Errors generated by a point-of-care CD4+ T-lymphocyte analyser: a retrospective observational study in nine countries

نویسندگان

  • Emmanuel Fajardo
  • Carol Metcalf
  • Erwan Piriou
  • Monique Gueguen
  • David Maman
  • Pascale Chaillet
  • Vivian Cox
  • Maryam B Rumaney
  • Syanness Tunggal
  • Cara Kosack
  • Teri Roberts
چکیده

OBJECTIVE To estimate the proportion of invalid results generated by a CD4+ T-lymphocyte analyser used by Médecins Sans Frontières (MSF) in field projects and identify factors associated with invalid results. METHODS We collated 25,616 CD4+ T-lymphocyte test results from 39 sites in nine countries for the years 2011 to 2013. Information about the setting, user, training, sampling technique and device repair history were obtained by questionnaire. The analyser performs a series of checks to ensure that all steps of the analysis are completed successfully; if not, an invalid result is reported. We calculated the proportion of invalid results by device and by operator. Regression analyses were used to investigate factors associated with invalid results. FINDINGS There were 3354 invalid test results (13.1%) across 39 sites, for 58 Alere PimaTM devices and 180 operators. The median proportion of errors per device and operator was 12.7% (interquartile range, IQR: 10.3-19.9) and 12.1% (IQR: 7.1-19.2), respectively. The proportion of invalid results varied widely by country, setting, user and device. Errors were not associated with settings, user experience or the number of users per device. Tests performed on capillary blood samples were significantly less likely to generate errors compared to venous whole blood. CONCLUSION The Alere Pima CD4+ analyser generated a high proportion of invalid test results, across different countries, settings and users. Most error codes could be attributed to the operator, but the exact causes proved difficult to identify. Invalid results need to be factored into the implementation and operational costs of routine CD4+ T-lymphocyte testing.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

B and T Lymphocyte Attenuator is a Target of miR-155 during Naive CD4+ T Cell Activation

Background: MicroRNA-155 (miR-155) is upregulated during T cell activation, but the exact mechanisms by which it influences CD4+ T cell activation remain unclear. Objective: To examine whether the B and T lymphocyte attenuator (BTLA) is a target of miR-155 during naïve CD4+ T cell activation. Methods: Firefly luciferase reporter plasmids pEZX-MT01-wild-type-BTLA and pEZX-MT01-mutant-BTLA were ...

متن کامل

O 1: The Effects of Vitamin D Supplementation on the T cell Compartment in Multiple Sclerosis

Multiple sclerosis (MS) is a complex neurological disease and its prevalence is about 2 million in the world. Neuroinflammation plays a key role in MS. Vitamins are essential nutrients that have effective role on immune system including activation of lymphocyte and differentiation of T-helper cell. Vitamin D is a micronutrient that is effective on immune function. Deficiently of Vitamin D is a ...

متن کامل

An Observational Study of Errors Related to the Preparation and Administration of Medications Given by Infusion Devices in a Teaching Hospital

Since there is no detailed hospital based incident reporting system, this study was designed to evaluate the medication errors associated with infusion pumps in intensive care unit (ICU). The investigation was conducted in a Teaching hospital in the form of a prospective, observational study. A sample size of 43 doses administered to ICU patients was chosen to enable reliable estimate of error ...

متن کامل

Medication Errors in an Emergency Department in a Large Teaching Hospital in Tehran

Medication errors have important effects on increased length of hospitalization, increased mortality and costs. We assessed the incidence of medication errors and characterize the error types in an emergency department in a large teaching hospital in Tehran. We also investigated the effect of Emergency Department pharmacists on patient safety with regard to recovery of potentially harmful m...

متن کامل

An Observational Study of Errors Related to the Preparation and Administration of Medications Given by Infusion Devices in a Teaching Hospital

Since there is no detailed hospital based incident reporting system, this study was designed to evaluate the medication errors associated with infusion pumps in intensive care unit (ICU). The investigation was conducted in a Teaching hospital in the form of a prospective, observational study. A sample size of 43 doses administered to ICU patients was chosen to enable reliable estimate of error ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 93  شماره 

صفحات  -

تاریخ انتشار 2015