Simultaneous measurement of regional cerebral blood flow by perfusion CT and stable xenon CT: a validation study.
نویسندگان
چکیده
BACKGROUND AND PURPOSE Knowledge of cerebral blood flow (CBF) alterations in cases of acute stroke could be valuable in the early management of these cases. Among imaging techniques affording evaluation of cerebral perfusion, perfusion CT studies involve sequential acquisition of cerebral CT sections obtained in an axial mode during the IV administration of iodinated contrast material. They are thus very easy to perform in emergency settings. Perfusion CT values of CBF have proved to be accurate in animals, and perfusion CT affords plausible values in humans. The purpose of this study was to validate perfusion CT studies of CBF by comparison with the results provided by stable xenon CT, which have been reported to be accurate, and to evaluate acquisition and processing modalities of CT data, notably the possible deconvolution methods and the selection of the reference artery. METHODS Twelve stable xenon CT and perfusion CT cerebral examinations were performed within an interval of a few minutes in patients with various cerebrovascular diseases. CBF maps were obtained from perfusion CT data by deconvolution using singular value decomposition and least mean square methods. The CBF were compared with the stable xenon CT results in multiple regions of interest through linear regression analysis and bilateral t tests for matched variables. RESULTS Linear regression analysis showed good correlation between perfusion CT and stable xenon CT CBF values (singular value decomposition method: R(2) = 0.79, slope = 0.87; least mean square method: R(2) = 0.67, slope = 0.83). Bilateral t tests for matched variables did not identify a significant difference between the two imaging methods (P >.1). Both deconvolution methods were equivalent (P >.1). The choice of the reference artery is a major concern and has a strong influence on the final perfusion CT CBF map. CONCLUSION Perfusion CT studies of CBF achieved with adequate acquisition parameters and processing lead to accurate and reliable results.
منابع مشابه
Vascular reserve in chronic cerebral ischemia measured by the acetazolamide challenge test: comparison with positron emission tomography.
PURPOSE To determine the value of the acetazolamide challenge test with stable xenon-enhanced CT (Xe CT) for making therapeutic decisions in patients with chronic cerebrovascular disease. METHODS We compared the Xe CT-measured acetazolamide response with various measures obtained by positron emission tomography. We performed both a positron emission tomographic scan and a Xe CT study in 11 pa...
متن کاملComparison of methods of attenuation and scatter correction in brain perfusion SPECT.
OBJECTIVE The radioactivity count distribution in the brain must be determined accurately to accurately measure cerebral blood flow (CBF). Scatter and attenuation are factors that compromise the accuracy of determining radioactivity counts in the brain. METHODS We compared regional CBF in patients by an autoradiographic method using N-isopropyl-p-[123I] iodoamphetamine when, first, attenuatio...
متن کاملThe in vivo autoradiographic measurement of regional cerebral blood flow using stable xenon and computerized tomography: the effect of tissue heterogeneity and computerized tomography noise.
We have studied the effect of brain tissue (gray matter-white matter) heterogeneity and computerized tomography (CT) noise on the accuracy of xenon CT measurements of regional cerebral blood flow (rCBF) based upon "autoradiographic" and multiple-scan washin protocols. The results of our mathematical analysis indicate that both protocols are associated with a variety of measurement errors that l...
متن کاملLocal cerebral blood flow measured by CT after stable xenon inhalation.
A safe, practical, clinically applicable, noninvasive method for measuring local cerebral blood flow (LCBF) using inert xenon (Xes) and a 60 sec CT scanner has been developed in the baboon. Direct measurement of expired Xes concentration after short inhalation (4--7 min) of 40% Xes and the use of computer-programmed autoradiographic formulas allow accurate, reproducible measurements of LCBF usi...
متن کاملComparison of xenon-enhanced CT with ultrafast CT for measurement of regional cerebral blood flow.
PURPOSE To compare an ultrafast CT method for estimating regional cerebral blood flow with a more commonly used xenon-enhanced CT method. METHODS Xenon CT and ultrafast CT were used to estimate regional cerebral blood flow (rCBF) in 12 healthy beagle dogs. Measurements were obtained for left and right hemisphere, cortical gray matter, basal ganglia, and deep white matter. The ability of each ...