Dose-area product measurements in panoramic dental radiology.

نویسندگان

  • B Poppe
  • H K Looe
  • A Pfaffenberger
  • N Chofor
  • F Eenboom
  • M Sering
  • A Rühmann
  • A Poplawski
  • K Willborn
چکیده

In this study, dose-area product (DAP) measurements in panoramic dental radiology have been performed in Germany. The results obtained in this study were proposed as diagnostic reference levels (DRLs). A representative number of dental panoramic units, both with digital and conventional image receptors, have been chosen. Common statistical parameters such as mean, standard deviation and 3rd quartile have been calculated. For four different standard programmes, 'large adult', 'adult male', 'adult female' and 'child', the proposed DRLs are 101, 87, 84 and 75 mGy cm(2), respectively. No clear tendency to a generalised dose reduction from the transition to digital techniques has been observed. Effective doses have been calculated from E/DAP conversion factors published in literature. Even though these values differ by a factor of approximately 3, upper limits of 15.8-21.2 microSv for the four different exposure settings were derived from the data.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of radiation-induced cancer risk to patients undergoing intra-oral and panoramic dental radiographies using experimental measurements and Monte Carlo calculations

Background: Radiation doses associated with the conventional dental radiographies are relatively low, but their number is high. Therefore, justification is necessary to ensure that radiation doses to patients, particularly children, are kept as low as reasonably achievable. Materials and Methods: The exposure factors applied for real patients in four age groups (5, 10, 15-year-old and adult) we...

متن کامل

Diagnostic reference levels at intraoral and dental panoramic examinations

Background: The aim of this study was to measure and determine diagnostic reference levels (DRLs) of programs used at intraoral and dent panoramic examinations in Kosovo. Materials and Methods: To determine DRL, were performed measurements for dose area product (DAP) at intraoral and dental panoramic radiology systems. This study has covered over 47% of X-ray units, which are in use for intraor...

متن کامل

Measurement of dose in panoramic dental radiology.

The National Radiological Protection Board (NRPB) has recommended the introduction of dose-width product (DWP) for the measurement of patient dose in panoramic dental radiology and has proposed a reference level of 65 mGy mm for adult exposures. This paper describes a method for measuring DWP and dose-area product (DAP) using thermoluminescent dosemeters (TLDs). The technique was used on 16 set...

متن کامل

Comparison of ionisation chamber and semiconductor detector devices for measurement of the dose-width product for panoramic dental units.

Doses for panoramic dental radiography are assessed in terms of the dose-width product (DWP) or dose-area product, which gives a measure of the radiation through a whole exposure. The DWP can be measured using a pencil ionisation chamber (IC) similar to that used for computed tomography dose assessment. However, ICs are sensitive to radiation incident from all directions and so backscatter from...

متن کامل

Effect of Head Misalignment in Horizontal & Vertical Dimensions on the Magnification Rate of Digital Panoramic Radiography in Different Parts of the Jaws

Background and aim: Considering the widespread development of implants in dental treatment plans, linear measurements on panoramic radiography are of especial importance. In the present study we investigated the effect of head misalignment up to 15° around vertical and horizontal axes on the magnification rate of digital panoramic radiography in each part of upper and lower jaws. Materia...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Radiation protection dosimetry

دوره 123 1  شماره 

صفحات  -

تاریخ انتشار 2007