Pulmonary cement embolism after augmentation of pedicle screws with bone cement.

نویسندگان

  • Marc Röllinghoff
  • Jan Siewe
  • Peer Eysel
  • Karl-Stefan Delank
چکیده

The authors present the very rare complication of widespread pulmonary artery cement embolism after pedicle screw augmentation with bone cement for multilevel spine fusion. A 69-year-old woman with severe osteoporosis underwent a posterior T12S1fusion because of lumbar scoliosis. After two months the superior pedicle screws loosened, and a revision spondylodesis T8L1 was performed with bone cement augmentation of the pedicle screws. Although cement leakage was seen paravertebrally, the patient was asymptomatic and reported distinct pain relief, so that no further investigations were initiated. Three months later, instability was identified in the adjacent superior segment. A CT-scan of the chest now revealed multiple pulmonary cement embolisms. Corpectomy T7 and extension spondylodesis T6T9 with an anterior single rod were performed. The pulmonary embolisms remained clinically silent and lung function was normal 18 months after surgery. The risk for cement embolism after pedicle screw augmentation has been established at 1/119 or 0.8%. After vertebroplasty and kyphoplasty it ranges from 3 to 23%. The existing literature offers no clear strategy for prevention or treatment of pulmonary cement embolism. The authors feel, as far as prevention is concerned, that creating a cavity by means of balloon distention before pedicle screw augmentation allows to inject the cement under lower pressure, so that the incidence of cement embolism might be reduced. Treatment options include, besides surgical removal in symptomatic embolism, heparin or warfarin treatment for 3 to 6 months.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Pulmonary cement embolism during cement augmentation of pedicle screw fixation: a case report.

We present a case of pulmonary cement embolism during cement augmentation of pedicle screw fixation for severe degenerative scoliosis. During intra-operative C-arm fluoroscopy, a radio-opaque substance was incidentally noted to pulsate in synchrony with the patient's heartbeat. A diagnosis of pulmonary cement embolism was made. The patient was treated with coumarin for 6 months. He made a good ...

متن کامل

Foreign body granuloma after bone cement augmentation of pedicle screws for osteoporotic spinal surgery: A case report

Osteoporosis causes a decrease in bone mineral density. To overcome the decrease in fixation strength, a variety of techniques and devices have been developed, including cement augmentation of pedicle screws. Polymethyl methacrylate (PMMA), is commonly known as bone cement, and is widely used for implant fixation in various orthopaedic and spine surgery. In general, PMMA augmentation of pedicle...

متن کامل

Minimally Invasive Technique for PMMA Augmentation of Fenestrated Screws

PURPOSE To describe the minimally invasive technique for cement augmentation of cannulated and fenestrated screws using an injection cannula as well as to report its safety and efficacy. METHODS A total of 157 cannulated and fenestrated pedicle screws had been cement-augmented during minimally invasive posterior screw-rod spondylodesis in 35 patients from January to December 2012. Retrospecti...

متن کامل

Alternative technique of cement augmentation of loosened pedicle screws -- technical note and presentation of two cases.

An alternative cement augmentation technique for pedicle screws is described, which was applied in two patients with mono- and bisegmental non-union after preceding multisegmental lumbar fusion. The correctly placed pedicle screws in S1 with diameters of 6 and 8.5 mm had severely enlarged the screw cavities due to segmental instability. Revision screws with 10 mm diameter demonstrated sufficien...

متن کامل

Possible usage of cannulated pedicle screws without cement augmentation

BACKGROUND: The use of pedicle screws is becoming increasingly popular for spinal surgery practice as the technology advances. Screw pullout due to bone quality and loading conditions is one of the most common problems observed after pedicle screw fixation. Several solutions were studied to prevent screw pullout. These can be investigated under three main categories: screw design, expandable sc...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Acta orthopaedica Belgica

دوره 76 2  شماره 

صفحات  -

تاریخ انتشار 2010