The pressure exerted on the tracheal wall by two endotracheal tube cuffs: A prospective observational bench-top, clinical and radiological study

نویسندگان

  • Alex Doyle
  • Ramai Santhirapala
  • Martin Crowe
  • Mark Blunt
  • Peter Young
چکیده

BACKGROUND The Lotrach endotracheal tube has a unique low-volume, low-pressure (LVLP) cuff, which has been designed to prevent pressure injury to the tracheal wall. We aimed to estimate the pressure exerted on the tracheal wall by the LVLP cuff and a conventional cuff in a bench-top, clinical and radiological study. METHOD In the bench-top study, a model trachea was intubated with the LVLP cuff and the conventional cuff. The cuff pressure was controlled using a constant pressure device. We assessed the pressure exerted on the tracheal wall by measuring the ability of the cuffs to support a column of water using a standard protocol. In the clinical study, we tested the ability of both cuffs to prevent air leak during a staged recruitment manoeuvre. In the radiological study, we recorded the degree of anatomical distortion of the trachea from both cuffs in the antero-posterior (AP) and transverse tracheal diameters. We performed statistical analysis using non-inferiority tests. RESULTS In the bench-top study, the LVLP cuff achieved a plateau at a mean height of 25.2 cmH2O (SD 0.34). In contrast, the conventional cuff failed to maintain any water above the cuff and a plateau could not be measured. In the clinical study, the mean pressure at which air leak occurred was 30.0 +/- 0.8 cmH2O (SD 3.8) using the LVLP cuff and 32.4 +/- 0.7 cmH2O (SD 3.0) using the conventional cuff. In the radiological study, the mean degree of anatomical distortion of the trachea in AP and transverse tracheal diameter was 2.9 +/- 2.2 mm (SD 2.1) and 1.8 +/- 1.4 mm (SD 1.4) using the LVLP cuff and 4.4 +/- 1.3 mm (SD 1.4) and 2.6 +/- 1.5 mm (SD 1.6) using the conventional cuff. CONCLUSIONS The bench-top and clinical studies both demonstrated that the LVLP cuff exerted approximately 30 cmH2O of pressure on the tracheal wall. These results are supported by our radiological study. We conclude that the LVLP cuff exerts an acceptable amount of pressure on the tracheal wall when it is operated at the recommended intracuff pressure.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of tracheal tube cuff inflation pressure among anesthesia team with different gradual level

Background & Objectives: Tracheal intubation is frequently performed in operating room. Over inflation of the endotracheal tube cuff is a risk factor for tracheal ischemia and subsequent complications. On the other hand inadequate cuff pressure increases the risk of aspiration of gastric contents. In this study, the endotracheal tube cuff pressure filled by the anesthesia team has been compared...

متن کامل

Pressure on the tracheal mucosa from cuffed tubes.

During cuffed intubation, damage to the trachea is least likely when the lateral wall pressure exerted by the cuff does not exceed the mean capillary perfusion pressure of the mucosa. A study was carried out of eight different types of endotracheal tubes. At the seal point the traditional red rubber tube and the armoured latex and Softway tubes exerted pressures above the mean systemic arterial...

متن کامل

Leakage of fluid around endotracheal tube cuffs: a cadaver study

BACKGROUND The aim of the study was to evaluate the leakage of liquid past the cuffs of tracheal tubes in fresh frozen human heads. METHODS Six truncated fresh frozen heads were used and intubated with 8.0 mm endotracheal tubes. The intracuff pressures tested were 30 and 100 cmH2O. Subsequently, 20 ml of each of two oral antiseptic rinses (0.2% chlorhexidine and octenidine [octenidol®, Schülk...

متن کامل

A method for comparing endotracheal cuffs. A controlled study of tracheal trauma in dogs.

Damage to the trachea produced in dogs by large and small residual volume cuffs during 6 h of IPPV was compared using a specially designed endotracheal tube. The cuffs under evaluation were adjusted to exert similar average pressures on the tracheal wall, so that many of the variables believed responsible for tracheal injury were controlled. The true compliance of the cuff was measured with the...

متن کامل

Tracheal size following tracheostomy with cuffed tracheostomy tubes: an experimental study.

In view of the severe damage caused by unyielding, low residual volume cuffs, various modifications to the cuff of an intratracheal tube have been introduced. The merits of two low-pressure cuffs were assessed in an experimental study in dogs; both cuffs produced little visible damage to the tracheal wall in dogs intubated continuously over a two-week period. A modified technique of producing t...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 10  شماره 

صفحات  -

تاریخ انتشار 2010