Measurement of Intrahepatic Pressure during Microwave Ablation in an Ex Vivo Bovine Liver Model

نویسندگان

  • Hae Jin Kim
  • Hyunchul Rhim
  • Min Woo Lee
  • Woo Kyoung Jeong
چکیده

BACKGROUND/AIMS We experimented with different ablation methods and two types of microwave antennas to determine whether microwave ablation (MWA) increases intrahepatic pressure and to identify an MWA protocol that avoids increasing intrahepatic pressure. METHODS MWA was performed using either a single-step standard ablation or a stepwise increment ablation paired with either a 16-gauge (G) 2-cm antenna or a 14G 4-cm antenna. We compared the maximum pressures and total ablation volumes. RESULTS The mean maximum intrahepatic pressures and ablation volumes were as follows 16G single-step 37±33.4 mm Hg and 4.63 cm(3); 16G multistep 31±18.7 mm Hg and 3.75 cm(3); 14G single-step 114±45.4 mm Hg and 15.33 cm(3); and 14G multistep 106±43.8 mm Hg and 10.98 cm(3). The intrahepatic pressure rose during MWA, but there were no statistically significant differences between the single and multistep methods when the same gauge antennae were used. The total ablation volume was different only in the 14G groups (p<0.05). CONCLUSIONS We demonstrated an increase in intrahepatic pressure during MWA. The multistep method may be used to prevent increased intrahepatic pressure after applying the proper power.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

An Investigation of Brightness Changes in Ultrasound Images due to Temperature Variations in Liver Tissue during Radiofrequency Ablation

Introduction: one of the minimally invasive methods of treatmenting liver malignancies is Radiofrequency Ablation (RFA) which can be applied to primary or secondary tumors. Nevertheless, the disease recurrence is probable after RFA application partially due to the inadequate capability of temperature monitoring of the target tissue and visualizing the thermal damage. The goal of this work was t...

متن کامل

Intra-Tissue Pressure Measurement in Ex Vivo Liver Undergoing Laser Ablation with Fiber-Optic Fabry-Perot Probe

We report the first-ever intra-tissue pressure measurement performed during 1064 nm laser ablation (LA) of an ex vivo porcine liver. Pressure detection has been performed with a biocompatible, all-glass, temperature-insensitive Extrinsic Fabry-Perot Interferometry (EFPI) miniature probe; the proposed methodology mimics in-vivo treatment. Four experiments have been performed, positioning the pro...

متن کامل

Moving-Shot versus Fixed Electrode Techniques for Radiofrequency Ablation: Comparison in an Ex-Vivo Bovine Liver Tissue Model

OBJECTIVE To compare the ablation characteristics of the moving-shot technique (MST) and the fixed electrode technique (FET) for radiofrequency (RF) ablation in an ex-vivo bovine liver tissue model. MATERIALS AND METHODS We performed RF ablation using FET in 110 bovine liver blocks using 11 different ablation times ranging from 5 seconds to 5 minutes (10 blocks per each time duration). Ten bo...

متن کامل

Fluid preinjection for microwave ablation in an ex vivo bovine liver model assessed with volumetry in an open MRI system.

PURPOSE We aimed to detect possible differences in microwave ablation (MWA) volumes after different fluid preinjections using magnetic resonance imaging (MRI). MATERIALS AND METHODS MWA volumes were created in 50 cuboid ex vivo bovine liver specimens (five series: control [no injection], 10 mL water, 10 mL 0.9% NaCl, 10 mL 6% NaCl, and 10 mL 12% NaCl preinjections; n=10 for each series). The ...

متن کامل

Fiber-optic combined FPI/FBG sensors for monitoring of radiofrequency thermal ablation of liver tumors: ex vivo experiments.

We present a biocompatible, all-glass, 0.2 mm diameter, fiber-optic probe that combines an extrinsic Fabry-Perot interferometry and a proximal fiber Bragg grating sensor; the probe enables dual pressure and temperature measurement on an active 4 mm length, with 40 Pa and 0.2°C nominal accuracy. The sensing system has been applied to monitor online the radiofrequency thermal ablation of tumors i...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 9  شماره 

صفحات  -

تاریخ انتشار 2015