A versatile facility for the calibration of X-ray polarimeters with polarized and unpolarized controlled beams
نویسندگان
چکیده
We devised and built a versatile facility for the calibration of the next generation X-ray polarimeters with unpolarized and polarized radiation. The former is produced at 5.9 keV by means of a Fe radioactive source or by X-ray tubes, while the latter is obtained by Bragg diffraction at nearly 45 degrees. Crystals tuned with the emission lines of X-ray tubes with molybdenum, rhodium, calcium and titanium anodes are employed for the efficient production of highly polarized photons at 2.29, 2.69, 3.69 and 4.51 keV respectively. Moreover the continuum emission is exploited for the production of polarized photons at 1.65 keV and 2.04 keV and at energies corresponding to the higher orders of diffraction. The photons are collimated by means of interchangeable capillary plates and diaphragms, allowing a trade-off between collimation and high fluxes. The direction of the beam is accurately arranged by means of high precision motorized stages, controlled via computer so that long and automatic measurements can be done. Selecting the direction of polarization and the incidence point we can map the response of imaging devices to both polarized and unpolarized radiation. Changing the inclination of the beam we can study the systematic effects due to the focusing of grazing incidence optics and the feasibility of instruments with large field of view.
منابع مشابه
Unpolarized nucleon structure studies utilizing polarized electromagnetic probes
By the mid-1980s, measurements of the nucleon form factors had reached a stage where only slow, incremental progress was possible using unpolarized electron scattering. The development of high quality polarized beams, polarized targets, and recoil polarimeters led to a renaissance in the experimental program. I provide an overview of the changes in the field in the last ten years, which were dr...
متن کاملLow-Energy Deuteron Polarimeter
We built a Low Energy Deuteron Polarimeter (LDP) which measures the spin-polarization of deuteron beams in the energy range of 25 to 80 keV. The LDP works by measuring azimuthal asymmetries in the D(~ d, n)He reaction at θlab = 90 ◦ and comparing them to analyzing powers. We built this polarimeter for two reasons. Firstly, to crosscheck other polarimeters at KVI (LSP and IBP). Secondly to test ...
متن کاملImpact of Photon Spectra on the Sensitivity of Polymer Gel Dosimetry by X-Ray Computed Tomography
Introduction: The purpose of the current study was to investigate the effect of X-ray spectra on the sensitivity of a polymer gel dosimeter imaged with a conventional computed tomography (CT) scanner. Material and Methods: The whole process of CT imaging of an irradiated polymer gel was simulated by MCNPX Monte Carlo (MC) code. The imaging of polyacrylamide gel was accomplished by means of a co...
متن کاملX-Ray Polarimetry: Historical Remarks and Other Considerations
We briefly discuss the history of X-ray polarimetry for astronomical applications including a guide to the appropriate statistics. We also provide an introduction to some of the new techniques discussed in more detail elsewhere in these proceedings. We conclude our discussion with our concerns over adequate ground calibration, especially with respect to unpolarized beams, and at the system level.
متن کاملComparison Between the NIST and the KEBS for the Determination of Air Kerma Calibration Coefficients for Narrow X-Ray Spectra and 137Cs Gamma-Ray Beams
Air kerma calibration coefficients for a reference class ionization chamber from narrow x-ray spectra and cesium 137 gamma-ray beams were compared between the National Institute of Standards and Technology (NIST) and the Kenya Bureau of Standards (KEBS). A NIST reference-class transfer ionization chamber was calibrated by each laboratory in terms of the quantity air kerma in four x-ray referenc...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره شماره
صفحات -