Optimum Spatial Resolution and Number of Averages of Diffusion Tensor Imaging in Prostate Cancer Detection
نویسندگان
چکیده
This study aimed to investigate the diagnostic performance of trace apparent diffusion coefficient (tADC) maps in prostate cancer (PCA) detection at different spatial resolutions and number of averages (NAV), and to determine a practical protocol that could provide satisfactory diagnostic performance for PCA detection. The tADC maps were reconstructed into three different isotropic resolutions, i.e. 1 mm, 2 mm and 3 mm. For each resolution there were three different NAVs, i.e. 2, 4, and 6 averages. The diagnostic performance of tADC maps for each data set was determined by comparing with the TRUS biopsy results core by core. We found that the voxel size of (2 mm) with two number of averages showed satisfactory diagnostic performance. Acquisition time for this protocol was around 4 to 5 minutes and could be a practical tool for PCA detection. Introduction Magnetic resonance (MR) diffusion imaging has been applied to the detection of prostate cancer [1]. Recently we have shown that trace apparent diffusion coefficient (tADC) value was significantly lower in prostate cancer (PCA) than that in normal peripheral zone tissue, and that 1-mm isotropic tADC maps were feasible for PCA detection [2]. To obtain tADC maps at 1-mm isotropic resolution, however, requires multiple averages to attain sufficient SNR, inevitably prolonging the scanning time. In this study, nine combinations of image resolution and number of averages of diffusion-weighted images (DWI) were investigated. By comparing with transrectal ultrasound guided (TRUS) biopsy results, we aimed to determine a practical protocol that could provide satisfactory diagnostic performance for PCA detection. Materials and Methods Fifteen male patients (57-74 years; average, 63 years; median, 63 years) with elevated prostate specific antigen (PSA, mean: 9.7 ng/ml) were recruited. They received endorectal MRI study followed by TRUS biopsy within one month. MR images were acquired on a 1.5T scanner (GE, Echo Speed, Milwaukee, WI, USA). Diffusion tensor imaging (DTI) was acquired using spin-echo echo planar imaging (EPI) with multiple transaxial slices of the prostate from base to apex. Imaging parameters: TR/TE = 17000/79 ms; slice thickness = 1mm; slice gap = 0, in-plane resolution = 1mm x 1mm; six diffusion-sensitive gradients at {±1,0,1}, {0,1,±1},{±1,1,0} with b = 500 s mm; number of averages (NAV) = 6. The acquired DWI images were reconstructed into three different isotropic resolutions, i.e. 1 mm, 2 mm and 3 mm. For each resolution there were three different NAVs, i.e. 2, 4, and 6 averages. The DWI data sets were entitled as D12, D14, D16, D22 D24, D26, D32, D34, and D36, where Dij represented the data set with isotropic resolution = i mm and NAV = j. TRUS biopsy was performed by sampling 12 cores systemically in the prostate gland, from right lateral, right medial, left medial to left lateral aspects at three levels at base, mid and apex. For image analysis, the peripheral zone of the prostate was identified and categorized into 12 regions as those in the TRUS biopsy. Trace ADC was determined by calculating the mean of the eigenvalues of the diffusion tensor. According to our previous work, a tADC value of 1.0 μm/ms was used as a threshold for PCA detection [2]. The diagnostic performance of tADC maps for each data set was determined by comparing with the TRUS biopsy results core by core. Results In a total of 180 cores, 39 cores in 7 patients were found to contain PCA in pathological results. Figure 1 shows tADC maps of one patient derived from 9 DWI data sets. The positive nodules (indicated by arrows) are those detected based on tADC < 1.0 μm/ms. One can see that as the resolution decreased, the number of positive nodules also decreased. Table 1 shows the diagnostic performance of nine data sets with nine different combinations of image quality. The best diagnostic accuracy was found in the data sets at 2-mm resolution, with little improvement as NAV increased. Discussion and Conclusions In this study, we have compared diagnostic performance of PCA detection among nine DWI data sets with different combinations of image quality, three spatial resolutions and three NAVs. From Fig. 1, the tADC positive nodules decrease with the voxel sizes. This can be explained by the partial volume effect. Increased NAV could reduce the number of positive nodules on tADC maps especially at 1-mm resolution. High NAV used at 1-mm resolution, however, would prolong the acquisition time and lead to higher failure rate and more motion artifacts. Comparing with the TRUS biopsy results (Table 1), we can see that as NAV increases, false positive nodules decrease but false negative nodules increase. Although the data set at 3-mm resolution has better signal-to-noise ratio, it does not help in increasing the sensitivity under the condition of inadequate spatial resolution. To conclude, higher NAV yields higher specificity whereas smaller voxel sizes yield higher sensitivity. In this study, the voxel size of (2 mm) with two number of averages showed satisfactory diagnostic performance. This protocol only requires 4~5 minutes acquisition time and could serve as a practical tool for PCA detection. Fig.1 Positive nodules (yellow arrows) with quantitative tADC value of less than 1.0μm/ms rendered from 9 data sets of DWI with different combinations of spatial resolution and number of averages. 83% 92% 60% 87% 72% 122 11 19 28 D36 83% 92% 59% 86% 74% 121 10 20 29 D34 83% 92% 59% 86% 74% 121 10 20 29 D32 86% 98% 61% 84% 92% 118 3 23 36 D26 85% 98% 60% 83% 92% 117 3 24 36 D24 85% 98% 60% 83% 92% 117 3 24 36 D22 84% 99% 58% 81% 97% 114 1 27 38 D16 77% 99% 49% 72% 97% 101 1 40 38 D14 69% 100% 41% 61% 100% 86 0 55 39 D12 Accuracy NPV PPV Specificity Sensitivity True negative False negative False positive True positive Table 1 The results of 9 imaging resolution levels using tADC value of1.0μm/ms. References1. Miao H et al. Prostate cancer detection with 3-T MRI: Comparison of diffusion-weightedand T2-weighted imaging. Eur J Rad 2007;61:297-302.2. Chen YJ et al. Prostate Cancer Detection in Patients with Intermediate Prostate SpecificAntigen Level Using Combined Trace Apparent Diffusion Coefficient and Nodular Size:Comparison with Transrectal Core Biopsy. ISMRM 2006;2930. Proc. Intl. Soc. Mag. Reson. Med. 17 (2009)2256
منابع مشابه
Signal Intensity of High B-value Diffusion-weighted Imaging for the Detection of Prostate Cancer
Background: Diffusion-weighted imaging (DWI) is a main component of multiparametric MRI for prostate cancer detection. Recently, high b value DWI has gained more attention because of its capability for tumor characterization. Objective: To assess based on histopathological findings of transrectal ultrasound (TRUS)-guided prostate biopsy as a reference, an...
متن کاملبررسی ارزش Diffusion Weighted Imaging در تصاویر MRI برای تشخیص سرطان پروستات
Background: Prostate cancer is the third leading cause of death and is the most common cause of cancer in elderly men. Regarding to the low accuracy of screening methods such as prostate-specific antigen (PSA), Digital Rectal Examination (DRE) and trans rectal ultrasound (TRUS) in detection and localization of tumor, Magnetic Resonance Imaging (MRI) and Diffusion Weighted Imaging (DWI) attracte...
متن کاملDetermination of Fiber Direction in High Angular Resolution Diffusion Images using Spherical Harmonics Functions and Wiener Filter
Diffusion tensor imaging (DTI) MRI is a noninvasive imaging method of the cerebral tissues whose fibers directions are not evaluated correctly in the regions of the crossing fibers. For the same reason the high angular resolution diffusion images (HARDI) are used for estimation of the fiber direction in each voxel. One of the main methods to specify the direction of fibers is usage of the spher...
متن کاملDetection of prostate cancer by an FDG-PET cancer screening program: results from a Japanese nationwide survey
Objective(s): The aim of this study was to analyze detection rates and effectiveness of 18F-fluorodeoxyglucose positron emission tomography (FDG-PET) cancer screening program for prostate cancer in Japan, which is defined as a cancer-screening program for subjects without known cancer. It contains FDG-PET aimed at detection of cancer at an early stage with or without additional screening tests ...
متن کاملDirect comparison of 99mTc-PSMA SPECT/CT and 68Ga-PSMA PET/CT in patients with prostate cancer
Objective(s): 99mTc-PSMA SPECT/CT is a cost effective alternative for 68Ga-PSMA PET/CT. The aim of this study was to directly compare these two techniques in patients with prostate cancer. Methods: 28 man with prostate cancer were studied using 99mTc-PSMA SPECT/CT and 68Ga-PSMA PET/CT in a shor...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره شماره
صفحات -
تاریخ انتشار 2008