First robotic tracheal intubations in humans using the Kepler intubation system.
نویسندگان
چکیده
BACKGROUND Intubation is one of the most important anaesthetic skills. We developed a robotic intubation system (Kepler intubation system, KIS) for oral tracheal intubation. METHODS In this pilot study, 12 patients were enrolled after approval of the local Ethics board and written informed consent. The KIS consists of four main components: a ThrustMaster T.Flight Hotas X joystick (Guillemot Inc., New York, NY, USA), a JACO robotic arm (Kinova Rehab, Montreal, QC, Canada), a Pentax AWS video laryngoscope (Ambu A/S, Ballerup, Denmark), and a software control system. The joystick allows simulation of the wrist or arm movements of a human operator. The success rate of intubation and intubation times were measured. RESULTS Eleven men and one woman aged 66 yr were included in this study. Intubation was successful in all but one patient using KIS at a total time of [median (inter-quartile range; range)] 93 (87, 109; 76, 153) s; in one patient, fogging of the video laryngoscope prevented intubation using KIS. CONCLUSIONS We present the first human testing of a robotic intubation system for oral tracheal intubation. The success rate was high at 91%. Future studies are needed to assess the performance and safety of such a system.
منابع مشابه
Accuracy and reliability of the self-inflating bulb to verify tracheal intubation in out-of-hospital cardiac arrest patients.
BACKGROUND To determine the sensitivity and specificity of the self-inflating bulb (SIB) to verify tracheal intubation in out-of-hospital cardiac arrest patients. METHODS Sixty-five consecutive adult patients with out-of-hospital cardiac arrest were enrolled. Patients were provided chest compression and ventilation by either ba-valve-mask or the esophageal tracheal double-lumen airway by ambu...
متن کاملTracheal intubation following training with the GlideScope compared to direct laryngoscopy.
SUMMARY Tracheal intubation using direct laryngoscopy has a high failure rate when performed by untrained medical personnel. This study compares tracheal intubation following direct laryngoscopy by inexperienced medical students when initially trained by using either the GlideScope, a video assisted laryngoscope, or a rigid (Macintosh) laryngoscope. Forty-two medical students with no previous e...
متن کاملEvaluation of rapid sequence intubation in the pediatric emergency department.
OBJECTIVES To describe the experience of the emergency department of a pediatric hospital with rapid sequence intubation (RSI) and to identify the factors associated with successful intubation. METHODS This prospective, observational, cross-sectional study conducted from July 2005 to December 2007 consisted of collection of data regarding tracheal intubations performed at the emergency depart...
متن کاملOut-of-hospital endotracheal intubation experience and patient outcomes.
STUDY OBJECTIVE Previous studies suggest improved patient outcomes for providers who perform high volumes of complex medical procedures. Out-of-hospital tracheal intubation is a difficult procedure. We seek to determine the association between rescuer procedural experience and patient survival after out-of-hospital tracheal intubation. METHODS We analyzed probabilistically linked Pennsylvania...
متن کاملLearning Curves of Macintosh Laryngoscope in Nurse Anesthetist Trainees Using Cumulative Sum Method
Background. Tracheal intubation is a potentially life-saving procedure. This skill is taught to many anesthetic healthcare professionals, including nurse anesthetists. Our goal was to evaluate the learning ability of nurse anesthetist trainees in their performance of orotracheal intubation with the Macintosh laryngoscope. Methods. Eleven nurse anesthetist trainees were enrolled in the study dur...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- British journal of anaesthesia
دوره 108 6 شماره
صفحات -
تاریخ انتشار 2012