Verification of dynamic and segmental IMRT delivery by dynamic log file analysis

نویسندگان

  • Dale W. Litzenberg
  • Jean M. Moran
  • Benedick A. Fraass
چکیده

A program has been developed to evaluate the delivered fluence of step-and-shoot segmental and sliding window dynamic multileaf collimator (MLC) fields. To automate these checks, a number of tools have been developed using data available from the dynamic log files that can be created each time a dynamic delivery occurs. Experiments were performed with a Varian 2100EX with a 120 leaf MLC equipped with dynamic capabilities. A dynamic leaf sequence is delivered and measured with film or an amorphous silicon imager. After delivery, the dynamic log file is written by the accelerator control system. The file reports the expected and actual position for each leaf and the dose fraction every 0.055 seconds. Leaf trajectories are calculated from this data and expected and actual fluence images are created from the difference of opposing leaf trajectories. These images can be compared with the expected delivery, measurements, and calculations of fluence. Tools have been developed to investigate other aspects of the delivery, such as specific leaf errors, beam hold-off flags sent by the control system to the MLC, and gap widths. This program is part of a semi-automated quality assurance (QA) system for pretreatment fluence verification and daily treatment verification of dynamic multileaf collimation (DMLC) delivery.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

[Testing of dynamic multileaf collimator by dynamic log file].

Intensity-modulated radiation therapy (IMRT) represents one of the most significant technical advances in radiation therapy. In the dynamic multileaf collimator (MLC) method of IMRT delivery, because of the relatively small gaps between opposed leaves and because most regions are shielded by leaves most of the time, the delivered dose is very sensitive to MLC leaf positional accuracy. A variati...

متن کامل

QA of intensity-modulated beams using dynamic MLC log files

To evaluate the utility of Dynalog file information for planar dose verification in IMRT QA, a program is developed to convert Dynalog file data to DMLC field files. For this study, five predefined fluencies are planned and delivered using Varian, Eclipse 3D planning system and 6MV photon beam of Varian, Clinac DMX linear accelerator. To measure planar dose distribution, Kodak, EDR2 films are e...

متن کامل

Maximum MLC opening effect in dynamic delivery of IMRT: leaf‐positional analysis

The analysis of dynamic multileaf collimator (MLC) positions for the delivered intensity-modulated radiotherapy (IMRT) plans is crucial in that it may capture dose delivery problems otherwise difficult to observe and quantify in the conventional dosimetric measurements with film or with an ionization chamber. In some IMRT systems, delivery of IMRT fields starts with a maximum MLC opening (rough...

متن کامل

Inverse Verification of the Dose Distribution for Intensity Modulated Radiation Therapy Patient-specific Quality Assurance Using Dynamic MLC Log Files

The aim of this study was to investigate a novel method for verification of the dose distribution for intensity modulated radiation therapy (IMRT) patient-specific quality assurance (QA) using dynamic multi-leaf collimator (DMLC) log files (Dynalog files). Dynalog files are recorded every 50 ms by using a MLC controller during the IMRT treatment. Dynalog files contain actual MLC positional info...

متن کامل

Commissioning and quality assurance for intensity‐modulated radiotherapy with dynamic multileaf collimator: Experience of the Pontificia Universidad Católica de Chile

The objective of this paper is to present our experience in the commissioning and quality assurance (QA) for intensity modulated radiotherapy (IMRT) using dynamic multileaf collimator (dMLC), sliding window technique. Using Varian equipment solution, the connectivity and operation between all IMRT chain components was checked. Then the following test were done: stability of leaf positioning and...