Assessment of new radioimmunoassay kit for determining urinary albumin at low concentrations: comparison with radial immunodiffusion.
نویسندگان
چکیده
The assay characteristics of a new radioimmunoassay kit for determining urinary albumin at low concentrations were studied. The sensitivity for urinary albumin was 2 mg/l, the analytical range 2 to 40 mg/l, and interassay coefficient of variation less than 12%. In a method comparison study entailing diabetic urine samples covering an albumin concentration of 2 to 150 mg/l the kit compared adequately with radial immunodiffusion (mean difference between methods = 2 mg/l; residual standard deviation = 4.6 mg/l), absolute variation between methods increasing with the concentration. The kit required much less skill than radial immunodiffusion but its capital and running cost were higher.
منابع مشابه
Immunoturbidimetry of Unnary Albumin: Prevention of Adsorption of Albumin; Influence of Other UrinaryConstituents
I describe a simple, rapid immunoturbidimetric assay for low concentrations of albumin in unne (2 to 260 mg/L). However, in this assay, the human serum albumin (HSA) in the standards binds nonspecifically to the polystyrene or glass tubes. This nonspecific binding cannot be prevented by adding bovine serum albumin (BSA) to standards, because the anti-HSA antibodycross reacts with BSA. Adding Tr...
متن کاملComparison of commercial kits for apoprotein A-I and apoprotein B with standardized apoprotein A-I and B radioimmunoassays performed at the Northwest Lipid Research Center.
There has recently been a proliferation of commercial kits available for apoproteins A-I and B. Since reference procedures for apoproteins have not yet been established we have elected to compare apoprotein kit methods with highly standardized apoprotein B and A-I radioimmunoassays developed at the Northwest Lipid Research Center. Commercial radial immunodiffusion kits for apoproteins A-I and B...
متن کاملMeasurement of the staphylococcal epidermolytic toxin: a comparison of bioassay, radial immunodiffusion, and radioimmunoassay.
Three methods of measuring the epidermolytic toxin Staphylococcus aureus-bioassay in newborn mice, radial immunodiffusion, and radioimmunoassay-were compared for reproducibility, specificity, and sensitivity. The bioassay is highly specific and remains the only functional assay. It is reproducible only if newborn mice of the same age are used. The time required for epidermolysis follows a dose-...
متن کاملA radioimmunoassay for retinol-binding protein in serum and urine.
In this radioimmunoassay for human retinol-binding protein (RBP), sample (serum, diluted as much as 3600-fold; or urine, diluted or undiluted) or calibrant (purified RBP in phosphate buffer containing bovine serum albumin) is incubated with 125I-labeled RBP and rabbit anti-human RBP antiserum for 24 h before separation with second antibody/polyethylene glycol. The assay's sensitivity is 5 micro...
متن کاملUrinary albumin radioimmunoassay using a solid phase second antibody and solid phase albumin iodination.
The measurement of microalbuminuria in diabetics is now clinically important as a predictor of later overt clinical proteinuria, reflecting advanced diabetic nephropathy [l-3]. Urine albumin assays suitable for routine use in the clinical laboratory are therefore important. Measurement of microalbuminuria by current techniques including immunodiffusion [4] and electroimmunoassay [5] are time-co...