Incomplete seating of the liner with the Trident acetabular system: a cause for concern?

نویسندگان

  • A J Langdown
  • R J Pickard
  • C M Hobbs
  • H J Clarke
  • D J N Dalton
  • M L Grover
چکیده

We reviewed the initial post-operative radiographs of the Trident acetabulum and identified a problem with seating of the metal-backed ceramic liner. We identified 117 hips in 113 patients who had undergone primary total hip replacement using the Trident shell with a metal-backed alumina liner. Of these, 19 (16.4%) were noted to have incomplete seating of the liner, as judged by plain anteroposterior and lateral radiographs. One case of complete liner dissociation necessitating early revision was not included in the prevalence figures. One mis-seated liner was revised in the early post-operative period and two that were initially incompletely seated were found on follow-up radiographs to have become correctly seated. There may be technical issues with regard to the implanting of this prosthesis of which surgeons should be aware. However, there is the distinct possibility that the Trident shell deforms upon implantation, thereby preventing complete seating of the liner.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Canted seating of the Stryker Modular Dual Mobility liner within a Trident hemispherical acetabular shell

A 75-year-old woman who suffered a left femoral neck fracture underwent a left total hip arthroplasty using a Stryker Trident (Kalamazoo, MI) hemispherical acetabular shell and Modular Dual Mobility (MDM) metal liner. Post-operative radiographs demonstrated canted seating of the liner. The patient was taken immediately back to the operating room where the acetabular liner appeared well seated s...

متن کامل

Do thin acetabular shells increase the disassociation risk of ceramic liners?

This study suggests that contemporary modular acetabular shell designs employing ceramic liners deform during implantation where acetabular bed under-reaming is employed. The liner retention strength did not degrade by comparison to non-deformed shells using ASTM static push-out methods. A larger question, however, is the influence of this deformation on liner retention strength over time when ...

متن کامل

Late liner disassociation of a Pinnacle system acetabular component.

This article describes a case of late locking mechanism failure and disassociation of a Pinnacle acetabular cup (DePuy, Warsaw, Indiana) and Marathon polyethylene liner (DePuy) 53 months after routine primary total hip arthroplasty in an active patient. Following an uncomplicated initial postoperative recovery, the patient felt a pop while kneeling for gardening activities. The patient had no p...

متن کامل

Wear of a sequentially annealed polyethylene acetabular liner

BACKGROUND AND PURPOSE We previously reported on a randomized controlled trial (RCT) that examined the effect of adding tobramycin to bone cement after femoral stem migration. The present study examined femoral head penetration into both conventional and highly crosslinked polyethylene acetabular liners in the same group of RCT patients, with a minimum of 5 years of postoperative follow-up. P...

متن کامل

Metallosis after Exchange of the Femoral Head and Liner following Ceramic Acetabular Liner Dissociation in Total Hip Arthroplasty with a Modular Layered Acetabular Component

The type of bearing material that should be used in revision surgery after the failure of ceramic-on-ceramic total hip arthroplasty (THA) remains controversial. In the case of ceramic fracture, the residual ceramic particles can cause consequent metallosis when metal implants are used for revision THA. On the other hand, in the case of THA failure without ceramic fracture, revision THA with a m...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The Journal of bone and joint surgery. British volume

دوره 89 3  شماره 

صفحات  -

تاریخ انتشار 2007