Accurate and reproducible mitral valvular blood flow measurement with three-directional velocity-encoded magnetic resonance imaging.
نویسندگان
چکیده
A new method for quantifying the transvalvular flow through the mitral valve (MV) based on three-directional velocity-encoded magnetic resonance imaging (MRI) is presented. For thirty time phases during one cardiac cycle, the three-dimensional (3D) velocity vector field of the blood flow is reconstructed from the MRI measurement. Retrospectively, for each time phase, the MV-plane is indicated manually in the velocity data and the flow through this plane is determined, representing the MV flow. Measurements are performed in 10 healthy volunteers. The new method is compared to the conventional, one-directional velocity-encoded MRI method for which an acquisition plane is positioned at the mitral valve at end-systole and remains fixed during the acquisition. The flow measurements with the new method correlate very well with the flow measured in the aorta (r(p)=0.92, p<0.01), whereas the conventional method shows no statistically significant correlation (r(p)=0.15, p=0.68). The low differences between the flow measured at the MV and the flow in the aorta proves high accuracy of the new method. Also, the new method shows very low intra- and interobserver variation, proving the high reproducibility. Three-directional velocity-encoded MRI is a patient-friendly and easy-to-use method suitable for quantifying accurately and reproducibly the transvalvular MV flow.
منابع مشابه
Reliability and reproducibility of trans-valvular flow measurement by 4D flow magnetic resonance imaging in acute myocardial infarct patients: two centre study
Background Time resolved 3-dimensional 3-directional phase contrast (PC) Magnetic Resonance Imaging (4D flow MRI) has advantage over 2-dimensional PC-MRI as it allows accurate quantification of stroke volumes (SV) using retrospective valve tracking and subsequent reconstruction of the transvalvular flow velocity throughout the cardiac cycle. However, the reliability of this method has not been ...
متن کاملDetection and quantification of valvular heart disease with dynamic cardiac MR imaging.
Magnetic resonance (MR) imaging is rapidly gaining acceptance as an accurate, reproducible, noninvasive method for optimal assessment of structural and functional parameters in patients with valvular heart disease. The severity of valvular regurgitation can be evaluated with cine gradient-echo MR imaging, which allows measurement of the area of the signal void corresponding to the abnormal flow...
متن کاملInfluence of Spatial Resolution in Three-dimensional Cine Phase Contrast Magnetic Resonance Imaging on the Accuracy of Hemodynamic Analysis
INTRODUCTION We aim to elucidate the effect of spatial resolution of three-dimensional cine phase contrast magnetic resonance (3D cine PC MR) imaging on the accuracy of the blood flow analysis, and examine the optimal setting for spatial resolution using flow phantoms. MATERIALS AND METHODS The flow phantom has five types of acrylic pipes that represent human blood vessels (inner diameters: 1...
متن کاملThree-Directional Evaluation of Mitral Flow in the Rat Heart by Phase-Contrast Cardiovascular Magnetic Resonance.
PURPOSE Determination of mitral flow is an important aspect in assessment of cardiac function. Traditionally, mitral flow is measured by Doppler echocardiography which suffers from several challenges, particularly related to the direction and the spatial inhomogeneity of flow. These challenges are especially prominent in rodents. The purpose of this study was to establish a cardiovascular magne...
متن کاملBlood Flow Simulation in an Aorta with a mild coarctation Using Magnetic Resonance Angiography and Finite Volume Method
Coarctation of the aorta is one of the five main congenital cardiovascular failures, accounting for 6–8 percent of these failures. This research aimed to simulate the blood flow of a seventeen-year-old male teen with a mild coarctation at one-third of his aorta's descending branch. The simulation was performed by extracting the domain and the input pulsatile velocity signal as the boundary cond...