Quantification of perflutren microsphere contrast destruction during transit through an ex vivo extracorporeal membrane oxygenation circuit.

نویسندگان

  • David G Platts
  • Charles McDonald
  • Kiran Shekar
  • Darryl J Burstow
  • Daniel Mullany
  • Marc Ziegenfuss
  • Sara Diab
  • John F Fraser
چکیده

BACKGROUND Echocardiography is a key investigation in the management of patients on extracorporeal membrane oxygenation (ECMO). However, echocardiographic images are often non-diagnostic in this patient population. Contrast-enhanced echocardiography may overcome many of these limitations but contrast microspheres are hydrodynamically labile structures prone to destruction from shear forces and turbulent flow, which may exist within an ECMO circuit. This study sought to evaluate microsphere destruction (utilising signal intensity as a marker of contrast concentration) during transit through an ECMO circuit. METHODS Activated Definity® contrast was diluted to 50 ml with normal saline and infused into a crystalloid primed ex vivo ECMO with a Quadrox oxygenator at 150 ml/h. Imaging was performed on pre- and post-pump head/oxygenator sections of the circuit using a Philips iE33 scanner and S5-1 transducer. Five-millimetre regions of interest were placed in the centre of the ultrasound field. Average signal intensity (decibels) was calculated at speeds of 1000, 2000, 3000 and 4000 rpm and then repeated with an infusion rate of 300 ml/h. The oxygenator was then spliced out of the circuit and the measures repeated. RESULTS There was a significant reduction in contrast concentration during passage through the ECMO circuit at all speeds (with higher pump head speeds resulting in greater microsphere destruction). In a circuit with an oxygenator, relative decrease in signal intensity was 21.4 versus 5.2 % without an oxygenator. There was significant destruction of contrast microspheres during passage through the ECMO circuit at all pump head speeds. An oxygenator contributed to microsphere destruction at a significantly greater level than the pump head alone. There was no significant difference in mean signal intensity reduction in the circuit between an infusion of 150 or 300 ml/h (3.5 ± 3.2 versus 3.6 ± 2.5 dB, respectively, p = 0.79). CONCLUSIONS Flow of contrast through an ECMO circuit results in significant destruction of microspheres. Circuits with an oxygenator result in significantly greater levels of contrast destruction than by the pump head alone. Clinicians should be cognisant of the relationship between ECMO circuit configurations, pump head speed and contrast destruction when performing a contrast-enhanced echocardiogram in patients supported with ECMO.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The ECMO PK Project: an incremental research approach to advance understanding of the pharmacokinetic alterations and improve patient outcomes during extracorporeal membrane oxygenation

BACKGROUND Extracorporeal membrane oxygenation (ECMO) is a supportive therapy and its success depends on optimal drug therapy along with other supportive care. Emerging evidence suggests significant interactions between the drug and the device resulting in altered pharmacokinetics (PK) of vital drugs which may be further complicated by the PK changes that occur in the context of critical illnes...

متن کامل

Membrane Oxygenation: A Case Report

Amphotericin B treatment during extracorporeal membrane oxygenation is at times necessary for the treatment of secondary fungemia or, less commonly, to treat a primary fungal infection. The impact of the extracorporeal circuit on the pharmacokinetics of amphotericin B have not been described in vivo. This case report describes the pharmacokinetics of amphotericin B in a patient receiving extrac...

متن کامل

In-vitro quantification of gaseous microemboli in two extracorporeal life support circuits.

During the course of extracorporeal membrane oxygenation, patients are at constant risk of exposure to air emboli. Air emboli may enter the circuit during routine lab sampling, medication administration, air entrainment through the venous cannula, or via a circuit disruption. Circuit components have been designed and positioned to minimize the quantity of air that travels through the arterial l...

متن کامل

Propofol, midazolam, vancomycin and cyclosporine therapeutic drug monitoring in extracorporeal membrane oxygenation circuits primed with whole human blood

INTRODUCTION As a result of drug sequestration and increased volume of distribution, the extracorporeal membrane oxygenation (ECMO) procedure might lead to a decrease in drug concentrations during a patient's treatment. The aim of this study was to evaluate sedative, antibiotic and immunosuppressive drug loss in ECMO circuit using ex-vivo and in-vitro experiments. METHODS Blood concentrations...

متن کامل

Macro- and micronutrient disposition in an ex vivo model of extracorporeal membrane oxygenation

BACKGROUND Extracorporeal membrane oxygenation (ECMO) circuits have been shown to sequester circulating blood compounds such as drugs based on their physicochemical properties. This study aimed to describe the disposition of macro- and micronutrients in simulated ECMO circuits. METHODS Following baseline sampling, known quantities of macro- and micronutrients were injected post oxygenator int...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Intensive care medicine experimental

دوره 4 1  شماره 

صفحات  -

تاریخ انتشار 2016