Dosimetric performance of the new high‐definition multileaf collimator for intracranial stereotactic radiosurgery
نویسندگان
چکیده
The objective was to evaluate the performance of a high-definition multileaf collimator (MLC) of 2.5 mm leaf width (MLC2.5) and compare to standard 5 mm leaf width MLC (MLC5) for the treatment of intracranial lesions using dynamic conformal arcs (DCA) technique with a dedicated radiosurgery linear accelerator. Simulated cases of spherical targets were created to study solely the effect of target volume size on the performance of the two MLC systems independent of target shape complexity. In addition, 43 patients previously treated for intracranial lesions in our institution were retrospectively planned using DCA technique with MLC2.5 and MLC5 systems. The gross tumor volume ranged from 0.07 to 40.57 cm3 with an average volume of 5.9 cm3. All treatment parameters were kept the same for both MLC-based plans. The plan evaluation was performed using figures of merits (FOM) for a rapid and objective assessment on the quality of the two treatment plans for MLC2.5 and MLC5. The prescription isodose surface was selected as the greatest isodose surface covering >or= 95% of the target volume and delivering 95% of the prescription dose to 99% of target volume. A Conformity Index (CI) and conformity distance index (CDI) were used to quantifying the dose conformity to a target volume. To assess normal tissue sparing, a normal tissue difference (NTD) was defined as the difference between the volume of normal tissue receiving a certain dose utilizing MLC5 and the volume receiving the same dose using MLC2.5. The CI and normal tissue sparing for the simulated spherical targets were better with the MLC2.5 as compared to MLC5. For the clinical patients, the CI and CDI results indicated that the MLC2.5 provides better treatment conformity than MLC5 even at large target volumes. The CI's range was 1.15 to 2.44 with a median of 1.59 for MLC2.5 compared to 1.60-2.85 with a median of 1.71 for MLC5. Improved normal tissue sparing was also observed for MLC2.5 over MLC5, with the NTD always positive, indicating improvement, and ranging from 0.1 to 8.3 for normal tissue receiving 50% (NTV50), 70% (NTV70) and 90% (NTV90) of the prescription dose. The MLC2.5 has a dosimetric advantage over the MLC5 in Linac-based radiosurgery using DCA method for intracranial lesions, both in treatment conformity and normal tissue sparing when target shape complexity increases.
منابع مشابه
Impact of the high-definition multileaf collimator on linear accelerator-based intracranial stereotactic radiosurgery.
OBJECTIVES The impact of two multileaf collimator (MLC) systems for linear accelerator-based intracranial stereotactic radiosurgery (SRS) was assessed. METHODS 68 lesions formed the basis of this study. 2.5 mm leaf width plans served as reference. Comparative plans, with identical planning parameters, were based on a 5 mm leaf width MLC system. Two collimation strategies, with collimation fix...
متن کاملRadiation transmission, leakage and beam penumbra measurements of a micro‐multileaf collimator using GafChromic EBT film
Micro-multileaf collimator systems coupled to linear accelerators for radioneurosurgery treatments require a rigorous dosimetric characterization in order to be used in 3D conformal and intensity modulated stereotactic radiosurgery and radiotherapy applications. This characterization involves high precision measurements of leaf transmission, leakage and beam penumbra through the collimation sys...
متن کاملRadiochromic Film Dosimetry of Small Irregular Fields Shaped by a Micro-multileaf Collimator for Stereotactic Radiosurgery
متن کامل
SU-E-T-453: Optimization of Dose Gradient for Gamma Knife Radiosurgery.
PURPOSE The goals of stereotactic radiosurgery (SRS) are the ablation of target tissue and sparing of critical normal tissue. We develop tools to aid in the selection of collimation and prescription (Rx) isodose line to optimize the dose gradient for single isocenter intracranial stereotactic radiosurgery (SRS) with GammaKnife 4C utilizing the updated physics data in GammaPlan v10.1. METHODS ...
متن کاملDynamic conformal arc cranial stereotactic radiosurgery: implications of multileaf collimator margin on dose-volume metrics.
OBJECTIVE The effect of multileaf collimator (MLC) margin on target and normal tissue dose-volume metrics for intracranial stereotactic radiosurgery (SRS) was assessed. METHODS 118 intracranial lesions of 83 SRS patients formed the basis of this study. For each planning target volume (PTV), five separate treatment plans were generated with MLC margins of -1, 0, 1, 2 and 3 mm, respectively. Id...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 11 شماره
صفحات -
تاریخ انتشار 2010