Feasibility study of performing IGRT system daily QA using a commercial QA device
نویسندگان
چکیده
The purpose of this study was to investigate the feasibility of using a single QA device for comprehensive, efficient daily QA of a linear accelerator (Linac) and three image-guided stereotactic positioning systems (IGSPSs). The Sun Nuclear Daily QA 3 (DQA3) device was used to perform daily dosimetry and mechanical accuracy tests for an Elekta Linac, as well as daily image geometric and isocenter coincidence accuracy tests for three IGSPSs: the AlignRT surface imaging system; the frameless SonArray optical tracking System (FSA) and the Elekta kV CBCT. The DQA3 can also be used for couch positioning, repositioning, and rotational tests during the monthly QA. Based on phantom imaging, the Linac coordinate system determined using AlignRT was within 0.7 mm/0.6° of that of the CBCT system. The difference is attributable to the different calibration methods that are utilized for these two systems. The laser alignment was within 0.5 mm of the isocenter location determined with the three IGSPSs. The ODI constancy was ± 0.5 mm. For gantry and table angles of 0°, the mean isocenter displacement vectors determined using the three systems were within 0.7 mm and 0.6° of one another. Isocenter rotational offsets measured with the systems were all ≤ 0.5°. For photon and electron beams tested over a period of eight months, the output was verified to remain within 2%, energy variations were within 2%, and the symmetry and flatness were within 1%. The field size and light-radiation coincidence were within 1mm ± 1 mm. For dosimetry reproducibility, the standard deviation was within 0.2% for all tests and all energies, except for photon energy variation which was 0.6%. The total measurement time for all tasks took less than 15 minutes per QA session compared to 40 minutes with our previous procedure, which utilized an individual QA device for each IGSPS. The DQA3 can be used for accurate and efficient Linac and IGSPS daily QA. It shortens QA device setup time, eliminates errors introduced by changing phantoms to perform different tests, and streamlines the task of performing dosimetric checks.
منابع مشابه
2D linear array device as a quality assurance tool in brachytherapy applications
Background: External beam radiotherapy and brachytherapy plays a vital role in the management of cancer cervix. High dose rate brachytherapy is being presently used worldwide for the brachytherapy applications. At present, 2-Dimensional linear array detectors are the most common QA tool used for pretreatment patient specific quality assurance in external beam radiotherapy alon...
متن کاملDaily quality assurance phantom for ultrasound image guided radiation therapy
A simple phantom was designed, constructed, tested, and clinically implemented for daily quality assurance (QA) of an ultrasound-image-guided radiation therapy (US-IGRT) system, the Restitu Ultrasound system (Resonant Medical, Montreal, QC). The phantom consists of a high signal echogenic background gel surrounding a low signal hypoechoic egg-shaped target. Daily QA checks involve ultrasound im...
متن کاملA novel compound 6D‐offset simulating phantom and quality assurance program for stereotactic image‐guided radiation therapy system
A comprehensive quality assurance (QA) device cum program was developed for the commissioning and routine testing of the 6D IGRT systems. In this article, both the new QA system and the BrainLAB IGRT system which was added onto a Varian Clinac were evaluated. A novel compound 6D-offset simulating phantom was designed and fabricated in the Prince of Wales Hospital (PWH), Hong Kong. The QA progra...
متن کاملAutomated Quality Assurance for Image‐Guided Radiation Therapy
The use of image-guided patient positioning requires fast and reliable Quality Assurance (QA) methods to ensure the megavoltage (MV) treatment beam coincides with the integrated kilovoltage (kV) or volumetric cone-beam CT (CBCT) imaging and guidance systems. Current QA protocol is based on visually observing deviations of certain features in acquired kV in-room treatment images such as markers,...
متن کاملUse of TrueBeam developer mode for imaging QA
The purpose of this study was to automate regular Imaging QA procedures to become more efficient and accurate. Daily and monthly imaging QA for SRS and SBRT protocols were fully automated on a Varian linac. A three-step paradigm where the data are automatically acquired, processed, and analyzed was defined. XML scripts were written and used in developer mode in a TrueBeam linac to automatically...