Physical and dosimetric characteristic of high‐definition multileaf collimator (HDMLC) for SRS and IMRT

نویسندگان

  • Dayananda Shamurailatpam Sharma
  • Prabhakar M Dongre
  • Vaibav Mhatre
  • Malhotra Heigrujam
چکیده

Physical and dosimetric characteristics of HDMLC were studied for SRS6, 6, and 10 MV X-rays from Novalis Tx. This in-built tertiary collimator consists of 60 pairs (32 × 0.25 cm; 26 × 0.5 cm and 2 × 0.7 cm) of leaves. Properties of HDMLC studied included alignment, readout and radiation field congruence, radiation penumbra, accuracy and reproducibility of leaf position and gap width, static and dynamic leaf shift, tongue-and-groove effect, leaf transmission and leakage, leaf travel speed, and delivery of dynamic conformal arc and IMRT. All tests were performed using a calibrated ionization chamber, film dosimetry and DynaLog file analysis. Alignment of leaves with isocenter plane was better than 0.03 cm at all gantry and collimator positions. The congruence of HDMLC readout and radiation field agreed to within ± 0.03 cm for filed sizes ranging from 1 × 1 to 20 × 20 cm2. Mean 80% to 20% penumbra width parallel (perpendicular) to leaf motion was 0.24 ± 0.05 (0.21 ± 0.02) cm, 0.37 ± 0.12 (0.29 ± 0.07) cm, and 0.51 ± 0.13 (0.43± 0.07) cm for SRS6, 6, and 10 MV X-rays, respectively. Circular field penumbra was comparable to corresponding square field. Average penumbra of 1 × 20 cm2 field was effectively constant over off-axis positions of up to 12 cm with mean value of 0.16 (± 0.01) cm at 1.5 cm depth and 0.38 (± 0.04) cm at 10 cm depth. Minimum and maximum effective penumbra along the straight diagonal edge of irregular fields increased from 0.3 and 0.32 cm at 70° steep angle to 0.35 and 0.56 cm at 20° steep angle. Modified Picket Fence test showed average FWHM of 0.18 cm and peak-to-peak distance of 1.99 cm for 0.1 cm band and 2 cm interband separation. Dynamic multileaf collimation (DMLC) output factor remained within ± 1% for 6 MV and ± 0.5% for 10 MV X-rays at all gantry positions, and was reproducible within ± 0.5% over a period of 14 months. The static leaf shift was 0.03 cm for all energies, while dynamic leaf shift was 0.044 cm for 10 MV and 0.039 cm for both SRS6 and 6 MV X-rays. The dose depression and corresponding tongue-and-groove size were 24% and 0.17 cm for 6 MV and 19% and 0.20 cm for 10 MV X-rays. Average transmission through HDMLC was 1.09%, 1.14% and 1.34% for SRS6, 6 and 10 MV X-rays. Analysis of DynaLog files for leaf speed test in arc dynamic mode, delivery test of dynamic conformal arc, and step-and-shoot and sliding window IMRT showed at least 95% or more of the error counts had misplacements < 0.2 cm, with maximum root mean square (RMS) error value calculated at 0.13cm. Accurate and reproducible leaf position and gap width, and less leakage and small consistent penumbra over the fields demonstrate HDMLC suitable for high-dose resolution SRS and IMRT.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The Dosimetric Effects of Different Multileaf Collimator Widths on Physical Dose Distributions

Introduction: Geometric changes in the multileaf collimator (MLC) led to dosimetric considerations in intensity-modulated radiation therapy (IMRT) due to the number and size of the pixels in the intensity map, which are determined by the MLC leaf width. In this study, we evaluated the dosimetric effects of different MLC widths on physical dose distributions for IMRT plans. Materials and Method...

متن کامل

Beam characteristic of P.E collimators Add-on multileaf collimator

  Introduction:    Multileaf collimators (MLCs) are computer-controlled devices which are Useful and cost-effective tools for conformal therapy and intensity modulation radiotherapy (IMRT). Nowadays MLCs are important and standard configuration of the new medical linear accelerators which are substitute of conventional Blocks and has eliminated the ...

متن کامل

Physical and Dosimetric Aspect of Euromechanics Add-on Multileaf Collimator on Varian Clinac 2100 C/D

Background: Before treatment planning and dose delivery, quality assurance of multi-leaf collimator (MLC) has an important role in intensity-modulated radiation therapy (IMRT) due to the creation of multiple segments from optimization process.Objective: The purpose of this study is to assess the quality control of MLC leaves using EBT3 Gafchromic films.Material and Methods: Leaf Position accur...

متن کامل

Investigation of physical penumbra definition in treatment planning

Background: Due to the small size of the beamlets in IMRT (intensity modulated radiotherapy), physical penumbra is one of the important dosimetric parameters and small changes in the penumbra have a notable impact on the results. The physical penumbra width is defined as the lateral distance between two specified isodose curves at a specified depth of phantom. In this study, after demonstrating...

متن کامل

Is 5 mm MMLC suitable for VMAT‐based lung SBRT? A dosimetric comparison with 2.5 mm HDMLC using RTOG‐0813 treatment planning criteria for both conventional and high‐dose flattening filter‐free photon beams

The aim of this study is to assess the suitability of 5 mm millennium multileaf collimator (MMLC) for volumetric-modulated arc therapy (VMAT)-based lung stereotactic body radiotherapy (SBRT). Thirty lung SBRT patient treatment plans along with their planning target volumes (ranging from 2.01 cc to 150.11 cc) were transferred to an inhomogeneous lung phantom and retrospectively planned using VMA...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 12  شماره 

صفحات  -

تاریخ انتشار 2011