Fro Ba zC OBJECTIVES The purpose of this study was to compare the 1-year outcome between bioresorbable vascular scaffold (BVS) and everolimus-eluting metallic stent (EES) in ST-segment elevation myocardial infarction (STEMI) patients. BACKGROUND The Absorb BVS (Abbott Vascular, Santa Clara, California) is a polymeric scaffold approved for treatment of stable coronary lesions. Limited and not randomized data are available on its use in ST-segment elevation myocardial infarction (STEMI) patients. METHODS This study included 290 consecutive STEMI patients treated by BVS, compared with either 290 STEMI patients treated with EES or 290 STEMI patients treated with bare-metal stents (BMS) from the EXAMINATION (A Clinical Evaluation of Everolimus Eluting Coronary Stents in the Treatment of Patients With ST-segment Elevation Myocardial Infarction) trial, by applying propensity score matching. The primary endpoint was a device-oriented endpoint (DOCE), including cardiac death, target vessel myocardial infarction, and target lesion revascularization, at 1-year follow-up. Device thrombosis, according to the Academic Research Consortium criteria, was also evaluated. RESULTS The cumulative incidence of DOCE did not differ between the BVS and EES or BMS groups either at 30 days (3.1% vs. 2.4%, hazard ratio [HR]: 1.31 [95% confidence interval (CI): 0.48 to 3.52], p 1⁄4 0.593; vs. 2.8%, HR: 1.15 [95% CI: 0.44 to 2.30], p 1⁄4 0.776, respectively) or at 1 year (4.1% vs. 4.1%, HR: 0.99 [95% CI: 0.23 to 4.32], p 1⁄4 0.994; vs. 5.9%, HR: 0.50 [95% CI: 0.13 to 1.88], p 1⁄4 0.306, respectively). Definite/probable BVS thrombosis rate was numerically higher either at 30 days (2.1% vs. 0.3%, p 1⁄4 0.059; vs. 1.0%, p 1⁄4 0.324, respectively) or at 1 year (2.4% vs. 1.4%, p 1⁄4 0.948; vs. 1.7%, p 1⁄4 0.825, respectively), as compared with EES or BMS. CONCLUSIONS At 1-year follow-up, STEMI patients treated with BVS showed similar rates of DOCE compared with STEMI patients treated with EES or BMS, although rate of scaffolds thrombosis, mostly clustered in the early phase, was not negligible. Larger studies with longer follow-up are needed to confirm our findings. (J Am Coll Cardiol Intv 2015;8:189–97) © 2015 by the American College of Cardiology Foundation. m the *Thorax Institute, Hospital Clinic, University of Barcelona, Institut d’Investigacions Biomèdiques August Pi i Sunyer, rcelona, Spain; yMedizinische Klinik und Poliklinik für Kardiologie, Angiologie und innere Intensivmedizin, Mainz, Germany; ardiac Department, National University Heart Centre Singapore, National University Health System, Singapore; xCardiocenter, ABBR EV I A T I ON S
منابع مشابه
Everolimus-eluting bioresorbable stent vs. durable polymer everolimus-eluting metallic stent in patients with ST-segment elevation myocardial infarction: results of the randomized ABSORB ST-segment elevation myocardial infarction—TROFI II trial
AIMS Patients with ST-segment elevation myocardial infarction (STEMI) feature thrombus-rich lesions with large necrotic core, which are usually associated with delayed arterial healing and impaired stent-related outcomes. The use of bioresorbable vascular scaffolds (Absorb) has the potential to overcome these limitations owing to restoration of native vessel lumen and physiology at long term. T...
متن کاملBioresorbable Vascular Scaffolds Versus Metallic Stents in Patients With Coronary Artery Disease: ABSORB China Trial.
BACKGROUND The everolimus-eluting bioresorbable vascular scaffold (BVS) is designed to achieve results comparable to metallic drug-eluting stents at 1 year, with improved long-term outcomes. Whether the 1-year clinical and angiographic results of BVS are noninferior to current-generation drug-eluting stents has not been established. OBJECTIVES This study sought to evaluate the angiographic ef...
متن کاملBest of JACC: Cardiovascular Interventions
OBJECTIVES The purpose of this study was to compare the 1-year outcome between bioresorbable vascular scaffold (BVS) and everolimus-eluting metallic stent (EES) in ST-segment elevation myocardial infarction (STEMI) patients. BACKGROUND The Absorb BVS (Abbott Vascular, Santa Clara, California) is a polymeric scaffold approved for treatment of stable coronary lesions. Limited and not randomized d...
متن کامل1-Year Outcomes of Everolimus-Eluting Bioresorbable Scaffolds Versus Everolimus-Eluting Stents: A Propensity-Matched Comparison of the GHOST-EU and XIENCE V USA Registries.
OBJECTIVES The purpose of this study was to compare the 1-year outcomes of the ABSORB everolimus-eluting bioresorbable scaffold (BRS) (Abbott Vascular, Santa Clara, California) and the XIENCE everolimus-eluting stent (EES) (Abbott Vascular) in patients undergoing percutaneous coronary intervention. BACKGROUND Randomized studies of the ABSORB BRS have been performed in selected patient and les...
متن کاملLong-Term Efficacy and Safety of Everolimus-Eluting Bioresorbable Vascular Scaffolds Versus Everolimus-Eluting Metallic Stents
Second-generation metallic drug-eluting stents (DES) have been widely recommended for percutaneous coronary intervention. One of the main drawbacks of these devices is the risk of late adverse events (ie, beyond 1 year), particularly repeat revascularization, which are partly related to the persistence of the metallic stents in the coronary artery wall. In this context, the bioresorbable vascul...