[Evaluation of Allergen Specific IgE assay on ADVIA Centaur Immunoassay System.].
نویسندگان
چکیده
BACKGROUND Allergen specific IgE (sIgE) assay is an important aid in the diagnosis and treatment of allergy. We evaluated the analytical performance of a quantitative chemiluminescence immunoassay for sIgE using the continuous random access ADVIA Centaur. METHODS Six ADVIA Centaur sIgE reagents for common inhalant allergens in Korea, d1, d2, e1, e5, t3, and t7, were evaluated for precision, dilution recovery (parallelism), comparison with Pharmacia UniCAP sIgE assay and skin prick test, sample volume, and analytical speed according to the NCCLS guidelines (I/LA20-A, EP5-A2). Commercialized positive and negative quality control materials were used for a precision study, and samples from a total of 110 patients were used for dilution recovery and comparison studies. RESULTS Within-run coefficients of variation (CV) of the 6 items were 3.45-6.14% and within-device CVs (total CVs) of all items were below 10%. Interdilutional CVs of all items were 2.84-11.95%, which showed a good linearity and parallelism over its measuring range. Positive/negative concordance rates of the 6 items with UniCAP sIgE assay were 76.3-96.1% (d1, 88.2%; d2, 96.1%; e1, 91.0%; e5, 77.0%; t3, 90.5%; and t7, 76.3%). Concordance rates of the six items with skin prick test were all above 80%. The quantity of sample volume (25 microL/test) needed was relatively small, and a high throughput (120 tests/hr) and rapid turnaround time (47 min) could be achieved. CONCLUSIONS The ADVIA Centaur sIgE assay was thought to be a convenient and efficient method to be used in medium- to large-sized laboratories.
منابع مشابه
Evaluation of the ADVIA Centaur TSH-3 assay.
An analytical evaluation of the thyroid stimulating hormone (TSH-3) assay on the Bayer ADVIA Centaur immunoassay system was performed. General analytical requirements (linearity, resistance to typical interferences, absence of a carry-over effect) were fulfilled and reproducibility was satisfactory. Inter-assay coefficient of variation (CV) of a human serum pool with a concentration of 0.014 mU...
متن کاملPreliminary evaluation of the AxSYM B-type natriuretic peptide (BNP) assay and comparison with the ADVIA Centaur BNP assay.
and fluorescence de-rivatisation with isonicotinic acid hydrazide. Chromatographia 2000;52:810 – 4. 15. Murphy BEP. Clinical evaluation of urinary cor-tisol determinations by competitive protein-binding radioassay. Some studies of the protein-binding of steroids and their application to the routine micro and ultramicro measurement of various steroids in body fluids by competitive protein-bindin...
متن کاملShould functional sensitivity of a new thyroid stimulating hormone immunoassay be monitored routinely? The ADVIA Centaur TSH3-UL assay experience.
OBJECTIVES We emphasize the importance of routine follow-up of subnormal TSH concentrations with QC materials. DESIGN AND METHODS The functional sensitivity (FS) of the ADVIA Centaur system TSH assay was assessed. We report the values yielded for QC materials in two clinical laboratories. RESULTS The FS was <0.02 mIU/L. The low-TSH QC (a serum pool) showed unacceptable between-lot imprecisi...
متن کاملEvaluation of the siemens HIV antigen-antibody immunoassay.
Fourth-generation assays for the simultaneous detection of human immunodeficiency virus (HIV) antigen and antibodies are available on the international market and are currently used for blood donor screening and for HIV diagnosis. In this study we evaluated the performance of the novel automated fourth-generation ADVIA Centaur® HIV Ag/Ab Combo assay. The assay detected seroconversion at the sam...