Time-resolved in vivo luminescence dosimetry for online error detection in pulsed dose-rate brachytherapy.

نویسندگان

  • Claus E Andersen
  • Søren Kynde Nielsen
  • Jacob Christian Lindegaard
  • Kari Tanderup
چکیده

PURPOSE The purpose of this study is to present and evaluate a dose-verification protocol for pulsed dose-rate (PDR) brachytherapy based on in vivo time-resolved (1 s time resolution) fiber-coupled luminescence dosimetry. METHODS Five cervix cancer patients undergoing PDR brachytherapy (Varian GammaMed Plus with 192Ir) were monitored. The treatments comprised from 10 to 50 pulses (1 pulse/h) delivered by intracavitary/interstitial applicators (tandem-ring systems and/or needles). For each patient, one or two dosimetry probes were placed directly in or close to the tumor region using stainless steel or titanium needles. Each dosimeter probe consisted of a small aluminum oxide crystal attached to an optical fiber cable (1 mm outer diameter) that could guide radioluminescence (RL) and optically stimulated luminescence (OSL) from the crystal to special readout instrumentation. Positioning uncertainty and hypothetical dose-delivery errors (interchanged guide tubes or applicator movements from +/-5 to +/-15 mm) were simulated in software in order to assess the ability of the system to detect errors. RESULTS For three of the patients, the authors found no significant differences (P>0.01) for comparisons between in vivo measurements and calculated reference values at the level of dose per dwell position, dose per applicator, or total dose per pulse. The standard deviations of the dose per pulse were less than 3%, indicating a stable dose delivery and a highly stable geometry of applicators and dosimeter probes during the treatments. For the two other patients, the authors noted significant deviations for three individual pulses and for one dosimeter probe. These deviations could have been due to applicator movement during the treatment and one incorrectly positioned dosimeter probe, respectively. Computer simulations showed that the likelihood of detecting a pair of interchanged guide tubes increased by a factor of 10 or more for the considered patients when going from integrating to time-resolved dose verification. The likelihood of detecting a +/-15 mm displacement error increased by a factor of 1.5 or more. CONCLUSIONS In vivo fiber-coupled RL/OSL dosimetry based on detectors placed in standard brachytherapy needles was demonstrated. The time-resolved dose-rate measurements were found to provide a good way to visualize the progression and stability of PDR brachytherapy dose delivery, and time-resolved dose-rate measurements provided an increased sensitivity for detection of dose-delivery errors compared with time-integrated dosimetry.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Characterization of a fiber-coupled Al2O3:C luminescence dosimetry system for online in vivo dose verification during Ir brachytherapy

A prototype of a new dose-verification system has been developed to facilitate prevention and identification of dose delivery errors in remotely afterloaded brachytherapy. The system allows for automatic online in vivo dosimetry directly in the tumor region using small passive detector probes that fit into applicators such as standard needles or catheters. The system measures the absorbed dose ...

متن کامل

EGSnrc Monte Carlo calculated dosimetry parameters for Ir and Yb brachytherapy sources

This study presents the results of EGSnrc Monte Carlo calculations of the dose distribution surrounding a high dose rate Yb brachytherapy source and 14 high dose rate and pulsed dose rate Ir brachytherapy sources. Energy-weighted spectra of emitted photons, a full set of TG-43 dosimetry parameters, along-away dose tables, and a description of the materials and geometry used for each source are ...

متن کامل

Online in vivo dosimetry in high dose rate prostate brchytherapy with MOSkin detectors: in phantom feasibility study.

MOSkin detectors were studied to perform real-time in vivo dose measurements in high dose rate prostate brachytherapy. Measurements were performed inside an urethral catheter in a gel phantom simulating a real prostate implant. Measured and expected doses were compared and the discrepancy was found to be within 8.9% and 3.8% for single MOSkin and dual-MOSkin configurations, respectively. Result...

متن کامل

Determining TG-43 brachytherapy dosimetry parameters and dose distribution for a 131Cs source model CS-1

Background: Monte Carlo determination of TG-43 brachytherapy dosimetry parameters and dose distribution calculation for 131Cs source model CS-1 are presented in this study. Materials and Methods: The dose distribution was calculated around the 131Cs Model CS-1 located in the center of 30 cm ×30 cm ×30 cm water, and soft tissue phantoms cube using MCNP code by Monte Carlo method. The percentage ...

متن کامل

Rectal and Bladder Dose Measurements in the Intracavitary Applications of Cervical Cancer Treatment with HDR Afterloading System: Comparison of TPS Data with MOSFET Detector

Background: Intracavitary brachytherapy plays a major role in management of cervical carcinoma. Assessment of dose received by OAR’s therefore becomes crucial for the estimation of radiation toxicities in HDR brachytherapy. Objective: Purpose of this study is to evaluate the role of in vivo dosimetry in HDR brachytherapy and to compare actual doses delivered to OAR’ s with those calculate...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Medical physics

دوره 36 11  شماره 

صفحات  -

تاریخ انتشار 2009