Assessment of MRI issues at 3-Tesla for metallic surgical implants: findings applied to 61 additional skin closure staples and vessel ligation clips

نویسندگان

  • Amreeta Gill
  • Frank G Shellock
چکیده

PURPOSE Metallic skin closure staples and vessel ligation clips should be tested at 3-Tesla to characterize MRI issues in order to ensure patient safety. Therefore, metallic surgical implants were assessed at 3-Tesla for magnetic field interactions, MRI-related heating, and artifacts. METHODS A skin closure staple (Visistat Skin Stapler, staple, Polytetrafluoroethylene, PTFE, coated 316L/316LVM stainless steel; Teleflex Medical, Durham, NC) and a vessel ligation clip (Hemoclip Traditional, stainless steel; Teleflex Medical, Durham, NC) that represented the largest metallic sizes made from materials with the highest magnetic susceptibilities (i.e., based on material information) among 61 other surgical implants (52 metallic implants, 9 nonmetallic implants) underwent evaluation for magnetic field interactions, MRI-related heating, and artifacts using standardized techniques. MRI-related heating was assessed by placing each implant in a gelled-saline-filled phantom with MRI performed using a transmit/receive RF body coil at an MR system reported, whole body averaged SAR of 2.9-W/kg for 15-min. Artifacts were characterized using T1-weighted, SE and GRE pulse sequences. RESULTS Each surgical implant showed minor magnetic field interactions (20- and 27-degrees, which is acceptable from a safety consideration). Heating was not substantial (highest temperature change, ≤ 1.6°C). Artifacts may create issues if the area of interest is in the same area or close to the respective surgical implant. CONCLUSIONS The results demonstrated that it would be acceptable for patients with these metallic surgical implants to undergo MRI at 3-Tesla or less. Because of the materials and dimensions of the surgical implants that underwent testing, these findings pertain to 61 additional similar implants.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Assessment of MRI Issues at 3 Tesla for a New Metallic Tissue Marker

Purpose. To assess the MRI issues at 3 Tesla for a metallic tissue marker used to localize removal areas of tissue abnormalities. Materials and Methods. A newly designed, metallic tissue marker (Achieve Marker, CareFusion, Vernon Hills, IL) used to mark biopsy sites, particularly in breasts, was assessed for MRI issues which included standardized tests to determine magnetic field interactions (...

متن کامل

Laparoscopic Vascular Control Techniques in Donor Nephrectomy: Effects on Vessel Length

BACKGROUND Various techniques for vascular control have been used during urologic laparoscopic procedures. The importance of optimizing the vessel length and securing reliable vascular control are critical for procedures like laparoscopic donor nephrectomy. We aimed to determine the length of vessel lost by using 4 common techniques of vascular control in a fresh human cadaveric vascular model....

متن کامل

Ventricular Assist Device implant (AB 5000) prototype cannula: In vitro assessment of MRI issues at 3-Tesla

PURPOSE To evaluate MRI issues at 3-Tesla for a ventricular assist device (VAD). METHODS The AB5000 Ventricle with a prototype Nitinol wire-reinforced In-Flow Cannula and Out-Flow Cannula attached (Abiomed, Inc., Danvers, MA) was evaluated for magnetic field interactions, heating, and artifacts at 3-Tesla. MRI-related heating was assessed with the device in a gelled-saline-filled, head/torso ...

متن کامل

A revolutionary advance in skin closure compared to current methods.

Six pigs were used to evaluate the influence of three separate modalities on contaminated wounds. Full-thickness skin wounds on the abdomen were contaminated with 10(4) or 10(5) Staphylococcus aureus and then closed with one of three methods. The three closure modalities included (1) a new absorbable staple (Insorb) placed in the subcuticular tissue, (2) a braided Vicryl suture, and (3) percuta...

متن کامل

Assessment of MRI issues at 7 T for 28 implants and other objects.

OBJECTIVE Metallic implants are currently a contraindication for volunteer subjects and patients referred for 7-T examinations because of concerns related to magnetic field interactions and MRI-related heating. Artifacts may also be problematic. Therefore, the purpose of this investigation was to evaluate these MRI issues for 28 implants and other objects in association with a 7-T MR system. ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 14  شماره 

صفحات  -

تاریخ انتشار 2012