A new technique to measure micromotion distribution around a cementless femoral stem.
نویسندگان
چکیده
The interfacial micromotion is closely associated to the long-term success of cementless hip prostheses. Various techniques have been proposed to measure them, but only a few number of points over the stem surface can be measured simultaneously. In this paper, we propose a new technique based on micro-Computer Tomography (μCT) to measure locally the relative interfacial micromotions between the metallic stem and the surrounding femoral bone. Tantalum beads were stuck at the stem surface and spread at the endosteal surface. Relative micromotions between the stem and the endosteal bone surfaces were measured at different loading amplitudes. The estimated error was 10 μm and the maximal micromotion was 60 μm, in the loading direction, at 1400 N. This pilot study provided a local measurement of the micromotions in the 3 direction and at 8 locations on the stem surface simultaneously. This technique could be easily extended to higher loads and a much larger number of points, covering the entire stem surface and providing a quasi-continuous distribution of the 3D interfacial micromotions around the stem. The new measurement method would be very useful to compare the induced micromotions of different stem designs and to optimize the primary stability of cementless total hip arthroplasty.
منابع مشابه
Initial Stability of Cementless Femoral Stems: An In Vitro Technique to Measure Micromotion and Gap Around the Loaded Stem
Currently available techniques used to measure experimentally micromotion directly at the bone-implant interface rely on LVDT sensors [2]. They allow a good precision but they also include the bone deformation between the device fixation point and the measurement site. Moreover, the reported maximum number of simultaneous measurement points on bone with sensors was six [3]. Introduction Valérie...
متن کاملSimultaneous and multisite measure of micromotion, subsidence and gap to evaluate femoral stem stability.
The initial stability of cementless femoral components is crucial for the long-term success of total hip arthroplasty. This has been reported in animal and clinical studies. Until now, the stability was evaluated by the measurement of relative micromotion on a few simultaneous locations around the stem in cadaveric experiments. This paper presents an extended experimental setup to measure simul...
متن کاملFull-field measurement of micromotion around a cementless femoral stem using micro-CT imaging and radiopaque markers.
A good primary stability of cementless femoral stems is essential for the long-term success of total hip arthroplasty. Experimental measurement of implant micromotion with linear variable differential transformers is commonly used to assess implant primary stability in pre-clinical testing. But these measurements are often limited to a few distinct points at the interface. New techniques based ...
متن کاملPrimary Stability Recognition of the Newly Designed Cementless Femoral Stem Using Digital Signal Processing
Stress shielding and micromotion are two major issues which determine the success of newly designed cementless femoral stems. The correlation of experimental validation with finite element analysis (FEA) is commonly used to evaluate the stress distribution and fixation stability of the stem within the femoral canal. This paper focused on the applications of feature extraction and pattern recogn...
متن کاملMicromotion-induced peri-prosthetic fluid flow around a cementless femoral stem.
Micromotion-induced interstitial fluid flow at the bone-implant interface has been proposed to play an important role in aseptic loosening of cementless implants. High fluid velocities are thought to promote aseptic loosening through activation of osteoclasts, shear stress induced control of mesenchymal stem cells differentiation, or transport of molecules. In this study, our objectives were to...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Journal of biomechanics
دوره 44 3 شماره
صفحات -
تاریخ انتشار 2011