Isolation of polyacetal wear particles from periprosthetic tissue of isoelastic femoral stems.

نویسندگان

  • A Minovic
  • I Milowsev
  • V Pisot
  • A Cör
  • V Antolic
چکیده

We analysed revised Mathys isoelastic polyacetal femoral stems with stainless-steel heads and polyethylene acetabular cups from eight patients in order to differentiate various types of particle of wear debris. Loosening of isoelastic femoral stems is associated with the formation of polyacetal wear particles as well as those of polyethylene and metal. All three types of particle were isolated simultaneously by tissue digestion followed by sucrose gradient centrifugation. Polyacetal particles were either elongated, ranging from 10 to 150 microm in size, or shred-like and up to 100 microm in size. Polyethylene particles were elongated or granules, and were typically submicron or micron-sized. Polyacetal and polyethylene polymer particles were differentiated by the presence of BaSO4, which is added as a radiopaque agent to polyacetal but not to polyethylene. This was easily detectable by back-scattered SEM analysis and verified by energy dispersive x-ray analysis. Two types of foreign-body giant cell (FBGC) were recognised in the histological specimens. Extremely large FBGCs with irregular polygonal particles showing an uneven, spotty birefringence in polarised light were ascribed to polyacetal debris. Smaller FBGCs with slender elongated particles shining uniformly brightly in polarisation were related to polyethylene. Mononucleated histiocytes containing both types of particle were also present. Our findings offer a better understanding of the processes involved in the loosening of polyacetal stems and indicate why the idea of 'isoelasticity' proved to be unsuccessful in clinical practice.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Extensive metallosis and necrosis in failed prostheses with cemented titanium-alloy stems and ceramic heads.

We describe three prostheses with cemented titanium-alloy stems and Al2O3 ceramic femoral heads which had to be revised after a mean period of implantation of 78 months. In each case, the neck of the prosthesis had been so severely worn that the profile was elliptical rather than circular. There was severe metallosis of the periprosthetic tissues. Metal particles isolated from the tissues were ...

متن کامل

Wear particles and ions from cemented and uncemented titanium-based hip prostheses—A histological and chemical analysis of retrieval material

Wear debris-induced inflammation is considered to be the main cause for periprosthetic osteolysis in total hip replacements (THR). The objective of this retrieval study was to examine the tissue reactions and exposure to metal ions and wear particles in periprosthetic tissues and blood samples from patients with titanium (Ti)-based hip prostheses that were revised due to wear, osteolysis, and/o...

متن کامل

Eleven-year follow-up of second-generation metal-on-metal total hip arthroplasty.

PURPOSE To investigate the cause of failure in matte-surface cemented stems in second-generation metal-on- metal total hip arthroplasty (THA). METHODS Records of 26 men and 11 women (39 hips) aged 29 to 72 years who underwent primary cementless THAs by a single surgeon using second-generation metal-on-metal prostheses and were followed up for a mean of 122 (range, 120-141) months were reviewe...

متن کامل

Wear Debris Characterization and Corresponding Biological Response: Artificial Hip and Knee Joints

Wear debris, of deferent sizes, shapes and quantities, generated in artificial hip and knees is largely confined to the bone and joint interface. This debris interacts with periprosthetic tissue and may cause aseptic loosening. The purpose of this review is to summarize and collate findings of the recent demonstrations on debris characterization and their biological response that influences the...

متن کامل

Extensively porous-coated stems for femoral revision: reliable choice for stem revision in Paprosky femoral type III defects.

Between January 1999 and August 2008, ninety-six femoral revisions were performed with extensively porous-coated stems in Paprosky type III femoral defects (89 type IIIA and 7 type IIIB defects). Seven type IIIB defects with a mean canal of 16.5 mm were observed; 6 defects achieved stable bone ingrowth and 1 achieved stable fibrous condition. Average postoperative Harris Hip Score was 92.3±8 (r...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • The Journal of bone and joint surgery. British volume

دوره 83 8  شماره 

صفحات  -

تاریخ انتشار 2001