Measurementof Plasma Phenytoinby EMIT Ifl the Monarch CentrifugalAnalyzer : Studies on the Eliminationof Within

نویسنده

  • Heather M. Barbour
چکیده

We investigated the measurement of phenytoin in plasma by the EMIT#{174} method, in the Monarch centrifugal analyzer. When we used the standard protocol supplied by Instrumentation Laboratory, significant drift was observed across a full rotor loaded with 31 replicates of a single specimen. Although the cause remains obscure, the drift was eliminated by using a load-spin-reload-spin procedure. This makes it possible to analyze large batches of samples without excessive use of standards and reagents.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Drug screening with EMIT reagents: a quantitative approach to quality control.

We investigated the precision, linearity, accuracy, and stability of quantitative results for five drugs of abuse [amphetamines, benzoylecgonine, opiates, phencyclidine, and the cannabinoid-tetrahydrocannabinol (THC)-9-acid metabolite], analyzed in control specimens by using EMIT d.a.u. reagents (Syva Co.) with a Monarch 2000 analyzer with a nonlinear interpolation curve-fitting algorithm. The ...

متن کامل

Studies on measurement of plasma magnesium: application of the Magon dye method to the "Monarch" centrifugal analyzer.

The "Magnesium Liquid Stable Reagent Set" from Medical Analysis Systems, Inc., is evaluated. The method, which involves Magon dye binding and bichromatic absorbance measurements, was used in a Monarch centrifugal analyzer. Results were compared with those by atomic absorption spectrometry. The calibration curve for the Magon method is linear to 2.5 mmol/L, with a 2-microL sample volume. Analyti...

متن کامل

3D Simulation of the Effects of the Plasma Actuator on the Unsteady, Turbulent and Developing Flow within a Circular Duct

The objective of current paper is 3D simulation of turbulent, developing flow and unsteady within a circular duct in presence of the body force vector persuaded by Dielectric barrier discharge (DBD) plasma actuator inside the surface of geometry for the first time. This article aims at investigating of applying plasma actuator to control separation with special arrangement of electrodes. For th...

متن کامل

Isofemale Line Analysis of Meloidogyne incognita Virulence to Cowpea Resistance Gene Rk.

Isofemale lines (IFL) from single egg masses were studied for genetic variation in Meloidogyne incognita isolates avirulent and virulent to the resistance gene Rk in cowpea (Vigna unguiculata). In parental isolates cultured on susceptible and resistant cowpea, the virulent isolate contained 100% and the avirulent isolate 7% virulent lineages. Virulence was selected from the avirulent isolate wi...

متن کامل

Urine accurately reflects circulating isoflavonoids and ascertains compliance during soy intervention.

BACKGROUND Isoflavonoids (IFL) may protect against chronic diseases, including cancer. IFL exposure is traditionally measured from plasma (PL), but the reliability of urine is uncertain. We assessed whether IFL excretion in overnight urine (OU) or spot urine (SU) reliably reflects IFLs in PL and the usefulness of the three matrices to determine soy intake compliance. METHODS In a randomized, ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2004