Assessment of Spectral Doppler for an Array-Based Preclinical Ultrasound Scanner Using a Rotating Phantom

نویسندگان

  • David A. Kenwright
  • Tom Anderson
  • Carmel M. Moran
  • Peter R. Hoskins
چکیده

Velocity measurement errors were investigated for an array-based preclinical ultrasound scanner (Vevo 2100, FUJIFILM VisualSonics, Toronto, ON, Canada). Using a small-size rotating phantom made from a tissue-mimicking material, errors in pulse-wave Doppler maximum velocity measurements were observed. The extent of these errors was dependent on the Doppler angle, gate length, gate depth, gate horizontal placement and phantom velocity. Errors were observed to be up to 172% at high beam-target angles. It was found that small gate lengths resulted in larger velocity errors than large gate lengths, a phenomenon that has not previously been reported (e.g., for a beam-target angle of 0°, the error was 27.8% with a 0.2-mm gate length and 5.4% with a 0.98-mm gate length). The error in the velocity measurement with sample volume depth changed depending on the operating frequency of the probe. Some edge effects were observed in the horizontal placement of the sample volume, indicating a change in the array aperture size. The error in the velocity measurements increased with increased phantom velocity, from 22% at 2.4 cm/s to 30% at 26.6 cm/s. To minimise the impact of these errors, an angle-dependent correction factor was derived based on a simple ray model of geometric spectral broadening. Use of this angle-dependent correction factor reduces the maximum velocity measurement errors to <25% in all instances, significantly improving the current estimation of maximum velocity from pulse-wave Doppler ultrasound.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Testing of Doppler Ultrasound Systems

Blood and tissue velocities are measured and analyzed in cardiac, vascular, and other applications of diagnostic ultrasound. Errors in system performance might give invalid measurements. We developed two moving string test targets and a rotating cylinder phantom (Doppler phantoms) to characterize Doppler ultrasound systems. These phantoms were initially used to measure such variables as sample ...

متن کامل

Assessment of X-Ray Crosstalk in a Computed Tomography Scanner with Small Detector Elements Using Monte Carlo Method

Introduction: Crosstalk is a leakage of X-ray or light produced in a matrix of X-ray detectors or array of photodiodes in one element to other elements affecting on image contrast and spatial resolution. In this study, we assessed X-ray crosstalk in a computed tomography (CT) scanner with small detector elements to estimate the effect of various parameters such as X-ray tube voltage, detector e...

متن کامل

Wall-less Flow Phantom for High-Frequency Ultrasound Applications

There are currently very few test objects suitable for high-frequency ultrasound scanners that can be rapidly manufactured, have appropriate acoustic characteristics and are suitably robust. Here we describe techniques for the creation of a wall-less flow phantom using a physically robust konjac and carrageenan-based tissue-mimicking material. Vessel dimensions equivalent to those of mouse and ...

متن کامل

اندازه‌گیری شاخصهای طبیعی در سونوگرافی داپلر شریان سلیاک در افراد بالای 20 سال

Doppler ultrasound is one of the non-invasive methods for blood flow assessment in cardio-vascular system and celiac trunk as the first branch of abdominal aorta is an important splanchnic artery. According to the relative differences in the reports of normal indices from doppler ultrasound of the celiac artery in the different medical centers of the world, it was decided to evaluate this issue...

متن کامل

3d Sonoelastography for Prostate Tumor Imaging

Sonoelastography is a new imaging technique to detect hard tumors in soft tissue. In sonoelastography imaging low frequency (200-500 Hz) shear waves are propagated through a tissue sample while real time Doppler techniques are used to image the resulting vibration pattern on an ultrasound scanner. A sonoelastography image is therefore a mapping of the relative vibration amplitude of the tissue ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 41  شماره 

صفحات  -

تاریخ انتشار 2015