator shocks with dual-chamber therapy compared with single-chamber therapy: results of the randomized OPTION study.
OBJECTIVES The OPTION (Optimal Anti-Tachycardia Therapy in Implantable Cardioverter-Defibrillator Patients Without Pacing Indications) trial sought to compare long-term rates of inappropriate shocks, mortality, and morbidity between dual-chamber and single-chamber settings in implantable cardioverter-defibrillators (ICDs) patients. BACKGROUND The use of dual-chamber ICDs potentially allows be...
متن کاملImplantable cardioverter-defibrillator in a patient with dextrocardia situs inversus
Background: Dextrocardia is a congenital anomaly, which may have coexistent coronary artery disease (CAD), arrhythmias and conventional indications for device therapy. However, the implantation of transvenous leads can be technically challenging and the approach needs to be tailored to the patient's individual anatomy. Case presentation: A 54-year-old male with dextrocardia situs inversus and i...
متن کاملHeart rate variability before the onset of ventricular tachycardia: differences between slow and fast arrhythmias.
BACKGROUND We tested whether or not heart rate variability (HRV) changes can serve as early signs of ventricular tachycardia (VT) and predict slow and fast VT in patients with an implantable cardioverter defibrillator (ICD). METHODS AND RESULTS We studied the ICD stored 1000 beat-to-beat intervals before the onset of VT (131 episodes) and during a control time without VT (74 series) in 63 chr...
متن کاملVentricular tachycardia slower than the rate cut-off of a subcutaneous cardiac defibrillator sensed and successfully treated as a result of oversensing
Introduction The subcutaneous implantable cardioverter-defibrillator system (S-ICD) has been an important alternative to transvenous defibrillators for certain clinical situations, such as when there is a venous access problem or high risk of bacterial endocarditis. This device reliably detects ventricular tachyarrhythmias and effectively delivers shocks. However, oversensing is more common wit...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology
دوره 19 10 شماره
صفحات -
تاریخ انتشار 2017