Localization error analysis for stereo X-ray image guidance with probability method.

نویسندگان

  • H Jiang
  • W R Chen
  • G Wang
  • H Liu
چکیده

The mean value and standard deviation of localization error for the stereo imaging systems are derived based on probability theory. Compared with the maximum error analysis method used in our previous study, the new approach yields more informative and precise results as the guidance for X-ray imaging system design and protocol optimization. The prototype for our current study is a CCD based monoplane digital stereo X-ray imaging system. The imaging model consists of two X-ray sources and one detector plane. With perspective geometry, the least-square solution is derived to reconstruct 3-dimensional object points, such as a biopsy needle tip, from a pair of 2-dimensional digital radiographs. Under the conditions of our specific prototype, the measurement errors of interested points in the radiographs are modeled as random variables with Gaussian distribution. Such variables account for finite image system noise and positioning errors. Then, the 3D localization error, in terms of mean value and standard deviation, is formulated using measurement error, feature point location, and separation between the two X-ray sources and distance from source to detector. Both theoretical analysis and numerical simulation are performed. The mean value and standard deviation of the localization error are first evaluated using numerical simulation under practical imaging conditions. Then, the error estimates are given in simply analytic forms. Simulation and theoretical results are in excellent agreement. The results show that our prototype X-ray stereological imaging system is accurate and reliable to locate feature points in 3D for medical intervention. Imaging protocols can be effectively optimized through the 3D localization error analysis using the approximate formulas proposed in this study.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A localization algorithm and error analysis for stereo x-ray image guidance.

Stereo x-ray radiography attracts increasing attention in major clinical applications. The purpose of this paper is to analyze the 3D localization error for breast biopsy procedures and provide guidelines for improving its accuracy. Our prototype is a CCD based digital stereo x-ray imaging system. The mathematical model consists of two x-ray sources and one stationary detector plane. A closed f...

متن کامل

2D/3D image fusion for accurate target localization and evaluation of a mask based stereotactic system in fractionated stereotactic radiotherapy of cranial lesions.

The purpose of this study was to evaluate the accuracy of a two-dimensional (2D) to three-dimensional (3D) image-fusion-guided target localization system and a mask based stereotactic system for fractionated stereotactic radiotherapy (FSRT) of cranial lesions. A commercial x-ray image guidance system originally developed for extracranial radiosurgery was used for FSRT of cranial lesions. The lo...

متن کامل

Innovative Methodology High-Resolution Three-Dimensional Microelectrode Brain Mapping Using Stereo Microfocal X-ray Imaging

Cox DD, Papanastassiou AM, Oreper D, Andken BB, DiCarlo JJ. High-resolution three-dimensional microelectrode brain mapping using stereo microfocal X-ray imaging. J Neurophysiol 100: 2966–2976, 2008. First published September 24, 2008; doi:10.1152/jn.90672.2008. Much of our knowledge of brain function has been gleaned from studies using microelectrodes to characterize the response properties of ...

متن کامل

Image quality enhancement in digital panoramic radiograph

One of the most common positioning errors in panoramic radiography is palatoglossal air space above the apices of the root of maxillary teeth. It causes a radiolucency obscuring the apices of maxillary teeth. In the case of this positioning error, the imaging should be repeated. This causes the patient be exposed to radiation again. To avoid the repetition of exposing harmful X-rays to the pati...

متن کامل

Empirical correlation for porosity deduction from X-ray computed tomography (CT)

For obtaining reservoir petrophysical properties, for example porosity, non-destructive methods such as X-ray computed tomography, CT, seems to be precise and accurate. Porosity is deducted from the CT image with a single scan via different techniques, such as pore space detection by image segmentation techniques then correlation with porosity. More than one hundred samples with carbonate li...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Medical engineering & physics

دوره 23 8  شماره 

صفحات  -

تاریخ انتشار 2001