Advanced System Components for the Development of a Handheld Ion Trap Mass Spectrometer
نویسندگان
چکیده
Kevin Philip Schultze: Advanced System Components for the Development of a Handheld Ion Trap Mass Spectrometer (Under the direction of J. Michael Ramsey) This work describes the development and implementation of system components utilized in a handheld ion trap mass spectrometer (MS). Current state-of-the-art portable mass spectrometers fall in the 30 to 40 lb. range due to their large and heavy vacuum systems. The strategy used in this work to create lighter, handheld instruments is to eliminate much of the vacuum pumping requirements by operating the entire instrument under 1 Torr of background pressure, i.e. 3 orders of magnitude higher than conventional instruments. Reducing the ion trap size by a factor of 20 and operating these miniature traps at increased RF drive frequencies compensates for the ions' reduced mean free path, making mass analysis possible at these elevated pressures. The first component developed was a novel ion trap geometry. The design is a modification of the cylindrical ion trap (CIT), using three electrodes to produce an elongated two-dimensional trapping region capped by planar mesh electrode features on each end. The SLIT's increased dimensionality led to a tenfold increase in sensitivity versus a CIT with the same critical dimensions while maintaining identical resolution. This 5 mm long SLIT with critical dimensions x o = 500 µm and z o = 650 µm represents the smallest linear type ion trap reported to date and maintains higher resolution than their larger counterparts. SLIT iv operation in pressures exceeding 1 Torr of helium, nitrogen, and air buffer gasses also represents the highest pressure operation reported for linear ion traps. Ion detection at elevated pressures was achieved using a Faraday cup and charge sensitive amplifier. Mass spectra were collected using both a commercial amplifier and two custom made amplifiers. The smaller of the two custom Faraday cup amplifiers was incorporated into a miniature vacuum chamber along with the SLIT mass analyzer and glow discharge ion source. The operation of this MS system is the first reported miniature instrument operated without the use of turbomolecular vacuum pumps, and can in fact be operated with only a simple miniaturized roughing pump. Incorporating the miniature chamber, pump, and a custom miniature RF system, a prototype MS was developed weighing 2 lbs. and drawing 8.2 W. Finally, gas chromatography-mass spectrometry (GC-MS) experiments were carried out utilizing the newly developed miniature MS instrumentation with a full size GC to demonstrate …
منابع مشابه
Multistage Fragmentation of Ion Trap Mass Spectrometry System and Pseudo-MS3 of Triple Quadrupole Mass Spectrometry Characterize Certain (E)-3-(Dimethylamino)-1-arylprop-2-en-1-ones: A Comparative Study
A new approach was recently introduced to improve the structure elucidation power of tandem mass spectrometry simulating the MS(3) of ion trap mass spectrometry system overcoming the different drawbacks of the latter. The fact that collision induced dissociation in the triple quadrupole mass spectrometer system provides richer fragment ions compared to those achieved in the ion trap mass spectr...
متن کاملA Rapid, Simple, Liquid Chromatographic-Electrospray Ionization, Ion Trap Mass Spectrometry Method for the Determination of Finasteride in Human Plasma and its Application to Pharmacokinetic Study
A fast, accurate, sensitive, selective and reliable method using reversed-phase high performance liquid chromatography coupled to electrospray ionization ion trap mass spectrometry was developed and validated for the determination of finasteride in human plasma. After protein precipitation with perchloric acid, satisfactory separation was achieved on a Zorbax Eclipse® C8 analytical column u...
متن کاملA Rapid, Simple, Liquid Chromatographic-Electrospray Ionization, Ion Trap Mass Spectrometry Method for the Determination of Finasteride in Human Plasma and its Application to Pharmacokinetic Study
A fast, accurate, sensitive, selective and reliable method using reversed-phase high performance liquid chromatography coupled to electrospray ionization ion trap mass spectrometry was developed and validated for the determination of finasteride in human plasma. After protein precipitation with perchloric acid, satisfactory separation was achieved on a Zorbax Eclipse® C8 analytical column u...
متن کاملRapid protein identification using a microscale electrospray LC/MS system on an ion trap mass spectrometer.
A methodology has been developed for the rapid identification of gel separated proteins. Following in gel protein digestion with trypsin, the resulting peptide mixture is analyzed by on-line liquid chromatography, electrospray mass spectrometry (LC/MS). The mass spectral data containing either accurate mass values or sequence specific fragment ion information is then matched to a database of kn...
متن کاملDevelopment of a capillary high-performance liquid chromatography tandem mass spectrometry system using SWIFT technology in an ion trap/reflectron time-of-flight mass spectrometer.
In this work a capillary high-performance liquid chromatography/tandem mass spectrometry system has been developed in a quadrupole ion trap/reflectron time-of-flight mass spectrometer. The broadband excitation for isolating a precursor ion is achieved by generating the stored waveform inverse Fourier transform waveform in real time. Following the isolation and cooling periods, an auxiliary AC v...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره شماره
صفحات -
تاریخ انتشار 2014