[Results and complications of pertrochanteric hip fractures using an intramedullary nail with a helical blade (proximal femoral nail antirotation) in 200 patients].
نویسندگان
چکیده
OBJECTIVES Evaluation of the surgical management, outcome and complications in patients with pertrochanteric fractures treated with PFNA nail. MATERIAL AND METHOD A retrospective study was conducted on 200 patients treated consecutively between April 2010 and February 2012. Radiological assessments were performed before and after the surgery, and during the follow-up (fracture reduction, blade position, consolidation or collapse signs). A clinical evaluation was performed as regards walking capabilities. The results were compared with those of a previous study on 700 patients treated with gamma 3 and TFN nails. RESULTS The blade position was centre-centre in 64% of patients, and decreased to 53% in the mechanical complications group. Tip-apex distance was less than 25mm in 91.5%. The average hospital stay was 9.17 days, with a mean post-surgery stay of 5.95 days. Complications (7.5%): 2 cut out (1%), one cut through (0.5%), 4 cases of helical blade sliding (2%), one failure in distal locking procedure (0.5%), 2 cases with painful fasciae latae (1%), one union delay (0.5%), 2 cases of non-union with hardware failure (1%), one case of intense bleeding related to distal locking of the nail (0.5%), and one case of avascular necrosis (0.5%). CONCLUSIONS The PFNA helical blade system seems to reduce the incidence of cut out and cut through in osteoporotic bone. Blade position was one of the main parameters associated with mechanical complications.
منابع مشابه
Reason and treatment of failure of proximal femoral nail antirotation internal fixation for femoral intertrochanteric fractures of senile patients.
The cause of postoperative failure after the treatment of femoral intertrochanteric fracture with proximal femoral nail antirotation (PFNA) was analyzed, and the reoperative methods were examined. Nine failures of 308 femoral intertrochanteric fracture patients with PFNA were treated with femoral prosthesis total hip replacement (THR) and reoperative internal fixation. All nine patients were an...
متن کاملBackout of the helical blade of proximal femoral nail antirotation and accompanying fracture nonunion.
This article describes a case of backout of the helical blade, a rare complication of proximal femoral nail antirotation. A 31-year-old man had sustained a trochanteric fracture of his right femur. Fracture fixation using proximal femoral nail antirotation and autologous bone grafting 7 months later were performed at another hospital. However, bony union was not obtained, and the patient's pain...
متن کاملFailure of PFNA: helical blade perforation and tip-apex distance.
Since the clinical introduction of proximal femoral nail antirotation (PFNA) by AO group in 2004 and PFNA-II (Asian version) in 2008, this type of cephalomedullary nail with single head-neck helical blade were commonly used in unstable pertrochanteric and intertrochanteric fractures in geriatric patients. Recently, severe complications of helical blade perforation through the femoral head, into...
متن کاملTreatment of Proximal Femur Osteomyelitis Occurred after Proximal Femoral Nail Antirotation Fixation, with Antibiotic Cement-coated Tibia Intramedullary Nail: A Case Report
Antibiotic cement-coated intramedullary nails maintain a locally high antibiotic concentration while contributing to bone stability. We present a case of femoral subtrochanteric fracture in a patient with an infected nonunion who was successfully treated for an infection and nonunion using an antibiotic cement-coated tibial intramedullary nail. A 79-year-old woman with a right femoral subtrocha...
متن کاملShould the tip-apex distance (TAD) rule be modified for the proximal femoral nail antirotation (PFNA)? A retrospective study
BACKGROUND Unstable proximal femoral fractures are common and challenging for the orthopaedic surgeon. Often, these are treated with intramedullary nails. The most common mode of failure of any device to treat these fractures is cut-out. The Synthes proximal femoral nail antirotation (PFNA) is unique because it is the only proximal femoral intramedullary nail which employs a helical blade in li...