High-intensity focused ultrasound treatment in refractory glaucoma patients: results at 1 year of prospective clinical study.
نویسندگان
چکیده
PURPOSE To evaluate the safety and efficacy of the ultrasonic circular cyclo-coagulation procedure using high-intensity focused ultrasound by a miniaturized annular device containing 6 piezoceramic transducers in patients with refractory glaucoma. METHODS This was a prospective interventional noncomparative study of 20 eyes of 20 patients with refractory glaucoma. All eyes were treated with 6 activated transducers operating at 21 MHz. Ultrasound biomicroscopy and a complete ophthalmic examination were performed before the procedure and at 1 day, 1 week, and 1, 3, 6, and 12 months after the procedure. Primary outcomes were surgical success (defined as intraocular pressure (IOP) reduction from baseline ≥20% and IOP >5 mm Hg) at the last follow-up visit. Secondary outcomes were mean IOP at each follow-up visit compared to baseline, medication use, complications, and re-interventions. RESULTS Intraocular pressure was significantly reduced (p<0.01) from a mean preoperative value of 36.4 ± 5.7 mm Hg to a mean postoperative value of 22.5 ± 10.3 mm Hg at 12 months. Four patients needed to be re-treated. The mean IOP reduction achieved was 38%. Surgical success was achieved in 13 of 20 eyes (65%). No major intraoperative or postoperative complications occurred. CONCLUSIONS Ultrasonic circular cyclo-coagulation using high-intensity focused ultrasound delivered by a circular miniaturized device containing 6 piezoceramic transducers is an effective and well-tolerated method to reduce IOP in patients with refractory glaucoma.
منابع مشابه
Cyclocoagulation of the ciliary bodies by high-intensity focused ultrasound: a 12-month multicenter study.
PURPOSE To evaluate the safety and efficacy of high-intensity focused ultrasound (HIFU) cyclocoagulation in reducing intraocular pressure (IOP) in patients with refractory glaucoma by using a novel miniaturized delivery device (EyeOP1). METHODS We conducted a 12-month open-label multicenter prospective study (EyeMUST1 Study). Patients with primary (primary open-angle glaucoma [POAG]) or secon...
متن کاملTherapeutic effects of high intensity focused ultrasound for in patients with benign uterine tumors with regard to lifestyle and obstetric characteristics
Background: The present investigation compared the decrease in the volume of benign tumor before and after High Intensity Focused Ultrasound (HIFU) treatment to evaluate the therapeutic effects of HIFU. in terms of Lifestyle and obstetric characteristics of in patients with Benign Uterine Tumors. Materials and Methods: To collect the data on each patients’ lifestyle and obstetric characteristic...
متن کاملUltrasound Cycloplasty in Glaucoma – Mechanisms of Action and their Possible Impact on Intraocular Pressure
U ltrasound cycloplasty (UCP) is a procedure based on high-intensity focused ultrasound (HIFU) for refractory and non-refractory glaucoma in cases for which surgery is not suitable. UCP may bring about a reduction in intraocular pressure (IOP) via multiple mechanisms of action that result in a decrease in aqueous humour production and increase in aqueous humour evacuation. UCP appears to have a...
متن کاملHigh-Intensity Focused Ultrasound Circular Cyclocoagulation in Glaucoma: A Step Forward for Cyclodestruction?
The ciliary body ablation is still considered as a last resort treatment to reduce the intraocular pressure (IOP) in uncontrolled glaucoma. Several ablation techniques have been proposed over the years, all presenting a high rate of complications, nonselectivity for the target organ, and unpredictable dose-effect relationship. These drawbacks limited the application of cyclodestructive procedur...
متن کاملNumerical Study for Optimizing Parameters of High-Intensity Focused Ultrasound-Induced Thermal Field during Liver Tumor Ablation: HIFU Simulator
Introduction High intensity focused ultrasound (HIFU) is considered a noninvasive and effective technique for tumor ablation. Frequency and acoustic power are the most effective parameters for temperature distribution and the extent of tissue damage. The aim of this study was to optimize the operating transducer parameters such as frequency and input power in order to acquire suitable temperatu...