Four dimensional spectral-spatial fat saturation pulse design.
نویسندگان
چکیده
PURPOSE The conventional spectrally selective fat saturation pulse may perform poorly with inhomogeneous amplitude of static (polarizing) field (B0 ) and/or amplitude of (excitation) radiofrequency field (B1 ) fields. We propose a four dimensional spectral-spatial fat saturation pulse that is more robust to B0/B1 inhomogeneity and also shorter than the conventional fat saturation pulse. THEORY The proposed pulse is tailored for local B0 inhomogeneity, which avoids the need of a sharp transition band in the spectral domain, so it improves both performance and pulse length. Furthermore, it can also compensate for B1 inhomogeneity. The pulse is designed sequentially by small-tip-angle approximation design and an automatic rescaling procedure. METHODS The proposed method is compared to the conventional fat saturation in phantom experiments and in vivo knee imaging at 3 T for both single-channel and parallel excitation versions. RESULTS Compared to the conventional method, the proposed method produces superior fat suppression in the presence of B0 and B1 inhomogeneity and reduces pulse length by up to half of the standard length. CONCLUSION The proposed four dimensional spectral-spatial fat saturation suppresses fat more robustly with shorter pulse length than the conventional fat saturation in the presence of B0 and B1 inhomogeneity.
منابع مشابه
Four-Dimensional Spectral-Spatial Fat Saturation Pulse Design at 3T
Fig. 2: The map and the ratio images by different pulses (every third slice) in the phantom experiment. Oil is on top of water, and the map is in Hz. 2117 Four-Dimensional Spectral-Spatial Fat Saturation Pulse Design at 3T Feng Zhao, Jeffrey A Fessler, and Douglas C Noll Biomedical Engineering, The University of Michigan, Ann Arbor, MI, United States, EECS, The University of Michigan, Ann Arbor...
متن کاملDixon Reconstruction
This approach is particularly important at lower field strengths, where it is the preferred method of lipid suppression. At 0.5T and below, chemically selective excitation or saturation is impractical since the small fat/water spectral shift would require a very long RF pulse. For example, the minimum time for a spectral-spatial pulse at 0.5T is about 24 ms. Inversion recovery methods become le...
متن کاملMagnetization preparation during the steady state: fat-saturated 3D TrueFISP.
A novel fat saturation scheme is proposed which combines spectral fat saturation with a steady-state 3D true Fast Imaging with Steady Precesson (TrueFISP) sequence. Fat saturation consisted of a conventional frequency-selective excitation pulse surrounded by spoiler gradients. This saturation block was periodically repeated within a continuously running TrueFISP sequence. Except for the fat sig...
متن کاملFat-suppression techniques for 3-T MR imaging of the musculoskeletal system.
Fat suppression is an important technique in musculoskeletal imaging to improve the visibility of bone-marrow lesions; evaluate fat in soft-tissue masses; optimize the contrast-to-noise ratio in magnetic resonance (MR) arthrography; better define lesions after administration of contrast material; and avoid chemical shift artifacts, primarily at 3-T MR imaging. High-field-strength (eg, 3-T) MR i...
متن کاملFast Fat Suppression Technique for Adaptive 3D Radial MRI Based on Multidimensional Golden Means
INTRODUCTION Adaptive sampling of k-space allows the reconstruction of images with various spatial and temporal resolutions from the same data set and is therefore suitable for dynamic MRI [1,2]. Golden-angle radial k-space sampling achieves this flexibility in-plane with relatively uniform angular sampling distribution for any time interval using samples incremented by the golden angle, φ = (√...