Reliability of cone counts using an adaptive optics retinal camera.

نویسندگان

  • Mélanie Bidaut Garnier
  • Mathieu Flores
  • Guillaume Debellemanière
  • Marc Puyraveau
  • Perle Tumahai
  • Mathieu Meillat
  • Claire Schwartz
  • Michel Montard
  • Bernard Delbosc
  • Maher Saleh
چکیده

BACKGROUND To assess the reproducibility and repeatability of cone imaging in healthy human eyes, using the RTx-1 Adaptive Optics Retinal Camera and its proprietary cone-counting software. DESIGN Single-centre, prospective study. PARTICIPANTS Ten healthy adults. METHODS Macular cones were imaged. Intrasession repeatability was assessed by comparing 10 consecutive acquisitions obtained by the same operator from each subject. For the intersession study, each subject was imaged five consecutive days. Interoperator reproducibility was also evaluated by comparing the images obtained from 10 different subjects by two independent operators. Finally, intergrader agreement was evaluated by comparing the cone counts measured by two masked graders. MAIN OUTCOME MEASURES Mean cone density (cells/mm(2) ), spacing between cells (μm) and percentage of cones with six neighbours calculated on Voronoi diagrams were measured. Correlation coefficients, intraclass correlation coefficients, and coefficients of variation were calculated. RESULTS Correlation coefficient and intraclass correlation coefficient were respectively 0.81 and 0.96 between operators, and 0.97 and 0.98 between the two graders. The intrasession and intersession coefficients of variation were under 7%. The percentage of cells with six neighbours and the spacing between cones varied in the same proportion (coefficients of variation ranged from 1.66 to 10.05%). CONCLUSIONS Overall, the test-retest variability of RTx-1 and its software was good in normal human eyes. Further studies in the normal clinical setting are mandatory.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Cone photoreceptor definition on adaptive optics retinal imaging

AIMS To quantitatively analyse cone photoreceptor matrices on images captured on an adaptive optics (AO) camera and assess their correlation to well-established parameters in the retinal histology literature. METHODS High resolution retinal images were acquired from 10 healthy subjects, aged 20-35 years old, using an AO camera (rtx1, Imagine Eyes, France). Left eye images were captured at 5° ...

متن کامل

Agreement in Cone Density Derived from Gaze-Directed Single Images Versus Wide-Field Montage Using Adaptive Optics Flood Illumination Ophthalmoscopy

Purpose We compared cone density measurements derived from the center of gaze-directed single images with reconstructed wide-field montages using the rtx1 adaptive optics (AO) retinal camera. Methods A total of 29 eyes from 29 healthy subjects were imaged with the rtx1 camera. Of 20 overlapping AO images acquired, 12 (at 3.2°, 5°, and 7°) were used for calculating gaze-directed cone densities...

متن کامل

Imaging cone photoreceptors in three dimensions and in time using ultrahigh resolution optical coherence tomography with adaptive optics

Cone photoreceptors in the living human eye have recently been imaged with micron-scale resolution in all three spatial dimensions using adaptive optics optical coherence tomography. While these advances have allowed non-invasive study of the three-dimensional structure of living human cones, studies of their function and physiology are still hampered by the difficulties to monitor the same cel...

متن کامل

Axial length and cone density as assessed with adaptive optics in myopia

AIM To assess the variations in cone mosaic in myopia and its correlation with axial length (AL). SUBJECTS AND METHODS Twenty-five healthy myopic volunteers underwent assessment of photoreceptors using adaptive optics retinal camera at 2° and 3° from the foveal center in four quadrants superior, inferior, temporal and nasal. Data was analyzed using SPSS version 17 (IBM). Multivariable regress...

متن کامل

Variability in parafoveal cone mosaic in normal trichromatic individuals

Parafoveal function is important for daily visual tasks such as reading. Here the variability in cone density along the four cardinal meridians in parafoveal regions of the retina was investigated in vivo using an adaptive optics fundus camera. Ten healthy normal trichromatic individuals were included in the study. There were significant differences in cone density between individuals at all fo...

متن کامل