Quantitative accuracy of 177Lu SPECT imaging for molecular radiotherapy

نویسندگان

  • Emilio Mezzenga
  • Vincenzo D'Errico
  • Marco D'Arienzo
  • Lidia Strigari
  • Koutla Panagiota
  • Federica Matteucci
  • Stefano Severi
  • Giovanni Paganelli
  • Andrew Fenwick
  • David Bianchini
  • Francesco Marcocci
  • Anna Sarnelli
چکیده

The purpose of this study is to investigate the optimal reference geometry for gamma camera calibration. Yet another question of interest was to assess the influence of the number of 3D Ordered Subsets Expectation Maximization (3D-OSEM) updates on activity quantification for SPECT imaging with 177Lu. The accuracy of 177Lu activity quantification was assessed both in small and in large objects. Two different reference geometries, namely a cylindrical homogeneous phantom and a Jaszczak 16 ml sphere surrounded by cold water, were used to determine the gamma camera calibration factor of a commercial SPECT/CT system. Moreover, the noise level and the concentration recovery coefficient were evaluated as a function of the number of 3D-OSEM updates by using the SPECT/CT images of the reference geometry phantoms and those of a cold Jaszczak phantom with three hot spheres (16ml, 8ml and 4ml), respectively. The optimal choice of the number of 3D-OSEM updates was based on a compromise between the noise level achievable in the reconstructed SPECT images and the concentration recovery coefficients. The quantitative accuracy achievable was finally validated on a test phantom, where a spherical insert composed of two concentric spheres was used to simulate a lesion in a warm background. Our data confirm and extend previous observations. Using the calibration factor obtained with the cylindrical homogeneous phantom and the Jaszczak 16 ml sphere, the recovered activity in the test phantom was underestimated by -16.4% and -24.8%, respectively. Our work has led us to conclude that gamma camera calibration performed with large homogeneous phantom outperforms calibration executed with the Jaszczak 16ml sphere. Furthermore, the results obtained support the assumption that approximately 50 OSEM updates represent a good trade-off to reach convergence in small volumes, meanwhile minimizing the noise level.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A comparison of 2D and 3D kidney absorbed dose measures in patients receiving 177Lu-DOTATATE

Objective(s): To investigate and compare quantitative accuracy of kidney absorbed dose measures made from both 2D and 3D imaging in patients receiving 177LuDOTATATE (Lutate) for treatment of neuroendocrine tumours (NETs). Methods: Patients receiving Lutate therapy underwent both whole body planar imaging and SPECT/CT imaging over the kidneys at time points 0.5, 4, 24, and 96-120 hours after inj...

متن کامل

Collimator-detector response compensation in molecular SPECT reconstruction using STIR framework

Introduction:It is well-recognized that collimator-detector response (CDR) is the main image blurring factor in SPECT.  In this research, we compensated the images for CDR in molecular SPECT by using STIR reconstruction framework. Methods: To assess resolution recovery capability of the STIR, a phantom containing five point sources along with a micro Derenzo p...

متن کامل

Impact of a commercial 3D OSEM reconstruction algorithm on the 177Lu activity quantification of SPECT/ CT imaging in a Molecular Radiotherapy trial

The purpose of this study was to investigate the influence of the Ordered Subsets Expectation Maximization (OSEM) reconstruction updates implemented in the 177Lu SPECT/CT imaging processing in molecular radiotherapy. A NEMA IEC Body PhantomTM was used to quantify activity in refillable spheres of five different sizes. Images were obtained with a hybrid dual-head SPECT-CT imaging system (Symbia ...

متن کامل

Design and Fabrication of Kidney Phantoms for Internal Radiation Dosimetry Using 3D Printing Technology.

Currently, the validation of multimodal quantitative imaging and absorbed dose measurements is impeded by the lack of suitable, commercially available anthropomorphic phantoms of variable sizes and shapes. To demonstrate the potential of 3-dimensional (3D) printing techniques for quantitative SPECT/CT imaging, a set of kidney dosimetry phantoms and their spherical counterparts was designed and ...

متن کامل

Determination of the optimum filter for qualitative and quantitative 99mTc myocardial SPECT imaging

Background: Butterworth, Gaussian, Hamming, Hanning, and Parzen are commonly used SPECT filters during filtered back-projection (FBP) reconstruction, which greatly affect the quality and size accuracy of image. Materials and Methods: This study involved a cardiac phantom in which 1.10 cm thick cold defect was inserted into its myocardium wall and filled with 4.0 μCi/ml (0.148 MBq/ml) 99mTc conc...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 12  شماره 

صفحات  -

تاریخ انتشار 2017