Optimized saturation recovery protocols for T1-mapping in the heart: influence of sampling strategies on precision
نویسندگان
چکیده
BACKGROUND T1-mapping has the potential to detect and quantify diffuse processes such as interstitial fibrosis. Detection of disease at an early stage by measurement of subtle changes requires a high degree of reproducibility. Initial implementation of saturation recovery (SR) T1-mapping employed 3-parameter fitting which was highly accurate but was quite sensitive to noise; 2-parameter fitting greatly reduced the sensitivity to noise at the expense of a small degree of systematic bias. A recently introduced implementation that uses a variable readout flip angle greatly reduces systematic errors in T1-measurement thereby making it feasible to use SR methods with 2-parameter fitting with improved accuracy and precision. SR T1 mapping techniques with multi-heartbeat recovery times have been proposed to better sample the T1 recovery curve, but have not been evaluated for 2-parameter fitting. METHODS An analytic formulation for calculating the standard deviation (SD) for SR T1-mapping with 2-parameter fitting is developed and validated using Monte-Carlo simulation. The coefficient of variation is compared for a brute force optimization of sampling and for several previously described sampling schemes for T1 measurement over several uncertainty ranges. Experimental validation is performed in phantoms over a range of T1, and in-vivo both native and post-contrast. Pixel-wise SD maps are calculated for SR T1-mapping. RESULTS Sampling schemes that use a non-saturated anchor image and multiple (N) measurements at a single fixed saturation delay are found to be near optimum for the case of known T1 and are close to the brute force optimized solution over wide ranges of native and post-contrast T1 values. The fixed delay sampling scheme is simple to implement and provides an improvement over uniformly distributed schemes. CONCLUSIONS Sampling strategies for saturation recovery methods for myocardial T1-mapping have been optimized and validated experimentally. Reduced SD, or improved precision, may be achieved by using fixed saturation delays when considering native myocardium and post-contrast T1 ranges. Pixel-wise estimates of T1 mapping errors have been formulated and validated for SR fitting methods.
منابع مشابه
On the selection of sampling points for myocardial T1 mapping.
PURPOSE To provide a method for the optimal selection of sampling points for myocardial T1 mapping, and to evaluate how this selection affects the precision. THEORY The Cramér-Rao lower bound on the variance of the unbiased estimator was derived for the sampling of the longitudinal magnetization curve, as a function of T1 , signal-to-noise ratio, and noise mean. The bound was then minimized n...
متن کاملTemporally resolved parametric assessment of Z-magnetization recovery (TOPAZ): Dynamic myocardial T1 mapping using a cine steady-state look-locker approach.
PURPOSE To develop and evaluate a cardiac phase-resolved myocardial T1 mapping sequence. METHODS The proposed method for temporally resolved parametric assessment of Z-magnetization recovery (TOPAZ) is based on contiguous fast low-angle shot imaging readout after magnetization inversion from the pulsed steady state. Thereby, segmented k-space data are acquired over multiple heartbeats, before...
متن کامل3D myocardial T 1 mapping using saturation recovery
PURPOSE To propose a 3D quantitative high-resolution T1 mapping technique, called 3D SASHA (saturation-recovery single-shot acquisition), which combines a saturation recovery pulse with 1D-navigator-based-respiratory motion compensation to acquire the whole volume of the heart in free breathing. The sequence was tested and validated both in a T1 phantom and in healthy subjects. MATERIALS AND ...
متن کاملT1-mapping in the heart: accuracy and precision
The longitudinal relaxation time constant (T1) of the myocardium is altered in various disease states due to increased water content or other changes to the local molecular environment. Changes in both native T1 and T1 following administration of gadolinium (Gd) based contrast agents are considered important biomarkers and multiple methods have been suggested for quantifying myocardial T1 in vi...
متن کاملMyocardial T1-mapping at 3T using saturation-recovery: reference values, precision and comparison with MOLLI
BACKGROUND Myocardial T1-mapping recently emerged as a promising quantitative method for non-invasive tissue characterization in numerous cardiomyopathies. Commonly performed with an inversion-recovery (IR) magnetization preparation at 1.5T, the application at 3T has gained due to increased quantification precision. Alternatively, saturation-recovery (SR) T1-mapping has recently been introduced...